     GPC225 Commercial/GPM225 Medical
     225 Watt Global Performance Switchers

                                                                    FEATURES:
                                                                    · Wide-range ac input 85-264 Vac
                                                                    · 2-year warranty
                                                                    · Conducted EMI exceeds FCC Class B and CISPR 22
                                                                      Class B (Commercial models) and CISPR 11 Class B
                                                                      (Medical models)
                                                                    · Commercial Approved to UL1950, IEC950 and
                                                                      CSA22.2-234 L3
                                                                    · Medical Approved to UL2601-1, IEC601-1 and
                                                                      CSA22.2 No. 601
                                                                    ·     marked to LVD




     SPECIFICATIONS:
     Ac Input
     85-264 Vac, 47-63 Hz single phase.                             line impedance and distortion. Other settings are avail-
                                                                    able to the user through adjustment of built-in potentiometer.
     Input Current
     Maximum input current at 120 Vac, 60 Hz with full rated        Output Noise
     output load: 5.5 A                                             0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential
                                                                    mode. Measured with noise probe directly across output
     Hold-Up Time                                                   terminals of the power supply.
     20 ms minimum from loss of ac input at full load, nominal
     line (115 Vac).                                                Transient Response
                                                                    500 µs typical response time for return to within 0.5% of
     Output Power                                                   final value for a 50% load step change. i/t<0.2 A/µs.
     175 W convection; 225 W with 26 cfm air; 250 W peak. Peak      Maximum voltage deviation is 3.5%. Startup/shutdown
     ratings are for 60 s maximum duration, 10% duty cycle.         overshoot less than 3%.
     Overload Protection                                            Remote Sense
     Fully protected against short circuit and output overload.     Provided as a standard feature.
     Short circuit protection is cycling type power limit.
     Recovery after fault is automatic. Factory set to begin        Voltage Adjustment
     power limiting at approximately 250 W.                         Built-in potentiometer adjusts voltage ±5%.
     Overvoltage Protection                                         EMI/EMC Compliance
     124% ± 12% typical.                                            All models include built-in EMI filtering to meet the
                                                                    following emissions requirements:
     Efficiency                                                     EMI SPECIFICATIONS          COMPLIANCE LEVEL
     70% at full rated load, nominal input voltage, depending       Conducted Emissions GPC225 EN55022 Class B; FCC Class B
     on model and load distribution.                                Conducted Emissions GPM225 EN55011 Class B; FCC Class B
                                                                    Static Discharge           EN61000-4-2, 6 kV contact, 8 kV air
     Input Protection                                               RF Field Susceptibility     EN61000-4-3, 3 V/meter
     Internal ac fuse provided. Designed to blow only if a          Fast Transients/Bursts     EN61000-4-4, 2 kV, 5 kHz
     catastrophic failure occurs in the unit.                       Surge Susceptibility       EN61000-4-5, 1 kV diff., 2 kV com.

     Inrush Current                                                 Commercial Leakage Current
     Inrush is limited by internal thermistors. Inrush at 240 Vac   0.95 mA 254 Vac @ 60 Hz input (with no deviations).
     under cold start conditions will not exceed 60 A.
                                                                    Commercial Safety
     Temperature Coefficient                                        Approved to UL1950, CSA22.2 No. 234 Level 3, IEC950,
     0.03%/°C typical on all outputs.                               EN60950 and CISPR22. UL file #E135803 commercial;
                                                                    CSA #LR46516. The output(s) are intended for safety
     Thermal Shutdown                                               earthed Signal Output and Intermediate Circuits only. All
     Provided as a standard feature. Designed to protect unit       dc outputs are SELV under normal and single fault
     from prolonged over temperature.                               conditions.
     Environmental                                                  Medical Leakage Current
     Designed for 0 to 50°C operation at full rated output          60 µA. 254 Vac @ 60Hz input.
     power; derate output current and total output power by
     2.5% per °C above 50°C. See Environmental and                  Medical Safety
     Packaging Specifications on next page.                         Approved to UL2601, CSA22.2 No. 601 Level 3 and
                                                                    IEC601. UL file E116994; CSA #LR46516. The output(s)
     Power Fail                                                     are intended for safety earthed Signal Output and
     TTL- or CMOS-compatible output goes low (< 0.5 V) 5 ms         Intermediate Circuits only. The output(s) are not accept-
     before output voltage drops more than 4% below nominal         able for patient connection without additional isolation. All
     voltage upon loss of ac power. The signal is factory set to    dc outputs are SELV under normal and single fault
     trip on 84 to 94 Vac brown-out depending upon incoming         conditions.

                                                   CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
                          Condor D.C. Power Supplies, Inc., 2311 15 01 - 0 ·Parkway,(089) 32 15 01 - 11
                                               Telefon (089) 32 Statham Telefax Oxnard, CA 93033
66                        800-235-5929 · 805-486-4565 · FAX 805-487-8911 · www.condorpower.com
                                               info@compumess.de · www.compumess.de · www.netzteile.de
GPC225 Commercial/GPM225 Medical 225 Watt Multiple Output

 Commercial           Medical                                   Output                Output                   Output
 Model                Model                 Output             Minimum              Maximum (B)              Maximum (C)                   Peak             Noise P-P             Total Regulation (A)
 GPC225-5             GPM225-5                5V                   0A                     34 A                     40 A                    45 A                50 mV                      2%

 GPC225-12            GPM225-12               12 V                 0A                    15.8 A                    19 A                    20 A                120 mV                     2%

 GPC225-15            GPM225-15               15 V                 0A                    12.7 A                    15 A                    16 A                150 mV                     2%

 GPC225-24            GPM225-24               24 V                 0A                    8.3 A                     9.4 A                   10 A                240 mV                     2%

 GPC225-28            GPM225-28               28 V                 0A                    7.1 A                      8A                     8.5 A               280 mV                     2%

A.Total regulation is defined as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B. Unrestricted natural convection cooling.
C. Requires 26cfm moving air.




GPC225/GPM225 MECHANICAL SPECIFICATIONS




 Environmental
                                      Operating                         Non-operating
 Specification
 Temperature (A)                See individual specs                     -40 to +85oC
 Humidity (A)                        0 to 95% RH                         0 to 95% RH
                                                                                                          A. Units should be allowed to warm up/operate under non-condensing conditions
 Shock (B)                               20 gpk                              40 gpk                          before application of power.
                                                                                                          B. Random vibration--10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
 Altitude                         -500 to 10,000 ft                   -500 to 40,000 ft                      orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
                                                                                                          C. Shock testing--half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes,
                                                     2
 Vibration (C)                  1.5 grms, 0.003 g /Hz                5 grms, 0.026 g2/Hz                     total 6 shocks.



                                                             CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
                                   Condor D.C. Power Supplies, 32 15 2311 Statham Parkway, Oxnard, CA 93033
                                                 Telefon (089) Inc., 01 - 0 · Telefax (089) 32 15 01 - 11
                                   800-235-5929 · 805-486-4565 · FAX 805-487-8911 · www.condorpower.com                                                                                                  67
                                                             info@compumess.de · www.compumess.de · www.netzteile.de