3010 :: 8PPM LABORATORY REFERENCE
1 YEAR BEST ACCURACY SPECIFICATIONS
FUNCTION FUNCTION MORE INFORMATION
Range : ±0 to 1020V
High power output for driving
Best Resolution : 0.1uV analogue multimeters.
Best 1 Year Accuracy : ±8ppm of
Voltage DC setting
kV Amplifier available to extend
output to 10kV
Click here for details
Range : ±0 to 30Amps
30A output with extended duty
Best Resolution : 100pA cycle
Best 1 Year Accuracy : ±50ppm of
setting (up to 10 minutes continuous
output at 20A)
Use with clamp coil option for
Current DC clamp meter
calibration up to 1500A
Transconductance Amplifier
available to extend output to 100A
Click here for details
Passive Range : 0 to 1GOhm
2 & 4 Wire Modes
Best 1 Year Accuracy : ±8ppm of
Resistance setting Passive decade resistors
Active DC 2 Wire Operation
Range : 0 to 30MOhm (7 ranges)
Resolution : 100ppm of range
Best 1 Year Accuracy : ±0.01% of Continuously variable across
Resistance setting
range
Range : 0 to 1020V
High power output for driving
Best Resolution : 1uV analogue multimeters.
Frequency Range (Sine) : 10Hz to
Voltage AC 500KHz
Best 1 Year Accuracy : ±0.012% of kV Amplifier available to extend
setting
output to 5kV
Click here for details
Range : 0 to 30A
30A output with extended duty
Best Resolution : 1nA cycle
Frequency Range (Sine) : 10Hz to
10KHz (up to 10 minutes continuous
Best 1 Year Accuracy : ±0.03% of output at 20A)
setting
Use with clamp coil option for
Current AC clamp meter
calibration up to 1500A
Transconductance Amplifier
available to extend output to 70A
Click here for details
High accuracy DDS generated
Range : 10Hz to 10MHz frequency
Frequency Best 1 Year Accuracy : referenced to a TCXO temperature
1ppm compensated reference
Range : 1nF to 100uF
Passive decade capacitors for the
Best 1 Year Accuracy : ±0.2% of best
Capacitance setting
compatibility with all bridges /
meters
Active Range : 1uF to 10mF
Best 1 Year Accuracy : ±0.25% of
Continuously variable across
Capacitance* setting
range
(Option)
Range : 1mH to 10H
Passive decade inductors for the
Inductance*
Best 1 Year Accuracy : ±0.5% of best
setting
compatibility with all bridges /
meters
Thermocouple Range : -210°C to 1800°C 7 Types Built-In
Best 1 Year Accuracy : ±0.04°C of Additional types available using
Source* setting
Virtual Front Panel Software
PRT (RTD) Range : -100°C TO 800°C (PT100) Additional RTD types available
Best 1 Year Accuracy : ±0.01% of using
Source* setting
Virtual Front Panel Software
Range : 0.3W to 1MW
DC Power* Best 1 Year Accuracy : ±0.08% of
setting
Range : 0.3W to 1MW
Phase Adjustment : 0 ± 180°
AC Power* Frequency Range : 40 to 400Hz
Best 1 Year Accuracy : ±0.06% of Incorporates Dynamic Digital
setting Phase Compenstation (DDPC) to
dynamically
Range : 0 to 359.9°
Eliminate errors caused by
Best Resolution : 0.1° capacitive or inductive loads
Best Life Accuracy : 0.2°
Choice of 5 waveforms for
Phase* calibration of power harmonics
function of power meters
Also includes upload function
for arbritrary waveform
generation.
Amplitude Range : 2mV to 50V / Div
(8mV to 400V Pk-Pk)
Best 1 Year Accuracy : ±0.01% of 2 high performance options for calibration
setting of oscilloscopes to 350MHz and a
higher bandwith option to 620MHz.
Timebase Range : 5s to 2ns / Div
Best 1 Year Accuracy : ±5ppm Both options provide a 1kHz square wave with a
maximum output of 400V Pk-Pk allowing
350MHz option calibration of oscilloscopes up to 50V/Div
Levelled Sweep Range : 5MHz to amplitude ranges at 8 graticules.
Oscilloscope 350MHz
High accuracy time markers provide a comb
Best 1 Year Accuracy : 1db
Calibration* waveform with 20% high / 80% low to provide easy
620MHz option triggering on both analogue and digital
Levelled Sweep Range : 5MHz to oscilloscopes.
620MHz
Best 1 Year Accuracy : 1db A dedicated BNC output provides all oscilloscope
calibration signals, avoiding lead changes during
50kHz Reference calibration.
Best 1 Year Accuracy (Frequency) :
30ppm A separate trigger output is also provided.
Best 1 Year Accuracy (Level) : ±0.5%
Optical Tacho Range : 240 to 60,000 RPM
Best 1 Year Accuracy : ±0.003%
Calibration*
Calibration Clamp Meter Calibration to 1500A
2/10/50 turn coil
Workstation Insulation Tester Calibration
with integrated
Continuity resistance to 1kOhm
Insulation resistance to 2GOhms /
1000V
2/10/50 Turn Insulation test voltage measure to
1000V
Clamp Coil Thermocouple Meter Calibration
7 Types simulated
Insulation Range : -210°C to 1800°C Combined all-in-one solution to
extend the workload of the
Tester Process Control Instruments
DCV measure 200mV, 2V & 20V calibrator to cover five additional
ranges
Calibration DCI measure 20mA
types of instrument in one
DCV/I source from calibrator convenient & cost effective
Thermocouple Optical Tacho Calibration workstation.
Simulation
240 to 60,000 RPM
Use with Virtual Front Panel
DCV/DCI software (supplied)
measurement or ProCal Calibration software
Only for use with a 3000 Series
function for Calibrator
process control
instrumentation
Optical
Tachometer
Calibration
Voltage Range : 0 to 33V
Best 1 Year Accuracy : ±0.02% of Use with Virtual Front Panel
setting software (supplied)
Power Supply
or ProCal Calibration software
Calibration* Current Range : 0 to 4A
Best 1 Year Accuracy : ±0.05% of
Only for use with a 2000 Series
setting Calibrator
GENERAL SPECIFICATIONS
Double the time since last used up to 20 minutes maximum
Warm Up Time
RS232
Standard
Interface
GPIB / USB
Optional
Interface
IGUARD protection - detects & protects against applied voltages on the output
Output
Protection
Storage : -5°C to +60°C
Temperature Operation : 0°C to +50°C
Calibration 18°C to 25°C
Performance
Specifications apply between 18°C and 25°C
Temperature Outside this range an allowance of 0.18 x 1 year specification per °C should be
Coefficient added
Operation <80% to 30°C, <70% to 40°C , <40% to 50°C
Relative Storage : <95% (non-condensing)
Humidity
Operation : 3000m (10,00ft) Maximum
Altitude Transit : 12000m (40,000ft) Maximum
The calibrator line inlet plug must be earthed
EMC & Safety See D.O.C. for full details
100V
Low Analogue
Isolation
Line voltage selectable : 110V / 230V
Line Power Line frequency : 50Hz to 60Hz
Line voltage variation : -6% +10%
28 Watts (Standby)
Power
Consumption
Calibrator Only : 43cm x 46cm x 19mm
Dimensions Calibrator in shipping box : 58cm x 56cm x 37cm
Calibrator in soft carry case : 49cm x 50cm x 19cm
Calibrator in hard transit case : 55cm x 56cm x 26cm
Calibrator Only : 14.5kg
Weight Calibrator in shipping box : 17kg
Calibrator in soft carry case : 16Kg
Calibrator in hard transit case : 24kg