Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
INTEGRATED ELECTRICAL EQUIPMENT CALIBRATION SOLUTION
Standard Interfaces RS232
Optional Interfaces GPIB (IEEE-488) : USB (Universal Serial Bus)
Temperature Performance Storage : -5°C to +60°C
Operation : 0°C to +50°C
Relative Humidity Operation : <80% to 30°C, <70% to 40°C, <40% to 50°C
Storage : <95%, non-condensing
Altitude Operation : 3000m (10,000ft) Maximum
Transit : 12000m (40,000ft) Maximum
EMC & Safety The calibrator line input plug must be earthed
See D.O.C for full details
Line Power Line Voltage Selectable : 110V / 230V
Line Frequency : 50Hz to 60Hz
Line Voltage Variation : -6% +10%
Power Consumption 28 Watts (Standby) 200 Watts (Maximum)
Low Analogue Isolation 100V
Connections Voltage / 2 Wire Resistance 1x Black : 1x White 4mm Safety sockets
Low Current (<=2A) 1x Black : 1x Red 4mm Safety sockets
High current (>2A) 1x Blue : 1x Yellow 4mm Safety sockets
Earth Connection 1x Green 4mm Safety Socket
Oscilloscope Functions 1x BNC terminal
Feature (Ext. Pod) 1x Female 'D' type socket
RS232 Interface 1x Female 'D' type socket
RS232 Settings Baud Rate 9600
Parity None
Data Bits 8
Stop Bits 1
Display Information Type Backlit Black on white film STN type
Viewing Area 124.3mm * 34mm
Resolution 256 * 94 dots
Backlight Type Cold fluorescent lamp
Brightness 70 to 90 cd/m2
Indicators Voltage / Current / High Current ed LED (between terminals)
R
Negative to ground Green LED (left of Earth terminal)
Oscilloscope Green LED (right of BNC Connector)
Feature Connector (Ext. Pod) Green LED (right of 'D' type connector)
Keyboard Membrane type with tactile feedback
Fuses Mains Inlet 3A A/S (240 Volt)
5A A/S (110 Volt operation)
Isolation Outputs are opto-isolated from mains earth and the RS-232 interface
Maximum common mode voltage between earth and the
Dimensions & Weights Calibrator Only
Calibrator in Shipping Box
Calibrator in Soft Carry Case
Calibrator in Hard Transit case
Warranty Period 3 Years (Parts & Labour)
Recommended Service Interval 1 Year
Supplied Connections 1x Serial Interface Connection
EXTENDED SPECIFICATIONS
TRANSMILLE LTD., UNIT 4 SELECT BUSINESS CENTRE, LODGE ROAD, STAPLEHURST, KENT. TN12 0QW. UK.
www.transmille.com : sales@transmille.com : Tel : +44 (0) 1580 890700 : Fax : +44 (0) 1580 890711
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
Warm Up Time Double the time since last used up to 20 minutes maximum
Standard Interfaces RS232
Optional Interfaces USB (Universal Serial Bus)
Temperature Performance Storage : -5°C to +60°C
Operation : 0°C to +50°C
Relative Humidity Operation : <80% to 30°C, <70% to 40°C, <40% to 50°C
Storage : <95%, non-condensing
Altitude Operation : 3000m (10,000ft) Maximum
Transit : 12000m (40,000ft) Maximum
EMC & Safety The calibrator line input plug must be earthed
See D.O.C for full details
Line Power Line Voltage Selectable : 110V / 230V
Line Frequency : 50Hz to 60Hz
Line Voltage Variation : -6% +10%
Power Consumption 28 Watts
Connections PAT Testing Connection 1x IEC Plug
LOOP & RCD Testing Connection1x UK / European / Australian type socket
Insulation Tester Connection 1xBlack : 1xRed 4mm Low Thermal Sockets
PAT Ground Connection 1x 4mm terminal post
RS232 Interface 1x Female 'D' type socket
RS232 Settings Baud Rate 9600
Parity None
Data Bits 8
Stop Bits 1
Display Information Type Backlit Black on white film STN type
Viewing Area 124.3mm * 34mm
Resolution 256 * 94 dots
Backlight Type Cold fluorescent lamp
Brightness 70 to 90 cd/m2
Indicators PAT Testing Connection Red LED above plug
LOOP & RCD Testing ConnectionRed LED above socket
Insulation Tester Connection Red LED above terminals
Keyboard Membrane type with tactile feedback
Fuses Loop (Live) 5A Anti-Surge
Mains 2A
Loop (Neutral) 5A Anti-Surge
RCD 2A
ACV 100mA
PAT 1A
Insulation Resistance 100mA
Isolation Outputs are opto-isolated from mains earth and the RS-232 interface
Maximum common mode voltage between earth and the
low terminals 30 Volts ac/dc.
Dimensions & Weights Calibrator Only 45cm x 44cm x 14cm : 10kgs
Warranty Period 1 Year
Recommended Service Interval 1 Year
Supplied Connections 1x Serial Interface Connection
1x 1m PAT Test Lead
Mounting Kit (optional) 3U rack mount kit
Case Colour Cream (RAL9002)
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
General Specifications : V1.00 Page 2
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
Continuity Resistance
Standard Accuracy - Relative to Calibration Standards Specifications
Range Type Resolution Accuracy
% ± m
0.1 to 20 Continuously
Variable 10m 1 ± 25
100 Fixed 10m 1 ± 25
1k Fixed 10m 1 ± 25
High Accuracy (option) - Relative to Calibration Standards Specifications
Range Type Resolution Accuracy
% ± m
0.1 to 20 Continuously
Variable 10m 0.25 ± 25
100 Fixed 10m 0.25 ± 25
1k Fixed 10m 0.25 ± 25
Continuity Current Measurement
Accuracy - Relative to Calibration Standards Specifications
Range Load Accuracy
% ± Counts
0 to 320mA 1 1.3 ± 6
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
Continuity Specifications : V1.00 Page 3
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
Insulation Resistance
Standard Accuracy - Relative to Calibration Standards Specifications
Range Type Resolution Maximum Accuracy
Voltage/Power1 %
0 to 5M Continuously
Variable 10k 1.1kV or 1 Watt 0.3
5M to 2G Continuously
Variable 10k 1.1kV or 1 Watt 3
High Accuracy (option) - Relative to Calibration Standards Specifications
Range Type Resolution Maximum Accuracy
Voltage/Power1 %
0 to 5M Continuously
Variable 10k 1.1kV or 1 Watt 0.1
5M to 2G Continuously
Variable 10k 1.1kV or 1 Watt 1
Note 1 : A 5kV option is available for Insulation Testers incorporating Active Guard .
10G Range (Option) - Relative to Calibration Standards Specifications
Range Type Resolution Maximum Accuracy
Voltage/Power1 %
2G to 10G Continuously
(Option) Variable 10k 1.1kV or 1 Watt 5
Note : Can be fitted to Standard or High accuracy models
Insulation Test Voltage Measurement
Accuracy - Relative to Calibration Standards Specifications
Ranges Current Resolution Accuracy
Measurement % ± Counts
50V · 100V · 250V · 500V · 1kV 0.5mA · 1mA 0.1V 1 ± 8
AC Voltage Output
Accuracy - Relative to Calibration Standards Specifications
Ranges Accuracy
% ± Counts
100V · 200V · 230V · 300V · 400V 0.2 ± 1
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
Insulation Specifications : V1.00 Page 4
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
RCD Time
Accuracy - Relative to Calibration Standards Specifications
Range Resolution Accuracy
20ms to 5s 10ms ± 0.7ms
RCD Current
Accuracy - Relative to Calibration Standards Specifications
Range Resolution Accuracy
to 200mA to 3A Setting % ± Counts
3mA to 3A 0.01mA 0.1mA <190ms 5 ± 20
up to 5s 1.2 ± 6
Additional Features
Timing Modes Immediate · Zero Crossing
Current Modes ½I · I ·2I ·5I
Display Modes 0° and 180° of Phase
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
RCD Specifications : V1.00 Page 5
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
Loop Resistance
Standard Accuracy - Relative to Calibration Standards Specifications
Nominal Resolution Accuracy
Resistance Values % ± m
0.05 0.1m 0.5 ± 4
0.1 0.1m 0.5 ± 4
0.22 0.1m 0.5 ± 4
0.33 0.1m 0.5 ± 4
0.5 0.1m 0.5 ± 4
1 0.1m 0.5 ± 4
5 0.1m 0.5 ± 4
10 0.1m 0.5 ± 4
100 1m 0.5 ± 4
1k 1m 0.5 ± 4
Auto Loop (Option)
The auto loop function automatically corrects for supplied loop impedance.
Maximum Correction 2
Resolution 100u
Accuracy ±18m
Measurement Current 4A
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
LOOP Specifications : V1.00 Page 6
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
PAT Earth Bond Resistance
Standard Accuracy - Relative to Calibration Standards Specifications
Nominal Resolution Accuracy
Resistance Values % ± m
0.05 0.1m 0.5 ± 4
0.1 0.1m 0.5 ± 4
0.22 0.1m 0.5 ± 4
0.33 0.1m 0.5 ± 4
0.5 0.1m 0.5 ± 4
1 0.1m 0.5 ± 4
5 0.1m 0.5 ± 4
10 0.1m 0.5 ± 4
100 1m 0.5 ± 4
1k 1m 0.5 ± 4
PAT Earth Bond Current Measurement
Range Resolution Accuracy
% ± Counts
100mA 1mA 1.5 ± 6
10A 10mA 1.5 ± 6
30A 10mA 1.5 ± 6
PAT Insulation Resistance
Standard Accuracy - Relative to Calibration Standards Specifications
Range Type Resolution Maximum Accuracy
Voltage/Power1 %
0 to 5M Continuously
Variable 10k 1.1kV or 1 Watt 0.3
5M to 2G Continuously
Variable 10k 1.1kV or 1 Watt 3
High Accuracy (Option) - Relative to Calibration Standards Specifications
Range Type Resolution Maximum Accuracy
Voltage/Power1 %
0 to 5M Continuously
Variable 10k 1.1kV or 1 Watt 0.1
5M to 2G Continuously
Variable 10k 1.1kV or 1 Watt 1
PAT Leakage Current
Range Resolution Accuracy
% ± Counts
240uA 1uA 1.5 ± 2
1mA 1.5 ± 2
3mA 1uA 1.5 ± 2
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
PAT Specifications : V1.00 Page 7
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
PAT Leakage Test Voltage
Range Resolution Accuracy
(RMS) % ± Counts
100V to 300V 0.1V 1.5 ± 9
PAT Load Testing
Range Accuracy
Ohms ± %
Short Circuit - - -
Open Circuit - - -
0.13kW 440 ± 5
PAT Flash Voltage Measurement
Class Range Resolution Load Accuracy
Resistance % ± Counts
Class 1 (1.5kV) 1kV to 1.8kV 1V 600k 4 ± 10
(2.5mA@1.5kV)
Class 2 (3kV) 2kV to 3.6kV 1V 1.2M 4 ± 10
(2.5mA@3kV)
PAT Flash Current Measurement
Range Resolution Accuracy
%
1mA to 3mA 10uA 5
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
PAT Specifications Cont'd : V1.00 Page 8
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
3200 ELECTRICAL TEST EQUIPMENT CALIBRATOR
Line Voltage Measurement
Range Resolution Accuracy
200V to 260V 0.1V 0.8% ± 6 Counts
Due to continuous development specifications may be subject to change.
3200 Extended Specifications
Mains Voltage Measure : V1.00 Page 9