Hier Klicken für das PDF Dokument!!!

            Standards                           Series 1409 Capacitor
USES:
·   Reference Standard                                Reference Standard of Capacitance
· Working Standard                              Introduction
· Verification of Calibration of LCR Meters     The 1409 Standard Capacitors are fixed mica capacitors of very high stability for use as two- or
· Capacitance Measurement Functions             three-terminal reference or working standards in the laboratory.
· Verification of Calibration of Multimeters
· For Calibrating Instrumentation               Description
                                                Typical capacitors, observed over more than 20 years, have shown random fluctuations of less
                                                than +0.01% in measured capacitance with no evidence of systematic drift.

FEATURES:                                       These capacitor units consist of a silvered-mica and foil pile, spring-held in a heavy metal clamp-
                                                ing structure for mechanical stability. The units are selected for low dissipation factor and are sta-
· Two and Three terminal calibration pro-       bilized by heat cycling. They are housed, with silica gel to provide continuous desiccation, in cast
    vided                                       aluminum cases, sealed with high-temperature potting wax. A well is provided in the wall of the
                                                case for the insertion of a dial-type thermometer. Three jack-top binding posts are provided on the
· Available in 0.001µF, 0.01µF, 0.1µF,          top of the case and removable plugs on the bottom, for convenient parallel connection without
   and 1.0µF                                    error.

· +0.01%/year stability
· Calibration accuracy +0.02%                                                                                       Figure 1
                                                                                                                     Change in capacitance as
                                                                                                                    a function of frequency for
                                                                                                                    typical 1409 Capacitors.
                                                                                                                    The 1-kHz value on the
                                                                                                                    plot should be used as a
                                                                                                                    basis of reference in estir-
                                                                                                                    nating frequency errors.




                                                                                                                    Figure 2
                                                                                                                    Dissipation factor as a
                                                                                                                    function of frequency.




                                                   HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS · DECADES · STANDARDS


                                                                                                       For more information about special
                                                                                                       purchase, rent & lease options, call

                                                           IET LABS, INC.

                                               Formerly manufactured by QuadTech
                                                   & GenRad (General Radio)

                                               CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                                                   Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
Series 1409 Capacitor

Calibration:           A certificate of calibration is supplied with each capacitor, giving the both two- and three-terminal measured
                       capacitances at 1kHz and at 23°±1°C. The measured value is the capacitance added when the standard is
                       plugged directly into the binding posts. This value is obtained by comparison, to a precision better than +/- 0.01%,
                       with working standards whose absolute values are known to an accuracy typically +/- 0.01%, determined and
                       maintained in terms of reference standards periodically measured by the National Institute of Standards and
                       Technology.



Adjustment Accuracy:       Within +/- 0.5% of the nominal capacitance value (two terminal) marked on the case.

Stability:                  Capacitance change is less than 0.01% per year.

Temperature Coefficient of Capacitance:
                               +35 +/-10ppm / °C between 10°C and 70°C.

Dissipation Factor:         Less than 0.0003 at 1kHz and 23° C (See Figure 2)
                                     Measured dissipation factor at 1 kHz is stated in the certificate to an accuracy of +/-0.00005.

Series Inductance:          0.050µH for 1409-F and 1409-L
                                     0.055µH for the 1409-T and 1409-Y

Series Resistance At 1MHz:
                                        0.02 V except for the 1409-Y, which is 0.03V

Frequency Characteristics:
                                        See Figure 1. Series resistance varies as the square root of the frequency for frequencies above
                       100kHz.

Approximate Terminal Capacitance:
                              From H Terminal to case (G), 12 to 50pF.
                                      From L Terminal (outside foils of capacitor) to case, 300 to 1300pF.

Leakage Resistance:         5000 ohm-farads or 100G V, whichever is less.

Maximum Voltage:            500V pk up to 10kHz.

Mechanical:                 Sealed Case.

Dimensions:                 (w x h x d):          1409-Y:   3.25 x 5.63 x 2.69in (83 x 143 x 69mm)
                                                            1409-F: 3.25 x 4 x 2in (83 x 102 x 51mm)
                                                                      1409-L: 3.25 x 4 x 2in (83 x 102 x 51mm)
                                                                                1409-T: 3.25 x 4 x 2in (83 x 102 x 51mm)

Weight:                     1.25lb (0.6kg) net, 4lb(1.9kg) shipping ; the 1409-Y is heavier by approximately 1lb (0.5kg).

  Ordering Information
   Reference Standard Capacitor

   Catalog Number     Item                                            Includes:
   1409-9706          1409-F, 0.001µF
   1409-9712          1409-L,0.01µF                                   Calibration Certificate Traceable to NIST
   1409-9720          1409-T, 0.1µF
   1409-9725          1409-Y, 1.0µF
                                                                      Optional Accessories:
                                                                      Calibration Data




                               IET LABS, INC.


                                           CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                                              Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de