     GLD150 Gold Performance Medical Switchers
     150 Watt Single Output

                                                                               FEATURES:
                                                                               · Compact 4.5" x 7" x 1.7" size
                                                                               · Power factor corrected to IEC 1000-3-2 Class A
                                                                               · Less than 300 µA leakage
                                                                               · EMI compliance to CISPR11, FCC Class B
                                                                               · Power fail and remote sense standard
                                                                               · Medical Approved to UL2601-1, IEC601-1/60601-1
                                                                                 and CSA-C22.2 No. 601.1
                                                                               · 2 year warranty
                                                                               ·     marked to LVD



     SPECIFICATIONS:
     Ac Input                                                                  EMI/EMC Compliance
     85-264 Vac, 47-63 Hz single phase.                                        All models include built-in EMI filtering to meet the EMC require-
     Input Current                                                             ments of IEC601-1. Unless otherwise stated, all tests are done
     2.8 A line current maximum, at 90 Vac, 60 Hz with full rated load,        at full load and 115 and 230 Vac input.
     power factor .99 typical, .96 minimum. Input current harmonic
     content meets the requirements of IEC1000-3-2.                            Conducted Emissions            EN55011, Class B; FCC Class B
     Output Power                                                              Static Discharge               EN61000-4-2, 6 kV contact 8 kV air
     150 W with convection cooling, 180 W with fan cooling.                    RF Field Susceptibility        EN61000-4-3, 3V/meter
     Efficiency                                                                Fast Transients / Bursts       EN61000-4-4, 2 kV, 5 kHz
     Minimum 80% at full rated load with 230 Vac Input. Approximately          Surge Susceptibility           EN61000-4-5, 1 kV diff., 2 kV com.
     3% less at 115 Vac.                                                       Conducted RF Susceptibility    EN61000-4-6, 3V
     Hold-Up Time                                                              Voltage Sags & Surges          EN61000-4-11
     Outputs will remain within regulation limits for 25 ms minimum from
     loss of ac input at full load, 10 ms before Power Fail indication.        Inrush Current
     Dc Ouput                                                                  Inrush 240 Vac is less than 37 A, averaged over the first ac half-
     Total regulation is the maximum deviation from the nominal voltage        cycle under cold start conditions. Limiting provided by internal
     for all steady state loading conditions. Peak ratings are for 60 s        thermistors.
     maximum duration, 10% duty cycle.                                         Fan Output
     Overload Protection                                                       An additional 12 Vdc, 250 mA output suitable for powering a dc
     Fully protected against short circuit and output overload. Short          fan is included in all models. The fan output is both current limited
     circuit protection is cycling type power limit.                           and thermally protected.
     Minimum Load                                                              Thermal Shutdown
     No minimum load required to maintain output specifications.               Provided as a standard feature. Designed to protect unit from
     Output Noise                                                              prolonged over temperature.
     0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode.                  Power Fail
     Measured with noise probe directly across output terminals of the         TTL / CMOS compatible output goes low (<0.5 V) 8 ms before
     power supply.                                                             output voltage drops more than 4% below nominal voltage upon
     Transient Response                                                        loss of ac power.
     Main Output - 500 µs typical response time for return to within           Power Good
     0.5% of final value for a 50% load step change, Di/Dt< 0.2 A µs.          TTL / CMOS compatible output goes high more than 100 ms
     Maximum voltage deviation is 3%.                                          after V1 reaches regulation and should assure that sufficient
     Remote Sense                                                              energy is stored in the input section to provide normal power
     Provided as a standard feature on all models.                             fail/shutdown.
     Overvoltage Protection                                                    Medical Approvals
     Built in on all models.                                                   All models are Certified to be in compliance with the
     Input Protection                                                          applicable requirements of UL2601-1, CSA-C22.2 No. 601.1,
     Internal ac fuses provided on both lines on all units.                    IEC601-1/60601-1.
     Voltage Adjustment                                                        Leakage Current
     Output Voltage is adjustable +/- 5% with user adjustable potentiometer.   70 µA, 132 Vac @ 60 Hz normal conditions. Single fault condi-
     Temperature Coefficient                                                   tions, 130 µA, 254 Vac @ 50 Hz.
     0.03% / °C typical on all outputs.                                        Design Verification Documents
     Overshoot                                                                 The "Gold" series has undergone rigorous review and design
     Less than 2% overshoot at turn-on under all conditions, less than         analysis. The following product documentation is available upon
     1% overshoot at turn-off under all conditions.                            request;
     Inhibit                                                                   1. MTBF study
     Inhibit signal is pulled to the V1 output common to reduce average        2. DVT Data
     output voltage to less than 5% of nominal.                                3. EMC / Susceptibility test results



                               CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                                   Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de

60
GLD150 Medical Switchers 150 Watt Multiple Output
Medical               Output            Output             Output            Voltage               Total                      OVP
                                                                                                                                                     Ripple/ Noise
Model                 Voltage          Current (A)        Current (B)       Adjustment           Regulation                  Setpoint
 GLD150-12             12 V             12.5 A               15 A             ± 5%                     2%                 14 + 1.1 V
                                                                                                                                                             1%
 GLD150-15             15 V                 10 A             12 A             ± 5%                     2%                18.5 + 1.5 V                        1%
 GLD150-24             24 V                 6.2 A            7.5 A            ± 5%                     2%                 28 + 2.5 V                         1%
 GLD150-28             28 V                 5.3 A            6.4 A            ± 5%                     2%                 34 + 2.8 V                         1%
 GLD150-48             48 V                 3.2 A           3.75 A            ± 5%                     2%                 55 + 4.0 V                         1%
Notes:
A. Maximum continuous current rating for unrestricted convection cooling.
B. Maximum continuous current rating with 150 LFM air or peak rating.
C. Add "C" suffix for cover option and derate convection rating to 130 W.




GLD150 MECHANICAL SPECIFICATIONS
INPUT
J1
MOLEX P.C.B. HEADER P/N: 39-30-2056
PIN 1) AC GROUND
PIN 2) N/C
PIN 3) AC NEUTRAL
PIN 4) N/C
PIN 5) AC LINE
MATING CONNECTOR MOLEX P/N
HOUSING 39-01-4051
CONTACT 39-00-0164

SIGNALS
J2
AMP P.C.B. HEADER P/N 641215-6
PIN 1) INHIBIT
PIN 2) +SENSE
PIN 3) POWER GOOD
PIN 4) -SENSE
PIN 5) COMMON
PIN 6) POWER FAIL
MATING CONNECTOR AMP P/N
HOUSING 770602-6
CONTACT 770666-6

OUTPUT
J3
MOLEX P.C.B. HEADER P/N: 39-29-9085
PINS 3,4,7,8) +Vout
PINS 1,2,5,6) RETURN
MATING CONNECTOR MOLEX P/N
HOUSING 39-01-2080
CONTACT 39-00-0163

FAN
AMP P.C.B. HEADER P/N: 641215-2
PINS 1) RTN
PINS 2) +12V
MATING CONNECTOR AMP P/N
HOUSING 770602-2
CONTACT 770666-02


 Environmental
                            Operating                 Non-operating
 Specification
                                   o
 Temperature (A)            0 to 50 C                  -40 to +85oC
 Humidity (A)              0 to 95% RH                 0 to 95% RH
                                                                               A. Units should be allowed to warm up/operate under non-condensing conditions
 Shock (B)                    20 gpk                       40 gpk                 before application of power.
                                                                               B. Random vibration--10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
 Altitude                -500 to 10,000 ft           -500 to 40,000 ft            orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
                                                                               C. Shock testing--half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes,
                                        2
 Vibration (C)         1.5 grms, 0.003 g /Hz         5 grms, 0.026 g2/Hz          total 6 shocks.



                      CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim