Hier Klicken für das PDF Dokument!!!

                                                     An                           Company




                                                                                             Lx Series
                                                                                        Power Systems
                                                                          3-18 kVA Programmable
                                                                         AC Power Source / Analyzer



                                          Integrated System
                                          The Lx Series represents a modern AC power source that addresses increasing demands
                                          on test equipment to perform more functions at a lower cost. By combining a flexible AC
  Backward Compatible                     power source with a harmonic power analyzer, the Lx Series systems are capable of handling
     CompatiblewithHP6834B&iLSeries       applications that would traditionally have required multiple instruments.
     ACSourcesFunction&buscompatible      The sleek integrated approach of the Lx Series avoids the cable clutter that is commonly
     withtheAgilentHP6834B&California     found in AC test setups. All connections are made internally and the need for external digital
     InstrumentsiLSeries                  multimeters, power harmonics analyzer and current shunts is completely eliminated.
  Three phase and Single phase modes      Using a state of the art Digital Signal Processor in conjunction with precision A/D converters,
     Ideallysuitedforavionicsanddefense   the Lx Series provides measurement accuracy and resolution similar to dedicated power
     applications                         analyzers. Since many components in the Lx Series are shared between the AC source and
                                          the power analyzer, the total cost of the integrated system is less than the typical cost of a
  3 KVA to 18 KVA Power Levels            multiple unit system.
     Matchpowersourceandcostto
     applicationrequirements              Easy To Use Controls
  Arbitrary Waveform Generator            The Lx Series is completely microprocessor controlled and can be operated from a simple
     Testproductsforharmonics             front panel keypad. An analog style control knob allows output voltage and frequency to be
     susceptibility                       slewed up or down dynamically. The control employs a dynamic rate change algorithm that
  Built-in Power Analyzer                 combines the benefits of precise control over small parameter changes with quick sweeps
                                          through the entire range. A decimal keypad makes direct parameter entries fast and simple.
     Performsvoltageandloadcurrent
     harmonicanalysisandwaveform
                                          Applications
     acquisition
                                          With precise output regulation and accuracy, high load drive current, multi or single phase
  Standard IEEE-488, USB & RS232          mode and built-in power analyzer measurement capabilities, Lx Series AC source/analyzers
     RemotecontrolinterfaceforATE         address many application areas for AC power testing. Additional features, like line arbitrary
     systemintegrationincluded            waveform generation and available RTCA/DO160, MIL-STD704, Airbus or Boeing test
                                          standards, make the Lx Series a good choice for avionics or defense applications. All Lx
                                          Series AC sources are equipped with IEEE-488 (GPIB), USB and RS232C remote control
                                          interfaces and support SCPI command language programming. An ethernet interface
                                          option is available.

                                          HP6834B and CI iL Series Compatibility
                                          The Lx Series offers functional and bus compatibility with the Agilent HP6834B AC power
                                          sources as well as the CI iL Series AC power sources and may be used in existing test systems
                                          without the need to modify program code.




www.california-instruments.com
                              An             Company



  Lx Series - AC Transient Generation
                                                 Standard Waveforms
                                                 The Lx Series provides three standard waveforms that are always available for output.
                                                 The standard waveforms are:
                                                 ·     Sinewave for normal AC applications.
                                                 ·     Squarewave for special applications.
                                                 ·     Clipped Sinewave - Simulates THD levels to test for harmonic distortion suscep-
                                                       tibility.
                                                 In addition to these standard waveforms, user defined waveforms can be downloaded
                                                 over the bus.

                                                 Harmonic Waveform Generation
Harmonicwaveform,Fund.,3rd,5th,7thand9th.        Using the latest DSP (Digital Signal Processing) technology, the Lx Series controller is
                                                 capable of generating harmonic waveforms to test for harmonics susceptibility of a unit
                                                 under test. With the help of the supplied Windows Graphical User Interface program,
                                                 defining harmonic waveforms is as easy as specifying the relative amplitude and phase
                                                 angle for each of up to 50 harmonics. The waveform data points are generated and
                                                 downloaded by the GUI to the AC source through one of the available interfaces and
                                                 remain in non-volatile memory. Up to twelve waveforms can be stored and given a
                                                 user defined name for easy recall.

                                                 Arbitrary Waveform Generation
                                                 Using the provided GUI program or custom software, the user also has the ability to
                                                 define arbitrary waveform data. Complex AC voltage anomalies can be simulated this
                                                 way. The GUI program provides a catalog of custom waveforms and also allows real-
                                                 world waveforms captured on a digital oscilloscope to be downloaded to one of the
                                                 AC source's waveform memories.
SimulationofsevereringingontheoutputofaUPS. Downloaded waveforms are retained in non-volatile memory for recall over the bus or
                                                 from the front panel. User defined waveform names make it easy to recall the desired
                                                 waveform when needed.


  Lx Series - Configuration Options
                                                Transient Programming
                                                To simulate common line disturbance occurrences, the Lx Series offers a list of transient
                                                steps. These steps can be programmed from the front panel or downloaded over the
                                                interface using the GUI program supplied. The GUI allows libraries of commonly used
                                                line disturbances to be created on disk for quick recall. Once downloaded, the transient
                                                program can be executed from the PC or from the front panel.
                                                AC transient generation allows the effect of rapid changes in voltage, frequency and
                                                waveform shape on the unit under test to be analyzed. The combination of transients
                                                and user defined arbitrary waveforms creates a powerful test platform for AC powered
                                                products.

Voltagesweeptransientcausesoutputvoltageto
changeataprogrammedrate.




                                                                                     www.california-instruments.com
                                                                                                                               An               Company



  Lx Series - Measurement and Analysis
 The Lx Series measurement system is based on real-time digitization of the voltage and current waveforms using a 4K sample buffer. The digitized
 waveform data is processed by a Digital Signal Processor to extract conventional load values such as rms voltage, rms current, real and apparent
 power. The same data is also used to perform Fast Fourrier Transformation (FFT) to extract the harmonic amplitude and phase angle of up to 50
 harmonics.
                                                    Standard Measurements
                                                    The following standard measurements are available from the front panel or via the bus:
                                                       · Frequency and Phase
                                                       · Voltage (rms)
                                                       · Current(rms) and Peak Current
                                                       · Crest Factor
                                                       · Neutral Current (rms)
                                                       · Real Power and Apparent Power
                                                       · Power Factor
                                                    Advanced Measurement Functions
                                                    In addition to standard load parameters, the Lx Series is capable of measuring voltage and
                                                    current amplitude and phase harmonics up to the 50th harmonic (for fundamental frequencies
                                                    up to 320 Hz). Total harmonic distortion of both voltage and current is also available.

 Standardmeasurementsforallphases.                  Harmonic analysis data can be displayed on the front panel display or on the PC using the
                                                    GUI program. The GUI can also be used to save and print harmonics data in tabular, bar
                                                    graph or time domain formats.
                                                    The acquired voltage and current time-domain waveforms for each output phase can be
                                                    displayed using the GUI program. Waveform displays on the PC include voltage and current
                                                    combined, three phase voltage, three phase current and true power. The time-domain data
                                                    is also available for transfer to a PC through IEEE-488, USB, RS232C, or Ethernet (option) when
                                                    using custom software.
                                                    Diagnostics Capability
                                                    The AC Source can perform a self test and report any errors. The self test will run until the first
                                                    error is encountered and terminate. The response to the self test query command will either
                                                    be the first error encountered or 0 if no error was found. (Self test passed).

                                                    Windows Graphical User Interface Control Software
                                                    A Windows Vista/2000/XPTM compatible Instrument Control Software (ICS) offers a soft front
  RelativeCurrentHarmonicsshownintableandchart.         panel interface for operation from a PC. The following functions are available:
                                                       ·   Steady state output control (all parameters).
                                                       ·   Create, run, save and print transient programs.
                                                       ·   Generate and save harmonic waveforms.
                                                       ·   Generate and save arbitrary waveforms.
                                                       ·   Download data from a digital storage oscilloscope.
                                                       ·   Measure and log standard measurements.
                                                       ·   Capture and display Voltage and Current waveforms.
                                                       ·   Measure, display, print and log harmonic voltage and current measurements.

                                                           Constant power mode chart shown applies to 3000Lx and 4500Lx models only.

                                                             Constant Power AC Mode - Available Max. AC Current
  SoftfrontpanelcontrolthroughWindowsGUI.
                                                                   125%
                                                             Current
                                                             (RMS) 100%


                                                                                                                                     Full
                                                                    50%                                                             Power


                                                                    20%


                                                                           10%                           50%                 80%             100%
                                                                                                                             Voltage (RMS)




www.california-instruments.com
                                            An                     Company



   Lx Series - Measurement and Analysis
 Output
Maximum Power per phase:                 3000Lx: 1 phase: 3000 VA, 3 phase: 1000 VA; 4500Lx: 1 phase 4500 VA, 3 phase 1500 VA; 6000Lx: 1 phase 5770 VA, 3 phase: 1923 VA
Power factor:                            0 to unity at full output VA

                                          Range:        V Low            V High
                                                                                                                    Programming Resolution:                       100 mV
Voltage Ranges:                           AC            0-150 V          0-300 V                                    Load Regulation:                              < 0.1 % FS ALC On or Reg.
                                         See-HVand-EHVoptionsforalternativevoltagerangepairs.                       Line Regulation:                              < 0.02 % for 10 % line change
Programming Accuracy                     Voltage (rms): ± (0.05% + 0.25) V from 5.0 V to FS, ALC On or Reg; Frequency: ± 0.025 %; Phase: ± 1° 45-100 Hz, ± (1° + 1°/kHz)
(25°C ±5°C):                             100 Hz-1kHz
Frequency Range:                         45 Hz - 1000 Hz (see -HF option for higher output frequencies) (Derated voltage from 17 Hz - 45Hz)
Frequency Resolution:                    0.01 Hz at < 81.9 Hz, 0.1 Hz at 82.0 to 819.1 Hz, 1 Hz at > 819 Hz

                                          Phases->            All       models                                      Model      3000Lx      3000Lx         4500Lx        4500Lx         6000Lx            6000Lx
                                                                                      <- Full Power, CP mode
                                          V Range:            3ø        1ø                                                     3ø          1ø             3ø            1ø             3ø                1ø
Max RMS Current:                          V low          12.8 A         38.4 A        At Full Scale Voltage ->      V Low      6.6 A       20.0 A         10.0 A        30.0 A         12.8 A            38.4 A
                                          V high         6.4 A          19.2 A                                      V High     3.3 A       10.0 A         5.0 A         15.0 A         6.4 A             19.2 A
                                         Note: Constant power mode on 3000Lx and 4500Lx provides increased current at reduced voltage; 6000Lx provides maximum power output.

Current Limit:                           Programmable from 0 Amps to maximum current for selected range
Peak Current:                            3000Lx: 5.7 X (lrms @ full scale voltage); 4500Lx: 3.8 X (lrms @ full scale voltage); 6000Lx: 3 X (lrms @ full scale voltage)
Output Noise (20 kHz to 1 MHz):          100mV rms typ.                                            Harmonic Distortion:            < 1% (at full scale voltage, full resistive load)
                                         250mV rms typ. for -HF option                             Output Relay:                   Push button controlled and bus controlled output relay
 Input
                                         ModeLx 3000Lx, 4500Lx, 9000Lx, 13500Lx: Standard: 208-230 ± 10% VAC, (L-L, 3 Phase); Option -400: 400 ± 10% VAC (L-L, 3 Phase);
Voltage:                                 ModeLx 6000Lx, 12000Lx, 18000Lx: Standard 208-230 + 10% VAC (L-L, 3 Phase)
                                         Notes: 1. Input must be specified when ordering. 2. -400 option not availble on 6000Lx, 12000Lx, 18000Lx. 3. 3000Lx can be operated from 1 phase AC.

                                                                                                                                                        Inrush Current       @ 180-254 V: 50 A peak
                                          Model                     3000Lx       3000Lx (1Phase)          4500Lx       6000Lx (@ 280V)
                                                                                                                                                        (Per phase):         @ 360-440 V: 83 A peak
Line Current (rms per phase):             187 VLL                   19 A         32 A                     31 A         38 A
                                                                                                                                                        Line Frequency: 47-440 Hz
                                          360 VLL                   10 A         n/a                      16 A         n/a

Efficiency:                               80% typical                                          Power Factor:           0.7 typical                                  Hold-up Time:              At least 10 ms
 Measurements
                                    Parameter          Frequency                   Phase            Voltage (AC)    Current (AC)         Crest Factor      Real Power        Apparent Power       Power Factor

                                    Range(s):          45.00-81.91 Hz              0.0° - 359.9°    0-400 V         0-50 A               1.00 - 10.0       0-6 kW            0-6 kVA              0.00-1.00
                                                       82.0-819.1 Hz
                                                       > 819 Hz

Measurements - Standard             Accuracy* (±)
                                    1 ø mode:                                      0.5° < 100 Hz                    0.1% + 0.15A         1.5%              0.15% + 9 W       0.15% + 9 VA         0.03
(ACMeasurements):                   3 ø mode:
                                                       0.1% + 1 digit
                                                                                   2° > 100 Hz
                                                                                                    0.05% + 0.25V
                                                                                                                    0.1% + 0.05A         1.5%              0.15% + 3 W       0.15% + 3 VA         0.01

                                    Resolution*:       0.01 Hz / 0.1 Hz / 1 Hz     0.1° / 1°        10 mV           1 mA                 0.01              1W                1 VA                 0.01

                                  * Accuracy specifications are in % of reading + scale error and apply above 100 counts. For multi-chassis configurations, current, power range and accuracy specifi-
                                  cations are times three. CF accuracy applies for Irms > 50 % of max. PF accuracy applies for PF > 0.5 and VA > 50% of max. Frequency measurement specification valid
                                  for output > 20 Vrms.

 System
Storage:                                Setup: 16 complete instrument setups / Transient List: 100 transient steps per list (SCPI mode)
Trigger Input/Output:                   Input: Triggers measurements or transient steps - SMA connector: 10K pull-up                                /     Output: SMA Connector: HCTTL output
 Protection
Overload/Temp/Voltage:                 Overload: Constant current or constant voltage mode; Over temperature: Automatic Shutdown; Over voltage: Automatic shutdown
Regulatory/RFI Suppresion:             IEC1010, EN50081-2, EN50082-2, CE, EMC, and safety mark requirements / RIF Suppression: CISPR 11, Group1, Class A



Note:      Specificationsaresubjecttochangewithoutnotice.Specificationsarewarrantedoveranambienttemperaturerangeof25°±5°C.Unlessotherwisenoted,specificationsareperphaseforasinewavewitharesistive
           loadandapplyaftera30minutewarm-upperiod.Forthreephaseconfigurations,allspecificationsareforL-N.Phaseanglespecificationsarevalidunderbalancedloadconditionsonly.




                                                                                                                                   www.california-instruments.com
                                                                                                                                                                                       An                     Company



       Lx Series - Specifications
 Remote Control
IEEE-488 Interface (option): IEEE-488 (GPIB) talker listener. Subset: AH1, C0, DC1, DT1, L3, PP0, RL2, SH1, SR1, T6, IEEE-488.2 SCPI Syntax
USB Interface & Ethernet:        Version: USB 1.1; Speed: 460 Kb/s maximum                   /    Ethernet Interface (Optional): specify -LAN option. 10BaseT, 100BaseT, RJ45
                                 Bi-directional serial interface; 9-pin D-shell connector. Handshake: CTS, RTS. Databits: 7 w/ parity, 8 w/o parity. Stopbits: 2. Baud rate: 9600 to
RS232C Interface:
                                 115200. Supplied with RS232C cable / Code and Format: SCPI / Not enabled if LAN option is present.
 Physical Dimensions
Dimensions (per chassis):        Height: 10.5" (267 mm), Width: 19" (483 mm), Depth: 23.7" (602 mm) (depth includes rear panel connectors)
Weight:                          Chassis: Net: 193 lbs / 87.7 Kg, Shipping: 280 lbs / 127.3 Kg (for /2 or /3 model configuarations multiply number of chassis).
Vibration and Shock:             Designed to meet NSTA project 1A transportation leveLx
Air Intake/Exhaust:              Forced air cooling, side air intake, rear exhaust
Temperature:                     Temperature: Operating: 0 to 35° C, full power / Storage: -40 to +85° C; Diagnostics: Built-in self test available over bus (*TST)
                                 * Three phase AC input and output terminal block with safety cover. * IEEE-488 (GPIB) connector, USB connector, RJ45 connector (with -LAN Op-
Rear Panel Connectors:           tion). * 9-pin D-Shell RS232C connector (RS232 DB9 to DB9 cable supplied). * Remote Inhibit (INH) and Discrete Fault Indicator (DFI). * Remote
                                 voltage sense terminal block. * Trigger In1 and Trigger Out1. * System interface connectors. * Auxilary Output (Option -AX)
 Harmonic Measurement Specifications
                                  Parameter            Frequency Fundamental Harmonics                   Voltage                                     Current
                                  Range                45-1000 Hz / 45 - 16 kHz                          Harmonics 2 - 50                            Harmonics 2 - 50
Measurements -
                                  Accuracy* (±)        0.1% + 1 digit / 0.1% + 2 digits                  0.1% + 0.1% /1 kHz +0.25V                   0.1% + 0.1% /1 kHz +0.05A
Harmonics:
                                  Resolution           0.01 Hz / 0.1 Hz                                  10 mV                                       10 mA

                                 * Accuracy specifications are in a percent of reading for single unit in 3-phase mode.

Waveforms:                       Pre defined: Sine, Square, Clipped; User defined: 1024 addressable data points; Storage: 50 user waveforms, non-volatile memory
Data Acquisition:                Parameters: Voltage & Current time domain, per phase; Resolution: 4096 data points, 10.4 usec (1ø) or 31.25 usec (3ø) sampling interval
 Option -AX Specifications
                                 Provides seperate isolated 26 VAC regulated and 5 Vac unregulated outputs. The 26 V is normally used for servo-synchro excitation, and the 5 V
Option -AX                       for lamp power. 26 Volt-Accuracy: ± 2%. Current capacity: 3 ARMS. Frequency: 360/440 Hz. Regulation ± 0.05%. 5 Volt-Accuracy: ± 5%. Current
                                 capacity: 5 ARMS
 Option -HV Specifications
Voltage Ranges:                  Low Range: 0-135 Volt; High Range: 0-270 Volt RMS
Max RMS Current at Full          3 Phase: High: 7.4 A, Low 14.8 A; 1 Phase: High: 22.2 A, Low: 44.4 A; Note: Constant power modes on 3000Lx and 4500Lx. Current available at
Power:                           reduced voltage for 3000Lx, 4500Lx, and max voltage for 6000Lx
Max RMS Current at Full          3000Lx: 3 Phase: High: 3.7 A, Low: 7.4 A; 1 Phase: High 11.1 A, Low: 22.2 A; 4500Lx: 3 Phase: High: 5.6, Low 11.1; 1 Phase: High: 16.7 A, Low: 33.3 A;
Scale Voltage:                   6000Lx: 3 Phase: High: 7.4 A, Low 14.8 A; 1 Phase: High: 22.2 A, Low: 44.4 A
 Option -EHV Specifications
Voltage Ranges:                  Low Range: 0-200 Volt; High Range: 0-400 Volt RMS
Max RMS Current at Full          3 Phase: High: 5.0 A, Low 10.0 A; 1 Phase: High: 15.0 A, Low: 30.0 A; Note: Constant power modes on 3000Lx and 4500Lx. Current available at
Power:                           reduced voltage for 3000Lx, 4500Lx, and max voltage for 6000Lx
Max RMS Current at Full          3000Lx: 3 Phase: High: 2.5 A, Low: 5.0 A; 1 Phase: High 7.5 A, Low: 15.0 A; 4500Lx: 3 Phase: High: 3.8, Low 7.5; 1 Phase: High: 11.3 A, Low: 22.5 A;
Scale Voltage:                   6000Lx: 3 Phase: High: 5.0 A, Low 10.0 A; 1 Phase: High: 15.0 A, Low: 30.0 A
 Option -HF Specifications
                                  Parameter         Frequency                 Phase              Voltage (AC)               Current (AC)      Crest Factor     Real Power     Apparent Power      Power Factor
Measurements:                     Range             45.00 - 5000 Hz           0.0°-359.9°        0-400 V                    0-50 A            1.00 - 10.0      0-6 kW         0-6kVA              0.00-1.00
                                                                                                 < 1000 Hz / > 1000 Hz
F < 1000 Hz: See standard         Accuracy* (±)                               0.5° < 100Hz                                  0.5% + 0.15A      1.5%             0.5% + 9 W     0.5% + 9 VA         0.03
                                                                                                 0.05% + 0.25V /
Lx Specifications;                1 ø mode          0.1% + 1 digit            2° < 2Khz
                                                                                                 0.1 % + 0.1%/kHz + 0.3V
                                  3 ø mode                                    5° > 2Khz                                     0.5% + 0.05A      1.5%             0.5% + 3 W     0.5% + 3 VA         0.01

F > 5000 Hz: See table --->       Resolution*       0.01 Hz / 0.1 Hz / 1 Hz   0.1°               10 mV                      1 mA              0.01             1W             1 VA                0.01

                                 * Accuracy specifications are in % of reading and apply above 100 counts. For multi-chassis configurations, current, power range and accuracy specifications are times three. CF ac-
                                 curacy applies for Irms > 50 % of max. Power factor accuracy applies for PF > 0.5 and VA > 50% of max. Frequency measurement specification valid for output > 20 Vrms.

Frequency Range:                 45 Hz- 5000 Hz (Derated voltage from 17 Hz - 45Hz)                                        Frequency Accuracy: ±0.025%
Output Noise:                    250 mVrms typical (20 kHz to 1 MHz)
    Note:      Specificationsaresubjecttochangewithoutnotice.Specificationsarewarrantedoveranambienttemperaturerangeof25°±5°C.Unlessotherwisenoted,specificationsareperphaseforasinewavewitharesistive
               loadandapplyaftera30minutewarm-upperiod.Forthreephaseconfigurations,allspecificationsareforL-N.Phaseanglespecificationsarevalidunderbalancedloadconditionsonly.




  www.california-instruments.com
                                               An                    Company

  Ordering Information                                                                                              Feature Comparison
  Model1                      Output           No of Output Phases            Nom. Input              -A350                Airbus ABD0100.1.8.1 tests.
                               Power               MODE                        Voltage2                                    Requires use of included
  3000Lx                        3 kVA                 1 and 3                  208-230 V                                   LxGui Windows software.
  3000Lx-400                     3 kVA                  1 and 3                 208-230 V             -B787                Boeing 787 tests. Requires
                                                                                                                           use of included LxGui Win-
  4500Lx                       4.5 kVA                  1 and 3                 208-230 V
                                                                                                                           dows software.
  4500Lx-400                   4.5 kVA                  1 and 3                      400 V            -LAN                 Ethernet Interface.
  6000Lx                         6 kVA                  1 and 3                 208-230 V             -MB                  Multi-box. Adds controller
  9000Lx/2                       9 kVA                  1 and 3                 208-230 V                                  to auxiliary chassis of multi-
  9000Lx/2-400                   9 kVA                  1 and 3                      400 V                                 chassis systems.
                                                                                                      -LKM1                Clock and Lock Master
  12000Lx/2                    12 kVA                   1 and 3                 208-230 V
                                                                                                      -LKS1,2              Clock and Lock Auxiliary
  13500Lx/3                   13.5 kVA                  1 and 3                 208-230 V
                                                                                                      -LNS2                Line Sync.
  13500Lx/3-400               13.5 kVA                  1 and 3                      400 V
                                                                                                      -EXS2                External Sync.
  18000Lx/3                    18 kVA                   1 and 3                 208-230 V
Note1:The/2or/3designationindicatesnumberofchassis.
                                                                                                      Cabinet Options
Note2:AllinputvoltagespecificationsareforLinetoLinethreephase,deltaorwye.Model3000Lx(208V
         input)canbeoperatedon230VL-Nsinglephaseifneeded.                                             -RMS       Rackmount Slides. Recom-
                                                                                                                 mended for rack mount ap-
Ordering Information                            Keypad Options                                                   plications.
Model                                           -RP        Ls Series style control                    C prefix   Cabinet System. Installed and
                                                           panel w/o numeric key-                                pre-wired in 19" cabinet.
   Refer to table shown for model
                                                           pad. See picture below.
   numbers and configurations.                                                                        Option Matrix:
Supplied with                                                                                         Note that some options are mutually ex-
   User / Programming Manual on                                                                       clusive as indicated in the table below. An
   CD-ROM, Software and RS232C                                                                        `o' means the options can be combined. An
   serial cable.                                                                                      `x' means they cannot.
Options
                                      Controller Options
Input Options                         -160        RTCA/DO-160 Rev D,
-400       400 ±10% Volt Line to Line             Change 2, EuroCAE-14D ,
           AC input.                              Rev E, Rev F firmware and
           [Not available on 6000Lx,              software [Section 16, AC
           12000Lx and 18000Lx                    only. Refer to -160 option
           Models]                                data sheet for details]
Output Options                                  -704           Mil-Std 704 rev D and E
-AX1       Auxiliary outputs, 26                               test firmware. [AC only,                         HF LF HV EHV LKM LKS EXS AX
           VAC, 5 VAC. Limits upper                            Refer to -704 option data               HF       -      x    o    o       x   x    o   x
           frequency to 800 Hz.                                sheet for details]                      LF       x      -    o    o       o   o    o   o
-EHV1      200/400 V output range.              -704F          Mil-Std 704 Rev A,B,C,F.                HV       o      o     -   x       o   o    o   o
-HV1       156/312 V output range.                             [AC Only]                               EHV o           o    x    -       o   o    o   o
-HF1
           Extends upper frequency              -ABD           Airbus Directive 0100.1.8               LKM x           o    o    o       -   x    o   o
           limit to 5000 Hz.                                   tests. [AC only]. Requires
                                                                                                       LKS      x      o    o    o       x   -    x   o
                                                               use of included LxGui
-LF1           Limits output frequency                                                                 EXS      o      o    o    o       o   x    -   o
                                                               Windows software.
               to 500 Hz.                                                                              AX       x      o    o    o       o   o    o   -
                                                -AMD           Airbus AMD24C tests.
                                                               Requires use of included
                                                               LxGui Windows software.
Note1:Seeoptionmatrix
Note2:-LKS,-LNSand-EXSaremutuallyexclusiveandwithExtTrigfunction.


9689 Towne Centre Drive, San Diego,GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
   CompuMess Elektronik CA 92121-1964 l (858) 677-9040 l FAX: (858) 677-0940 l sales@calinst.com
© Copyright 2008, AMETEK Programmable Power.        Specifications subject to change without notice                 Printed in the USA       DSLx 09/2008
       Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11 · www.compumess.de · www.netzteile.de
                                               www.california-instruments.com