Hier Klicken für das PDF Dokument!!!

3010 :: 8PPM LABORATORY REFERENCE




1 YEAR BEST ACCURACY SPECIFICATIONS

 FUNCTION                  FUNCTION                               MORE INFORMATION
                           Range : ±0 to 1020V
                                                                High power output for driving
                           Best Resolution : 0.1uV                 analogue multimeters.
                           Best 1 Year Accuracy : ±8ppm of
 Voltage DC                setting
                                                               kV Amplifier available to extend
                                                                     output to 10kV
                                                                       Click here for details

                           Range : ±0 to 30Amps
                                                                30A output with extended duty
                           Best Resolution : 100pA                           cycle
                           Best 1 Year Accuracy : ±50ppm of
                           setting                               (up to 10 minutes continuous
                                                                        output at 20A)
                                                                Use with clamp coil option for
  Current DC                                                             clamp meter
                                                                   calibration up to 1500A

                                                                  Transconductance Amplifier
                                                               available to extend output to 100A
                                                                       Click here for details

    Passive                Range : 0 to 1GOhm
                                                                     2 & 4 Wire Modes
                           Best 1 Year Accuracy : ±8ppm of
  Resistance               setting                                 Passive decade resistors

   Active DC                                                          2 Wire Operation
                           Range : 0 to 30MOhm (7 ranges)
                           Resolution : 100ppm of range
                           Best 1 Year Accuracy : ±0.01% of      Continuously variable across
  Resistance               setting
                                                                           range

                           Range : 0 to 1020V
                                                                High power output for driving
                           Best Resolution : 1uV                   analogue multimeters.
                           Frequency Range (Sine) : 10Hz to
  Voltage AC               500KHz
                           Best 1 Year Accuracy : ±0.012% of   kV Amplifier available to extend
                           setting
                                                                      output to 5kV
                                                                       Click here for details
                Range : 0 to 30A
                                                     30A output with extended duty
                Best Resolution : 1nA                             cycle
                Frequency Range (Sine) : 10Hz to
                10KHz                                 (up to 10 minutes continuous
                Best 1 Year Accuracy : ±0.03% of             output at 20A)
                setting
                                                     Use with clamp coil option for
 Current AC                                                   clamp meter
                                                        calibration up to 1500A

                                                      Transconductance Amplifier
                                                    available to extend output to 70A
                                                            Click here for details
                                                      High accuracy DDS generated
                Range : 10Hz to 10MHz                           frequency
 Frequency      Best 1 Year Accuracy :              referenced to a TCXO temperature
                1ppm                                      compensated reference
                Range : 1nF to 100uF
                                                     Passive decade capacitors for the
                Best 1 Year Accuracy : ±0.2% of                    best
Capacitance     setting
                                                      compatibility with all bridges /
                                                                  meters
   Active       Range : 1uF to 10mF
                Best 1 Year Accuracy : ±0.25% of
                                                      Continuously variable across
Capacitance*    setting
                                                                range
  (Option)
                Range : 1mH to 10H
                                                     Passive decade inductors for the
  Inductance*
                Best 1 Year Accuracy : ±0.5% of                    best
                setting
                                                     compatibility with all bridges /
                                                                  meters
Thermocouple    Range : -210°C to 1800°C                    7 Types Built-In
                Best 1 Year Accuracy : ±0.04°C of    Additional types available using
  Source*       setting
                                                      Virtual Front Panel Software

 PRT (RTD)      Range : -100°C TO 800°C (PT100)      Additional RTD types available
                Best 1 Year Accuracy : ±0.01% of                  using
  Source*       setting
                                                      Virtual Front Panel Software
                Range : 0.3W to 1MW
 DC Power*      Best 1 Year Accuracy : ±0.08% of
                setting




                Range : 0.3W to 1MW
                Phase Adjustment : 0 ± 180°
 AC Power*      Frequency Range : 40 to 400Hz
                Best 1 Year Accuracy : ±0.06% of     Incorporates Dynamic Digital
                setting                             Phase Compenstation (DDPC) to
                                                             dynamically
                  Range : 0 to 359.9°
                                                              Eliminate errors caused by
                  Best Resolution : 0.1°                     capacitive or inductive loads
                  Best Life Accuracy : 0.2°

                                                           Choice of 5 waveforms for
    Phase*                                               calibration of power harmonics
                                                            function of power meters
                                                          Also includes upload function
                                                             for arbritrary waveform
                                                                    generation.
                  Amplitude Range : 2mV to 50V / Div
                  (8mV to 400V Pk-Pk)
                  Best 1 Year Accuracy : ±0.01% of           2 high performance options for calibration
                  setting                                        of oscilloscopes to 350MHz and a
                                                                 higher bandwith option to 620MHz.
                  Timebase Range : 5s to 2ns / Div
                  Best 1 Year Accuracy : ±5ppm            Both options provide a 1kHz square wave with a
                                                             maximum output of 400V Pk-Pk allowing
                  350MHz option                              calibration of oscilloscopes up to 50V/Div
                  Levelled Sweep Range : 5MHz to                  amplitude ranges at 8 graticules.
 Oscilloscope     350MHz
                                                           High accuracy time markers provide a comb
                  Best 1 Year Accuracy : 1db
 Calibration*                                            waveform with 20% high / 80% low to provide easy
                  620MHz option                               triggering on both analogue and digital
                  Levelled Sweep Range : 5MHz to                           oscilloscopes.
                  620MHz
                  Best 1 Year Accuracy : 1db             A dedicated BNC output provides all oscilloscope
                                                         calibration signals, avoiding lead changes during
                  50kHz Reference                                            calibration.
                  Best 1 Year Accuracy (Frequency) :
                  30ppm                                     A separate trigger output is also provided.
                  Best 1 Year Accuracy (Level) : ±0.5%



Optical Tacho     Range : 240 to 60,000 RPM
                  Best 1 Year Accuracy : ±0.003%
Calibration*

  Calibration     Clamp Meter Calibration to 1500A
                  2/10/50 turn coil
 Workstation      Insulation Tester Calibration

with integrated
                  Continuity resistance to 1kOhm
                  Insulation resistance to 2GOhms /
                  1000V
 2/10/50 Turn     Insulation test voltage measure to
                  1000V
  Clamp Coil      Thermocouple Meter Calibration
                  7 Types simulated
  Insulation      Range : -210°C to 1800°C                 Combined all-in-one solution to
                                                              extend the workload of the
    Tester        Process Control Instruments
                  DCV measure 200mV, 2V & 20V             calibrator to cover five additional
                  ranges
  Calibration     DCI measure 20mA
                                                              types of instrument in one
                  DCV/I source from calibrator               convenient & cost effective
Thermocouple      Optical Tacho Calibration                          workstation.
  Simulation
                  240 to 60,000 RPM

                                                            Use with Virtual Front Panel
   DCV/DCI                                                       software (supplied)
measurement                                                or ProCal Calibration software
                                                           Only for use with a 3000 Series
  function for                                                       Calibrator
process control
instrumentation
    Optical
 Tachometer
  Calibration
                         Voltage Range : 0 to 33V
                         Best 1 Year Accuracy : ±0.02% of                  Use with Virtual Front Panel
                         setting                                                software (supplied)
 Power Supply
                                                                          or ProCal Calibration software
 Calibration*            Current Range : 0 to 4A
                         Best 1 Year Accuracy : ±0.05% of
                                                                          Only for use with a 2000 Series
                         setting                                                    Calibrator

GENERAL SPECIFICATIONS


                         Double the time since last used up to 20 minutes maximum
Warm Up Time
                         RS232
Standard
Interface
                         GPIB / USB
Optional
Interface
                         IGUARD protection - detects & protects against applied voltages on the output
Output
Protection
                         Storage : -5°C to +60°C
Temperature              Operation : 0°C to +50°C
                         Calibration 18°C to 25°C
Performance
                         Specifications apply between 18°C and 25°C
Temperature              Outside this range an allowance of 0.18 x 1 year specification per °C should be
Coefficient              added
                         Operation <80% to 30°C, <70% to 40°C , <40% to 50°C
Relative                 Storage : <95% (non-condensing)

Humidity
                         Operation : 3000m (10,00ft) Maximum
Altitude                 Transit : 12000m (40,000ft) Maximum
                         The calibrator line inlet plug must be earthed
EMC & Safety             See D.O.C. for full details
                         100V
Low Analogue
Isolation
                         Line voltage selectable : 110V / 230V
Line Power               Line frequency : 50Hz to 60Hz
                         Line voltage variation : -6% +10%
                         28 Watts (Standby)
Power
Consumption
                         Calibrator Only : 43cm x 46cm x 19mm
Dimensions               Calibrator in shipping box : 58cm x 56cm x 37cm
                         Calibrator in soft carry case : 49cm x 50cm x 19cm
                         Calibrator in hard transit case : 55cm x 56cm x 26cm
                         Calibrator Only : 14.5kg
Weight                   Calibrator in shipping box : 17kg
                         Calibrator in soft carry case : 16Kg
                         Calibrator in hard transit case : 24kg


 CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
      Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de