Hier Klicken für das PDF Dokument!!!
Digibridge 1715 LCR Meter
Passive Component Tester with RS-232
USES: Introduction
· Production Testing of LCR Components The QuadTech 1715 LCR Digibridge is an economically priced, easily pro-
grammable, micro-processor based passive component tester for fast, precise
· Component Design and Evaluation
impedance measurements. It measures eleven different parameters with a
· Component Inspection and Screening basic accuracy of 0.2%. The 1715 provides four different test frequencies and
· Quality Assurance Testing two test voltages for testing flexibility. For remote operation and automated
system applications, the unit comes standard with an RS-232 interface, plus
· Material Testing IEEE-488 and Handler interfaces are available as options.
Description
FEATURES: 11 Measurement Parameters: Measure and display two of 11 impedance
· 11 Measurement Parameters parameters simultaneously with a basic accuracy of 0.2%. Test results can also
be shown as Deviation or %Deviation from nominal value.
· 0.2% Basic Measurement Accuracy
Test Signal: Four user selectable test frequencies: 100Hz, 120Hz, 1kHz, and
· Selectable Test Frequency: 100Hz, 10kHz (9.6kHz), plus two test voltages: 0.25V and 1.0V.
120Hz, 1kHz, 10kHz Display: 40 x 4 character LCD display for user friendly menu programming
· Selectable Test Voltage: 0.25V, 1.0V and measured results. Simultaneously displays primary/secondary values and
· Measurement Speed to 25/second binning or comparator results.
Setup Storage: Storage of 1 setup with parameter, binning and compare infor-
· LCD Display mation saved. Auto recall of this setup upon power-up of instrument.
· RS-232 Interface Standard Auto Ranging: For maximum measurement integrity the 1715 automatically
· IEEE and Handler Interfaces, Optional selects the appropriate measurement range and instrument source impedance
· Automatic Open/Short Zeroing (25, 100, 1k, 10k or 100k), dependent upon the impedance of the
device under test.
· Automatic Hi/Lo Comparator Limits Pass/Fail Testing: The 1715 provides a programmable Hi/Lo comparator
· Pass/Fail Sorting (8 Bins) function as well as eight sorting bins for categorization of components.
· Storage of 1 Setup Open/Short Zeroing: Reduce the effects of error sources (stray
admittance/residual impedance) between the device under test and the cali-
· Automatic Source Impedance Selection
brated connection to the measuring instrument.
· Keypad Lockout Interfaces: For remote control, or adaption to a production environment, the
· Easy Operation 1715 instrument includes an RS-232 interface as standard equipment. Both an
· Programmable Delay Time IEEE-488 and a Handler interface are available as optional interfaces.
CompuMess Elektronik GmbH
Lise-Meitner-Str.1, 85716 Unterschleissheim
Tel 089-321501-0 Fax 089-321501-11
http://www.compumess.de oder http://www.netzteile.de
HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS
1715 Digibridge LCR Features
Parameter Measurement Range Interfaces (Standard): RS-232
Ls, Lp 0.01uH - 9.999kH Interfaces (Optional): IEEE-488 and Handler
Cs, Cp 0.01pF - 99.999mF
Rs, Rp 0.1m - 99.99M Test Terminals: Front: 4 BNC & 1 GND
D 0.0001 - 9999 Dimensions: (w x h x d): 12.5 x 4.0 x 8.0in
Q 0.0001 - 9999 (312.5 x 100.0 x 200.0mm)
lZl 0.1m - 99.99M
Weight: 10.85 lbs. (5kg) net, 15.2 lbs. (7kg) ship
Xs 0.1m - 99.99M
Phase -180.00° - +180.00° Environmental: Specifications: +15°C to +35°C, 75% RH
Accuracy: Basic LCR ±0.2% Operating: 10°C to +40°C
Storage: 0°C to +50°C
Basic DQ ±0.002
Humidity: 10 -90% RH
Test Frequency: 100Hz, 120Hz, 1kHz and 10kHz (9.6kHz) Pollution Degree 2
Accuracy ±0.2% Installation Category II
Test Signal: 0.25V or 1.0V RMS, ±(10% + 3mV) Power: · 90 - 125V AC
Measurement Speed: Fast: 25 measurements/second · 190 - 250V AC
Medium: 8 measurements/second · 48 - 62 Hz
Slow: 2 measurements/second · 50VA max
Measurement Mode: Continuous or Trigger
Trigger: Internal
External (IEEE or Handler)
Manual
Trigger Delay: 0 - 9999ms
Ranging: Automatic or Hold Range
Equivalent Circuit: Series or Parallel
Zeroing: Open and short circuit compensation
Source Impedance: 25, 100, 1k, 10k or 100k
(based on DUT impedance)
Comparator: Primary Nominal
Primary Hi/Lo Limits (Value or %)
Secondary Hi/Lo Limuts (Value or %)
Indication: Audible Alarm programmable: HI LO or
OFF for Pass or Fail Result
Binning: Primary Nominal
Primary Hi/Lo Limits (8 Bins in %)
Secondary Hi/Lo Limits (Value)
Display: 40 by 4 Character LCD display
Results Format: Value
Deviation
% Deviation
Bin Number
Setup Storage: Store 1 setup, Auto recall upon power-up
Lock: Keypad Lockout
Ordering Information
1715 LCR Digibridge Optional Accessories:
B/A Calibration Data
Includes:
1700-01 Axial/Radial Component Test Fixture
QuadTech P/N Item
1700-02 Axial/Radial Remote Test Fixture
150710 1715 Instruction Manual
1700-03 Lead Set: 4 BNC Connectors to 2 Kelvin Clips
1700-03 Lead Set: 4 BNC Connectors to 2 Kelvin Clips
1700-04 Lead Set: 4 BNC Connectors to 4 Banana Plugs
4200-0300 AC Power Cord
1700-05 Lead Set: 4 BNC Connectors to Chip Component Tweezers
520023/520025 Power Line Fuses (0.630A 250V/0.315A 250V)
7000-07 Low Voltage Chip Component Test Fixture
Calibration Certificate Traceable to NIST
700171 IEEE-488 & Handler Interface
CompuMess Elektronik GmbH
Lise-Meitner-Str.1, 85716 Unterschleissheim
Tel 089-321501-0 Fax 089-321501-11
http://www.compumess.de oder http://www.netzteile.de