Hier Klicken für das PDF Dokument!!!

              GXM60 Medical Single Output
              60 Watt-Range Input High Efficiency Desk-Top Supply

                                                                        FEATURES:
                                                                        · Universal Input
                                                                        · Small size, light weight
                                                                        · Medically approved to UL2601-1, CSA 22.2 No.
                                                                          601.1-M90, EN60601-1, and CB certified to IEC601-1.
                                                                        · Ultra efficient - Up to 90%
                                                                        · Spill proof external enclosure
                                                                        · 2 year warranty
                                                                        ·      marked to LVD




SPECIFICATIONS:
AC INPUT                                                                EFFICIENCY
90-264 Vac, 47-63 Hz single phase. Input power is less than 75          77 to 90% at full rated load depending upon model and ac line.
W when output power is less than 60 W, 230 Vac in. Meets                Units are typically > 85 % @ 230 Vac. Efficiency is measured at
EN61000-3-2 harmonics standard.                                         the supply output, cabling losses are not considered and will
OUTPUT POWER                                                            vary with different cables. Contact factory for assistance.
Total continuous output power is 60 W (55 for any unit less than        OVERSHOOT
8 V), 65 W peak for 60 s 10% duty cycle. Total power specified          Less than 3% overshoot at turn-on under nominal conditions.
from supply output, derate for cable losses.                            TURN-ON TIME
OUTPUT REGULATION                                                       Less than 1.5 s at 115 Vac, 25 °C
Supply regulation is +/- 1%. However, actual regulation limits are      INPUT PROTECTION
affected by cable losses on units without remote sense. Final           Both lines protected with internal ac fuses, provided on all units.
regulation limits must include cable losses. Contact factory for        Designed to blow only if a catastrophic failure occurs in the unit
assistance.                                                             -- Fuse does not blow on short circuit or overload.
INPUT CURRENT                                                           INRUSH CURRENT
Maximum input current at minimum input voltage and output               Inrush is limited by internal thermistor. The inrush at 240 Vac,
overload will be less than 2.0 A.                                       averaged over the first ac half-cycle under cold start conditions
HOLD-UP TIME                                                            will not exceed 60 A.
Output voltage stays within regulation for 20 ms from loss of ac        EMI / EMC COMPLIANCE
input at full load, from 120 Vac input.                                 All models include built-in EMI filtering.
MINIMUM LOAD                                                            Conducted Emissions              EN55011, Class B; FCC Class B
No minimum load required.                                               Static Discharge                 EN61000-4-2, 6 kV contact 8 kV air
OVERLOAD PROTECTION                                                     RF Field Susceptibility          EN61000-4-3, 3V/meter
Fully protected against short circuit and output overload. Short        Fast Transients / Bursts         EN61000-4-4, 2 kV, 5 kHz
circuit protection is cycling type power limit and will automatically   Surge Susceptibility             EN61000-4-5, 1 kV diff., 2 kV com.
recover after removal of fault.                                         Conducted RF Susceptibility EN61000-4-6, 3V
OUTPUT NOISE                                                            Voltage Sags & Surges            EN61000-4-11
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode.                MEDICAL SAFETY AGENCY
Measured with a scope probe directly across the output                  All models are approved to UL2601, CSA 22.2 No. 601.1-M90,
connector. Load terminated with a 0.1 µF capacitor in parallel          IEC 601-1 (1988), EN 60601-1
with a low Z 10 mF capacitor.
                                                                        LEAKAGE CURRENT
TRANSIENT RESPONSE
                                                                        80 µA under normal conditions (132 Vac @ 60 Hz ). Maximum
Response at supply is 1.5 ms typical for return to regulation for
                                                                        under single fault conditions (264 Vac @ 50 Hz) is 170 µA.
a 50% load step, i/t< 0.2 A/µs. Maximum voltage deviation is
3.5%. Cable resistance will effect maximum voltage deviation.            Environmental
                                                                                                           Operating                       Non-operating
OVER VOLTAGE PROTECTION                                                  Specification
Prevents output from producing excessive voltage in the event            Temperature (A)                    0 to 40ºC                        -40 to +85ºC
of a single fault failure. Voltage limited typically to 130% depend-
                                                                         Humidity (A)                     0 to 95% RH                        0 to 95% RH
ing upon model. OVP firing reduces output to less than 50% of
nominal voltage in less than 50 ms.                                      Shock (B)                            20 gpk                             40 gpk
OVERLOAD PROTECTION                                                      Altitude                       -500 to 10,000 ft                 -500 to 40,000 ft
Factory set to begin power limiting at approximately 75 W.
                                                                         Vibration (C)               1.5 grms, 0.003 g2/Hz               5 grms, 0.026 g2/Hz
                                                                        A. Units should be allowed to warm up/operate under non-condensing conditions
                                                                           before application of power.
                                                                        B. Random vibration--10 to 2000 Hz, 6 dB/octave roll-off from 350 to 2000 Hz, 3
                                                                           orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
                                                                        C. Shock testing--half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes,
                                                                           total 6 shocks.

                    CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                        Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
GXM60 Medical Single Output

Typical Standard Model Configurations:
 Medical                                      Output                  Current            Initial (1)           Total                   Output                 Remote          Ripple/
 Model                                         (V)                      (I)              Set Point           Regulation                Harness                Sense           Noise

 GXM60-5A11                                       5V                   11 A                 2%                    1%                     Molex                   Yes           1.5%

 GXM60-12A01                                      12 V                 5.0 A                2%                    1%                    2.1 mm                    No           1%

 GXM60-15A01                                      15 V                4.0 A                 2%                    1%                    2.1 mm                    No           1%

 GXM60-24A01                                      24 V                2.5 A                 2%                    1%                    2.1 mm                    No           1%

 GXM60-28A01                                      28 V                2.2 A                 2%                    1%                    2.1 mm                    No           1%

 GXM60-36A01                                      36 V                 1.7 A                2%                    1%                    2.1 mm                    No           1%

Notes: 1) Initial set point without load. Total regulation shown includes cable losses. When selecting cable options without remote sense, cable drop should be considered.


MODEL NUMBER CONFIGURATION
GXM60-XXXXX
                                                  Output Cable/Plug Option
                                                  Number


                                         Case Color:             A = Beige
                                                                 B = Black
                                  Output Voltages:
                                  Units are available standard with any
                                  output voltage from 5 to 35 Vdc in 1.0
                                  increments. Also available in 0.1
                                  increments data custom order.

    Option      Output Cord/Plug                                         Cable/Plug
     No.                                                               Maximum Rating
      01        Circular 5.6 mm O.D. , 2.1 mm I.D., 18 ga.             5 A, 80 V max.
      02        Circular 5.6 mm O.D., 2.5 mm I.D., 18 ga.              5 A, 80 V max.
      03        Circular 5.6 mm O.D., 2.1 mm I.D., RT angle, 18 ga.    5 A, 80 V max.
      04        Circular 5.6 mm O.D., 2.5 mm I.D., RT angle, 18 ga.    5 A, 80 V max.
      05        Circular 5.6 mm O.D., 2.1 mm I.D., 18 ga., shielded    5 A, 80 V max.
      06        Circular 5.6 mm O.D., 2.5 mm, I.D.,18 ga., shielded    5 A, 80 V max.
      07        Mini DIN 20 ga.                                        6 A, 36 V max.
      08        Mini DIN 20 ga. shielded                               6 A, 36 V max.
      09        5 Pin 20 ga. DIN plug                                  6 A, 36 V max.
      10        5 Pin 20 Ga. DIN plug, shielded                        6 A, 36 V max.
      11        Molex Mini-Fit, 18 ga.                                11 A, 480 V max.
      12        Power DIN                                              11 A, 20 V max.
      13        Power DIN., shielded                                   11 A, 20 V max.



MECHANICAL SPECIFICATIONS




                       CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                              Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
                                                                                                                                                                                        107