Series 9336
High Value Resistance Standards
Very High Stability High Value Calibration Laboratory Standards
G uildline 9336 series of Resistance Standards are designed
as very high stability calibration laboratory standards for
accurate resistance calibration in air, between 10 Megohms
and 100 Gigaohms.
These resistors are also suitable for use as calibration standards
for the Guildline Teraohmmeter, Model 6500A and the
calibration of other high ohms measuring instruments.
They can be used as working standards or reliable, ruggedized,
transportable transfer standards. They are extremely useful for
the calibration of resistance ranges of multi-function
calibrators and high accuracy digital multimeters, as well as
being used in more classical standards and calibration
laboratory applications where the need for high accuracy high
ohms values are required.
The resistor elements are securely mounted to the inside of a
rugged hermetically sealed, shielded, aluminum enclosure. A
pair of input/output Type N connectors provide the
termination for the standard.
9336 Features The 9336 Series Precision Resistance Standards
can be used for laboratory or portable applications
> Very High Stability requiring high accuracy at large resistance values.
> Range 10M to 100G
The "SOURCE" connector connects to the power supply of the
> Operating Range 18 ºC to 28 ºC measurement system, while the "OUTPUT" connector connects
to the measurement detector. If necessary, the ambient
> Hermetically sealed temperature of the enclosure may be monitored and a
> Report of Calibration traceable to NIST or correction factor applied to the value of the resistance.
INMS/NRCC included
> Compact & Ruggedized
> Low Temperature Coefficient
> Rated to 1000 V
> Low Power coefficient
> Suitable for calibration of Teraohmmeters,
Meggers, long scale DMMs, & Ohmmeters
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
9336 Resistance Standards
9336 Series Specifications
Model Nominal Nominal Calibration Stability Temp. Coeff. Voltage Coeff.
Resistance Initial Uncertainty 12 Months 18-28 0C (ppm/volt)
Value Tolerance @23 0C ±1 0C (± ppm) (ppm / 0C) (Note 3)
(Ohms) (± ppm) (Note1) (± ppm) (Note 2)
9336-10M 10M 25 15 10 <5 0.1
9336-100M 100M 50 25 25 <5 0.5
9336-1G 1G 100 80 35 <6 0.5
9336-10G 10G 200 100 100 <25 1
9336-100G 100G 500 500 200 <250 1
Note 1: Nominal Initial Tolerance is the maximum variation of resistance mean value as adjusted initially at the point of sale.
Note 2: Calibrated at 23 0C, referred to the unit of resistance as maintained by the National Research Council of Canada (NRCC) or the
National Institute of Standards and Technology (NIST) and expressed as a total uncertainty with a coverage factor of k =2.
A traceable calibration report stating the measured value and uncertainty is provided with each resistor.
Note 3: Maximum Voltage Rating: 1000 volts
Note 4: Special Values available on request
Ordering Information
9336-ohmic value Resistance Standard
TM 9336 Technical Manual (included)
Cert of Calibration (included)
Report of Calibration (included)
General Specifications Guildline is distributed by:
Environment: Operating 18 0 C to 28 0 C
< 50% RH non-condensing
Storage -20 0 C to 60 0 C,
15 to 80% RH CompuMess Elektronik GmbH
Dimensions: H 82 mm (3.2 in) Lise-Meitner-Str.1, 85716 Unterschleissheim
Tel 089-321501-0 Fax 089-321501-11
W 124 mm (4.9 in) http://www.compumess.de oder http://www.netzteile.de
D 79 mm (3.1 in)
Weight: 0.63 kg (1.4 lbs)
Guildline Instruments Limited
P.O. Box 99
21 Gilroy Street
Smiths Falls, Ontario Canada K7A 4S9
Phone: (613) 283-3000
FAX: (613) 283-6082
9336 V1.0 Copyright © 2002.08.08 Guildline Instruments Limited. All rights reserved. Subject to change without notice. Web: www.guildline.ca