                                                                                                                 146-1968


                            MORE WATTS PER VERTICAL INCH


                                Load up on the
                                     NEW
   SERIES EL ELECTRONIC LOADS
                                      from Kepco




Model EL 2K-200-200
Single Channel Unit


                                                                                                     Model EL 1K-200-100D
                        CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
                                                                                                        Dual Channel Unit
                        Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
                        info@compumess.de · www.compumess.de · www.netzteile.de




                      · 1KW - 5KW Modules · 4U Height · Parallel Operation
   · Up to 600 Volts DC · Up to 600 Amperes/Module · Constant E, I, P, I/E, E/I · Air Cooled
         · Local, Remote Analog Control and USB Standard · GPIB and Ethernet Optional


                                                  AN ISO 9001 COMPANY
THE NEW KEPCO SERIES EL ELECTRONIC LOADS
The EL Series is a line of modular air-                                   INDIVIDUAL LOAD                                    SYSTEM INTEGRATION
cooled, DC electronic loads used to test                                  CHARACTERISTICS                                    IS SIMPLE
power sources such as batteries, power                                    · Base Unit - Includes front panel                 · Up to 60KW into one 8 ft. rack
supplies, generators, chargers, fuel cells, etc.                            controls and indicators for Amps,                · Control up to 15 slaves with
Power ratings start at 1KW; standard models                                 Volts and KW                                       one master (75KW)
have maximum test capabilities of 50, 200,
                                                                          · Computer operated LabVIEWTM                      · Computer controlled and/or front
400 and 600 volts. High current and
                                                                            capable                                            panel operation with display for
power operation are achieved via parallel
connection in a master/slave configuration.                               · Constant current, voltage, power,                  voltage, current, power and status
Individual modules are capable of up to 5KW                                 resistance, conductance and                      · Simple operation - no deep menus
and 600 amperes (see chart).                                                short circuit modes                                to contend with
The EL Series offers unique Functional                                    · Readback voltage, current, power
Modularity. Functional Modularity allows                                    and load status                                  RELIABLE AND RUGGED
the load to be retrofitted to increase load                                                                                  · High power semiconductors are
                                                                          · USB and RS-232 are standard
power/current as required for undefined                                                                                        heat sinked for cool operation
future applications. Purchase the capability                              · GPIB and Ethernet Control via
                                                                                                                             · Very low internal resistance
needed today with assurance of future                                       single optional interface card
                                                                                                                               ensures fast and clean switching
power/current upgrades. The result is an                                  · Comprehensive SCPI command set                     operation
Electronic Load maximizing the most
                                                                          · Monitor load current via computer                · Power devices individually fused
important specification:
                                                                            or analog output                                   for complete protection
                     VALUE = $/Watt
                                                                          · Use unique test patterns from an                 · Internal thermal sensors ensure
When the requirements increase, the load                                    Arbitrary Waveform Generator via                   uniform power distribution for
may be upgraded.                                                            the external analog input                          long reliability
All EL loads operate as master or slave and                               · Remote voltage sense eliminates                  · Isolated test bus allows a
come with or without front panel control and                                error caused by voltage drop in                    wide range of source (DUT)
readouts. Up to 15 additional loads can be                                  test leads                                         configurations
driven with one acting as a master.
                                                                                                                               LabVIEW is a trademark of National Instruments Corp.




EXAMPLE OF SAFE OPERATING AREA FOR EL 5K-400-420
                                                              12.5 16.6
   1,000.00                                                                                                                                           The load may
                                                                                                                                                      be operated at
        400.00                                                                                As current increases, voltage must be
                                                                                             lower to keep the power dissipation at
                                                                                                                                                      any given input
        300.00
                                                                                             or below the upper limit of 5000 Watts.                  condition that
                         Operation at 400 Volts                                                                                                       is within the
        100.00        valid until 12.5 Amps, which                                                                                                    red line boundary.
                      is 5000 Watts, the maximum
                      power dissipation capability.
                                                                                                                                                      For example:
                                                                                                                                                      If a 300 Volt source
                                                                     This line represents the maximum current                                         is to be tested, select
VOLTS




         10.00                                                       capability (400 Amperes). The line stops at                                      the 300 Volt line (blue)
                                                                      5.05 Volts because the lowest resistance
                                                                             capability has been reached.                                             and note the range
                                                                                                                                                      of current allowable
                                                                                                                                                      along that line.
          1.00                  This line represents the                                                                                              In this example,
                            minimum resistance of the load.
                                                                                                                                                      the load may be
                                                                                                                                                      operated at currents
                                                                                                                                                      from near zero to
                                                                                                                                                      a maximum of 16.6
          0.10                                                                                                                                        Amperes @300V.
                 1                                       10                                       100                                      1000
                                                                          AMPERES

                      Refer to www.kepcopower.com/el-opcurve.pdf for operating curves of all EL models.

©2010, Kepco, Inc.                                                           CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
                                KEPCO, INC. · 131-38 Sanford Avenue · Flushing, NY 11355 USA · Tel: (718) 461-7000 · Fax: (718) 767-1102
Specifications subject to
change without notice.                                  Email: hq@kepcopower.com ·(089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
                                                                             Telefon www.kepcopower.com
                                                                                    info@compumess.de · www.compumess.de · www.netzteile.de
Model EL 2K-200-200                                                                                                                             Model EL 1K-200-100D


  SERIES EL MODEL TABLE

  MODEL NUMBER (1)                           OPERATING LIMITS                                       OVERLOAD PROTECTION (2)                        MINIMUM ON
                                                                                                                                                   RESISTANCE
                            RATED POWER       RATED VOLTAGE      RATED CURRENT          POWER (3)          VOLTAGE (4)        CURRENT (5)          ALL MODES
                                Watts             Volts            ALL MODES             Watts               Volts              Amps                  Ohms
                                                                     Amps

  EL 1K-50-100                 1000                50                 100               1050                 52.5               105                  0.008

  EL 1K-200-100                1000                200                100               1050                 210                105                  0.014

  EL 1K-400-70                 1000                400                70                1050                 420                73.5                 0.046

  EL 1K-600-30                 1000                600                30                1050                 630                31.5                 0.017

  EL 1K-50-100D                100 0 (6)           50 (6)             100 (6)           1050 (6)             52.5(6)            105 (6)              0.008 (6)

  EL 1K-200-100D               1000 (6)            200 (6)            100 (6)           1050 (6)             210 (6)            105 (6)              0.014 (6)

  EL 1K-400-70D                1000 (6)            400 (6)            70 (6)            1050 (6)             420 (6)            73.5 (6)             0.046 (6)

  EL 1K-600-30D                1000 (6)            600 (6)            30 (6)            1050 (6)             630 (6)            31.5 (6)             0.017 (6)

  EL 2K-50-200                 2000                50                 200               2100                 52.5               210                  0.004

  EL 2K-200-200                2000                200                200               2100                 210                210                  0.007

  EL 2K-400-140                2000                400                140               2100                 420                147                  0.023

  EL 2K-600-60                 2000                600                60                2100                 630                 63                  0.083

  EL 3K-50-300                 3000                50                 300               3150                 52.5               315                  0.005

  EL 3K-200-300                3000                200                300               3150                 210                315                  0.005

  EL 3K-400-210                3000                400                210               3150                 420               220.5                 0.015

  EL 3K-600-90                 3000                600                90                3150                 630                94.5                 0.056

  EL 4K-50-500                 4000                50                 500                525                 52.5               525                  0.002

  EL 4K-200-500                4000                200                500               2100                 210                525                  0.003

  EL 4K-400-350                4000                400                350               4200                 420               367.5                 0.009

  EL 4K-600-150                4000                600                150               4200                 630               157.5                 0.333

  EL 5K-50-600                 5000                50                 600               5250                 52.5               630                  0.002

  EL 5K-200-600                5000                200                600               5250                 210                630                  0.002

  EL 5K-400-420                5000                400                420               5250                 420                441                  0.008

  EL 5K-600-200                5000                600                200               5250                 630                210                  0.028

(1) For GPIB and Ethernet Interface, add suffix "G" to the model number.
(2) Temperature protection: Per FET safe junction temperature.
(3) Maximum operating power and/or current may be reduced if operating conditions (ambient temperature, available air flow, etc.) cause internal load temperature to
    exceed safe operating conditions for FETs (TEMP FAULT).
(4) Maximum operating power and/or current may be reduced when ambient temperature exceeds 25°C depending on air flow, duty cycle of load operation and
    other situational specific conditions.
(5) Individually fused FETs allow continued operation with device failure.
(6) Per channel.




                     KEPCO, INC. · 131-38 Sanford Avenue · Flushing,Elektronik USA · Tel: Lise-Meitner-Str. Fax: D-85716 Unterschleißheim
                                                     CompuMess NY 11355 GmbH · (718) 461-7000 · 1 · (718) 767-1102
                                             Email: hq@kepcopower.com · www.kepcopower.com 01 - 11
                                                     Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15
                                                             info@compumess.de · www.compumess.de · www.netzteile.de
 SERIES EL SPECIFICATIONS

 SPECIFICATION                                                      RATING/DESCRIPTION (1)                                 CONDITION

 MODE CHARACTERISTICS
 Linearity vs. Programming Command                                                                                    5-100% of Full Scale
   Constant Current                                                       ±0.25%
   Constant Power                                                           ±2%
   Constant Voltage                                                       ±0.25%
   Constant Resistance                                                      ±1%
   Constant Conductance                                                     ±1%

 Regulation
   Constant Current                                                       ±0.25%
   Constant Power                                                           ±1%

   Constant Voltage                                                       ±0.25%

   Constant Resistance                                                      ±1%
   Constant Conductance                                                     ±1%

 Resolution (Via Computer Control)                                         14 Bits

 Current Readback               Computer Accuracy                         ±0.25%
 (Current Mode)
                               Computer Resolution                        ±15 Bits

 Parameter Readback                        Accuracy                       ±0.25%
 (Current and Voltage)
                                          Resolution                      ±15 Bits

 CONTROL CHARACTERISTICS
 USB Interface                                                  Uses IEEE 488.2 and SCPI                     Requires no cost downloadable driver
                                                                 commands and queries

 RS 232 Interface                                               Uses IEEE 488.2 and SCPI                      Baud rate: 38400       Parity: None
                                                                 commands and queries                     Data Bits: 8    Stop Bits: 1      Echo: OFF

 GPIB Interface                                                 Uses IEEE 488.2 and SCPI                Optional card provides access to both interfaces.
 Ethernet Interface                                              commands and queries                           Add suffix "G" to model number
                                                                                                                         when ordering.

 GENERAL CHARACTERISTICS
 a-c Line Power Input                                           120V a-c ±10%, 50-60 Hz or                          120V is factory default,
                                                                 240V a-c ±10%, 50-60 Hz                      240V requires internal configuration

 Operating Temperature                                                  0°C to 40°C

 Dimensions (Load)                                                  19"W x 7"H x24.5"D

 Weight (Load)                                                             88 lbs.

 Storage Requirements                                           -20 to +60°C, 15 to 80% RH                      Relatively dust free environment

(1) Specifications measured @23°C ambient.
NOTES:
1. Regulation specified after 15 minutes of operation at set power level.                                                        PENDING

2. Individually fused FETs allow continued operation with device failure.
3. Maximum operating power and current may be reduced if operating conditions (ambient temperature, available air flow, etc.)
   cause internal load temperature to exceed safe operating conditions for FETs (TEMP FAULT).
4. Maximum operating power and/or current may be reduced when ambient temperature exceeds 25°C depending on air flow,
   duty cycle of load operation and other conditions.




                                                                 APPLICATIONS
                  AUTOMOTIVE - LIFE-CYCLE TESTING OF HYBRID CONTROL MODULES · FUEL CELL TEST
                 WINDMILL CHARACTERIZATION AND PERFORMANCE TESTING · WEAPONS BATTERY TESTS
                                  HELICOPTER POWER GENERATOR FLIGHT LINE TEST
                            POWER SUPPLY TESTING · HIGH PERFORMANCE BATTERY TEST



                 KEPCO, INC. · 131-38 Sanford Avenue · Flushing, NY 11355 USA · · Lise-Meitner-Str. · Fax: (718) 767-1102
                                                CompuMess Elektronik GmbH Tel: (718) 461-7000 1 · D-85716 Unterschleißheim
                                         Email: Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
                                                hq@kepcopower.com · www.kepcopower.com
                                                       info@compumess.de · www.compumess.de · www.netzteile.de
CONFIGURING AND OPERATING YOUR EL ELECTRONIC LOAD                                                          FRONT PANEL
The front and rear panels below show a typical single channel model                                        CONTROLS AND INDICATORS
                                                                                                           1. Power Switch
                                                                                                           2. Current Display (Autorange)
                                                                                                           3. Voltage Display (Autorange)
                      7       8   9                                                  1                     4. Power Display (Autorange)
                                                                                                           5. Multifunction Control Knob
                                                                                                               Press to Select Mode
                                                                                                               Turn for Set Point
 2                                                                                                         6. LOAD - Green for Standby
                                                                                                                       Amber for Engaged
 3                                                                                                                     Press to Engage/Disengage
                                                                                                                       Flashing Red for Fault
 4
                                                                                                           7. LED Constant Current Mode (CI)
 5                                                                                                         8. LED Constant Resistance Mode (CR)
                                                                                                               Blinks for Constant Conductance
                                                                                                                 Mode (CS)
                                                                                                           9. LED Constant Voltage Mode (CV)
                                                                                Front Panel                10. LED Constant Power Mode (CP)
              6       12 11 10                                                                             11. LED LAN (Lit When LAN Connected)
                                                                                                           12. LED Local

                                                                                                           All Mode LEDs Blinking ON - Short
                                                                                                           All Mode LEDs Blinking OFF - Mode Off
                  1       2               3           4        5 6    7

                                                                                                           REAR PANEL
                                                                                                           LAYOUT AND CONNECTIONS
                                                                                                 8
                                                                                                           1. a-c Input
                                                                                                           2. Fuse
                                                                                                 9         3. Laminated Copper Buss
                                                                                                10             LOAD Connection
                                                                                                11         4. Remote Sense ­, Rear Terminal ­
                                                                                                           5. Remote Sense +, Rear Terminal +
                                                                                                12         6. External Analog Control
                                                                                                13         7. Current Monitor Analog Output
                                                                                                           8. Ethernet Option (Not Shown)
                                                                                                           9. Master (Out to Slave)
                                                                                                           10. Slave Input
      Rear Panel
                                                                                                           11. Slave Output
                                                                                14
                                                                                                           12. RS-232 Interface
                                                                                                           13. USB Port
                                                                                                           14. GPIB Option (Not Shown)




                                                                          UPGRADABLE AND DYNAMICALLY CONFIGURABLE
                                                                          · Expand capability by adding Power FET modules
                                                                          · Upgradable due to load's modular nature
                                                                          · Interface enables field retrofits of firmware upgrades
                                                                            ensuring system is up-to-date
                                                                          · Ethernet option facilitates remote access


                                                                          Model EL 2K-200-200/Single Channel Unit
                                                                          This unit shows two FET modules in place. There is room for three
                                                                          more modules to be added expanding the capability of the load.




                  KEPCO, INC. · 131-38 Sanford Avenue ·CompuMess11355 USA · Tel: (718)Lise-Meitner-Str.(718)D-85716 Unterschleißheim
                                                        Flushing, NY Elektronik GmbH · 461-7000 · Fax: 1 · 767-1102
                                          Email: hq@kepcopower.com · 15 01 - 0 · Telefax (089)
                                                       Telefon (089) 32www.kepcopower.com32 15 01 - 11
                                                      info@compumess.de · www.compumess.de · www.netzteile.de
 TABLE OF SAMPLE SOFTWARE COMMANDS FROM OVER 1000 IN YOUR SYSTEM

 FUNCTION                                                         PURPOSE                                              COMMAND EXAMPLE

  OPERATION MANAGEMENT
  Mode                                                  Selection of operating mode                     MODE CURRent, MODE VOLTage, MODE POWer,
                                                                                                          MODE RESistance, MODE CONDuctance,
                                                                                                                MODE SHORt, MODE OFF

  Load ON/Load OFF                             Engage/Disengage load with device under test                       INPut 1=ON, INPut 0=OFF

  Set Parameter Value                    Set current, voltage, power, resistance, conductance values             CURRent xxxx, POWer xxxx

  Measure                                             Measure Voltage, Current, Power                      MEASure:CURRent?, MEASure:VOLTage?

  Protection                              Allows setting maximum parameter values to protect the               VOLTage:PROTection:UNDer xxxx
                                          device under test as needed as well as minimum voltage                VOLTage:PROTection:over xxxx

  Damping Selection                                12 different damping settings to allow                      SYSTem:PFModule:DAMPing x,
                                                          optimizing the user's test                                  SYSTem:DAMPing x
                                                                                                          Where x is a number from 0 to 5 (min - max)

  Front Panel Lock                                Locks or unlocks the front panel controls.                   LOCK where state is OFF;
                                              Allows three settings for Off, Lock or Emergency                    EMERgency; or LOCKed
                                                (Emergency allows front panel LOAD button
                                                           to disengage the load)

 SYSTEM MANAGEMENT
 Serial Number                                            Display EL serial number                                     SYSTem:SERN?

 Version                                          Display Master control board information                           SYSTem:VERSion?

 Power FET Modules Installed                       List power FET modules installed in EL                         SYSTem:PFModule:LIST?

 Display Power FET Status                    Display FET temperatures, currents and fuse status                  SYSTem:PFModule:STATus?

 External Analog Control On/Off                      Turn external analog input ON/OFF                            SYSTem:EXTernal 

NOTE: System responds to short forms of all commands and queries. See instruction manual for details.



                                                                                3/4 View                           A 60KW
                                                                                Single                       configuration.
                                                                                Channel Unit                Consists of one
                                                                                                           5kW master and
                                                                                                            11 5kW slaves.




                                                                                3/4 View
                                                                                Dual
                                                                                Channel Unit




                                                      For more information visit
                                                     www.kepcopower.com/el.htm