Hier Klicken für das PDF Dokument!!!

                 INFRATEK 31 HAND HELD POWER ANALYZER




The INFRATEK MODEL 31 HAND HELD, BATTERY OPERATED, SINGLE AND THREE PHASE
POWER ANALYZERS are valuable engineering tools designed to offer precision measurement on all types of
signals including electronic drives.

 UNRIVALED FEATURES                                                            The Infratek 31 Power Analyzers are controlled by six menu soft
                                                                               keys Ml through M6, five cursor soft keys, and one power-on/off soft
 · Main           · 100 mA-50 A, DC to 300 kHz                                 key. The operating procedure, to configure the display, to configure
   Features       · 1 V-1000 V, DC to 300 kHz                                  the interface, to set scaling factors, and to select the input, the ranges,
                   · 0.3 W-150 kW, no CT's required                            and many more functions are self-explanatory. If you desire you can
                   · Suitable for frequency inverter drives                    store your personal instrument configuration in nonvolatile memory
                   · Measures rms-, power-, and harmonic values                and have the unit start-up with your personal configuration at power-
                   · Line-to-line voltage, torque, efficiency                  on.
                                                                               Efficient data processing makes simultaneous measurements in a
                  · Harmonics 1-63, IEC1000-3-2
                                                                               three phase system possible. Three rms line voltages, rms line-to-line
                   · Data logging for dynamic processes                        voltages, rms currents, power, power factor, minimum, maximum,
                   · Accuracy grades 0.1%, 0.2% and 0.4% (low cost)            harmonics of 3 currents and 3 voltages, and all derived quantities are
                                                                               determined from the same measurement interval (minimum 250ms).
 · Display        · Stores user defined configurations                         You have the choice to display phase values including wave forms
    Features      · Combined numeric and graphic fields                        or bar graphs, or you can display total and average values in the 3
                                                                               phase system. You can also display the values of all three phases
 · Processing     · Simultaneous measurements in 3 phase system                including their sums, and averages. The Model 31 Power Analyzer
    Power         · All data, including harmonics in real time                 includes a data logging mode to transfer data from a dynamic
                  · Efficient data transfer to computer                        process to a personal computer for further analysis. Also, current
                                                                               harmonics according to IEC1000-3-2 in a 3 phase system are
 · Interface      · RS-232 as standard                                         determined. By means of the Infratek operating software simple pass-
                  · Operating software under Windows 95, 98, NT                fail criteria are generated.

 LOW COST, EASY TO OPERATE
 The Infratek Model 31 hand-held Power Analyzers are sophisticated
 measuring instruments, single phase and three phase, designed for
 laboratory and field use. Unlike other instruments in this price range
 the Model 31 Power Analyzers are suitable for measurements on
 frequency inverter drives, light ballasts, and other electronically
 controlled loads. The wide frequency range, the high common mode
 rejection, and the fact that you can measure currents of up to 50A
 directly without having to use clamps guarantees excellent accuracy for
 all types of measurements.
                                                CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                                                    Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
 SPECIFICATIONS
Voltage             7 ranges: 1 V, 3 V, 10 V, 30 V, 100 V, 300 V, 1000 V
                    Frequency range: DC- 100 kHz / Coupling: AC/AC+DC                                                              1 Hz- 100 kHz / DC- 100 kHz
                    Crest Factor 3:1 at 50 % full scale / Common Mode 50 Hz/ 100 kHz                                               130dB/70dB
                    Built-in star point network 500 kOhm            | Maximum Input: Hi to Lo/Lo to case                           1000V/600V
                    Accuracy 23° ±3 °K; rms, rdg=reading                                                                           Accuracy Grades: k=l,k=2, k=4
                    1 Hz-1 kHz                     ±k(0.1 % rdg ±0.07 % range)
                    1 kHz- 10 kHz                  ±k(0.3 % rdg ±0.2 % range)
                    10 kHz- 100 kHz                ±k(0.05 %/ kHz rdg + 0.3 % range)
Current             10 ranges: 100mA, 300mA, 1,3, 10 A; 1,3, 10,30, 100 A.                                                         Clamp: 1 A- 1000 A
                    Frequency range DC- 100 kHz                   | Coupling: AC, AC+DC                                            1 Hz- 100 kHz / DC- 100 kHz
                    Crest Factor 3:1 at 50 % full scale                | Common Mode 50 Hz/100 kHz                                 150dB/100dB
                    3 A input: 3 A cont./l0 A 5 s; 50 A input: 40 A cont./50 A 20 s.                                               RI=30 mOhm/3 mOhm
                                                                                                                                   Accuracy Grades k=l,k=2, k=4
                    Accuracy 23° ±3 °K; rms, rdg=reading, rng=range
                                            50 A/Clamp Input                            3 A Input
                    1 Hz-500 Hz          ±k(0. 1 % rdg + 0. 1 % rng)                 +k(0. 1 % rdg + 0. 1 % rng)
                    500 Hz-2 kHz          ±k(0.6 % rdg + 0.6 % rng)                   ±k(0.8 % rdg + 0.8 % rng)
                    2 kHz- 10 kHz         ±k(0.8 % rdg + 0.8 % rng)                  ±k(2 % rdg + 1 % rng)*
                    10 kHz- 100 kHz       ±(0. 1 %/ kHzrdg + 0.8 % rng)*             ±(0. 1 %/ kHzrdg + 2 % mg)*
                    Lowest range multiply percentage figures by 2.                  *typical

Power               70 ranges corresponding to the products of voltage ranges times current ranges.                                DC- 100 kHz
                    Accuracy 23° ±3 °K; 50 A/Clamp Input, 3 A Input                                                                PF = O t o ± l
                    1 Hz- 1 kHz                  Add accuracy percentage figures of current and voltage input
                    1 kHz- 100 kHz                Add accuracy percentage figures of current and voltage input,
                                                  add +2 % ( 1 -PF)/ kHz of range

Computed Values     Reactive Power: Var = ±(VA2-W2)1/2; Apparent Power: VA=ArmsVrms; Power Factor: PF=W/VA; Crest                  Add accuracy percentage figures of
                    Factor: CF=Ap/Arms, Vp/Vrms; Maximum: Ap, Vp; Minimum: -Ap, -Vp; PtP: Maximum -Minimum;                        values involved in computation.
                    Impedance: Z=Vrms/Arms; Total harm.Dist., THDl=(Irms2-Ifund2) 1/2/Irms, THD2=(I22 + I32 + ... In2) 1/2/Irms.


Mechanical          Total Input Power, Nm/s; Output Power to load, Nrn/s; Torque at axis of rotating machine, Nm; Speed,           Synchronous machines only.
                    rpm; Efficiency: Eta = Output Power / Input Power
Values

Harmonic            Frequency range of fundamental 4 Hz-50 kHz                                                                     Harmonic 1-63
Analysis            Accuracy: harmonic current and voltage, same as rms current and rms voltage
                    Computed Values: harmonic power; harmonic phase angle (power factor); harmonic impedance.                      Add accuracy percentage figures of
                                                                                                                                   values involved in computation.
Frequency           2Hz-100kHz; A; V:±0.1 %
Integrator          Energy, Accuracy Wh, Vah: Basic accuracy of integrated quantity
Data Logging        Output values, speed, and duration is programmable; maximum speed: 10 values from a 3-phase system in          Range of signal frequency 5 Hz to
                    150 ms.                                                                                                        1 kHz
Current             Current harmonics in a 3-phase system are determined according to IEC1000-3-2
Harmonics
Display             Display: Blue liquid crystal graphic display with EL backlight                                                 58xl08mm; 128x240 pixels
                    Charger: AC, 50-60 Hz, 1 15/230 V, 0.2 AF, 10 VA; sealed Lead gel battery, 6 hours
Power/Battery       Dielectric Strength: Current inputs to case; Voltage inputs to case; Line input to case
Dielectric                                                                                                                         3 kV; 2 kV; 3 kV 50 Hz/1 min
                    Dimensions: H x W x D                                                                                          75xl60x245mm; 1.5kg
Strength                                                                                                                           I
Dimensions
Interface           RS-232 as standard incl. Windows Software
Accessories         Current clamps 0-200 A/0-400 A




     CompuMess Elektronik GmbH
        Lise-Meitner-Str.1, 85716 Unterschleissheim
           Tel 089-321501-0 Fax 089-321501-11
    http://www.compumess.de oder http://www.netzteile.de