Hier Klicken für das PDF Dokument!!!
iX and i Series II
Power Systems
3-15 kVA Programmable
AC and DC
Power Source / Analyzer
Intergrated System
The iX Series II represents a new generation of AC and DC power source that addresses
Combination AC and DC Power increasing demands on test equipment to perform more functions at a lower cost. By
combining a flexible AC/DC power source with a high performance power analyzer, the
Source and Power Analyzer
iX Series II systems are capable of handling complex applications that have traditionally
Replaces multiple instruments with a
required multiple instruments.
single multifunction unit
The sleek integrated approach of the iX Series II avoids the cable clutter that is commonly
3000 VA - 15000 VA of Output Power found in AC test systems. The i/iX Series II is rackmountable with a 4U chassis design. All
Capable of handling a wide range of connections are made internally and the need for external digital multimeters, power
loads with 0 to 1 power factor harmonics analyzer and current shunts or clamps is completely eliminated.
Arbitrary Waveform Generation Using a state of the art digital signal processor in conjunction with precision high resolu-
User defined voltage waveform and tion A/D converters, the iX Series II provides more accuracy and resolution than can be
distortion programming found in some dedicated harmonic power analyzers. Since many components in the iX
Built-in Digital Power Analyzer Series II are shared between the AC/DC source and the power analyzer, the total cost of
Analyze frequency and time domain of the integrated system is less than the typical cost of a multiple unit system.
both voltage and current on all phases For less demanding applications, the i Series II provides similar output and transient
capabilities as the iX Series II, as well as basic measurements.
Scope Capture Capability
Built in voltage and current
Easy To Use Controls
waveform acquisition
capability (iX Series II) Both the iX Series II and i Series II are microprocessor controlled and can be operated
from an easy to use front panel keypad. Functions are grouped logically and are directly
EN61000-3-2 and EN61000-3-3 accessible from the keypad. This eliminates the need to search through various levels of
Meets source requirements for IEC menus and/or soft keys.
Harmonics and Flicker testing
A large analog control knob can be used to quickly slew output parameters. This knob
Powerful Programing Software is controlled by a dynamic rate change algorithm that combines the benefits of precise
Powerful, yet easy to use, program- control over small parameter changes with quick sweeps through the entire range.
ming software included (also includes
Avionics test software). Applications
With precise output regulation and accuracy, the iX Series II AC and DC sources address
Constant Power Mode
many application areas for AC and DC power testing. The iX also provides a high load
Provides additional current at lower
current capability, multi or single phase output modes, and built-in power analyzer
voltages (See current voltage curve)
measurements. Additional features including line distortion simulation (LDS), arbitrary
waveform generation, and programmable output impedance address requirements for
product quality and regulatory compliance testing.
www.california-instruments.com ISO 9001:2000
iX and i Series II - Multi-Function and Multi-Use
Product Evaluation and Test Remote Control
Increasingly, manufacturers of electronic equipment and appliances Standard IEEE-488 and RS232C remote control interfaces allow
are required to fully evaluate and test their products over a wide programming of all instrument functions from an external com-
range of input line conditions. The built-in output transient genera- puter. The popular SCPI command protocol is used for program-
tion and readback measurement capability offers the convenience ming. Drivers for several popular instrumentation programming
of an easy to use and integrated test system. environments are available to facilitate systems integration of the
i/iX Series II. Instrument drivers for popular programming environ-
Avionics
ments such as National Instruments LabViewTM are available to
With an output frequency range to 1000 Hz, at up to 150 VRMS, speed up system intergration.
the iX Series II is well suited for aerospace applications. Precise fre-
quency control and accurate load regulation are key requirements Application Software
in these applications. The standard IEEE-488 control interface and
SCPI command language provide for easy integration into existing
ATE systems. Since the iX Series II can eliminate the need for several
additional pieces of test equipment and only occupies 7 inches of
rack space (4U), it significantly saves cost and space. Options are
available for popular avionics test routines such as: DO-160, ABD-
0100, MIL-STD-704A-F, and Boeing 7E73B-0147.
Regulatory Testing
As governments are moving to enforce product quality standards,
regulatory compliance testing is becoming a requirement for a
growing number of manufacturers. The iX Series II is designed
to meet AC source requirements for use in Euronorm EN 61000
compliance testing. For flicker testing, the programmable output
impedance capability of the 3001iX, 5001iX and 15003iX can be used
to create the required IEC 725 reference impedance. Run IEC61000-
4-11, IEC61000-4-14 and IEC61000-4-28 test programs. Windows® application software is included with the iX and i Se-
ries II1 . This software provides easy access to the power source's
Multi-Box Configurations capabilities without the need to develop any custom code. The
For high power applications, two or three 5001i/iX chassis can be following functions are available through this GUI program:
· Steady state output control (all parameters)
· Create, run, save, reload and print transient programs
· Generate and save harmonic waveforms [iX only]
· Generate and save arbitrary waveforms [iX only]
· Download data from a digital storage oscilloscope
[iX only]
· Measure and log standard measurements
MODE-iX Option · Capture and display output voltage and current waveforms
[iX only]
combined to provide 10 to 15 kVA of single or three phase power.
· Measure, display, print and log harmonic voltage and current
measurements [iX only]
A 9003iX, 15003iX or 15003i three phase configuration can be
ordered with the MODE-iX option. This option allows automatic · Display IEEE-488 or RS232C bus traffic to and from the AC
switching between single or three phase output mode. In single Source to help you develop your own test programs.
phase mode, all current is available on phase A.
High Crest Factor 1. Requires PC running WindowsXPTM or Windows 2000TM.
With a crest factor of up to 5:1, the i/iX Series II AC source can drive
difficult nonlinear loads with ease. Since many modern products
use switching power supplies, they have a tendency to pull high
repetitive peak currents. The 5001iX can deliver up to 110 Amps of
repetitive peak current to handle such loads.
ISO 9001:2000 www.california-instruments.com
iX and i Series II - Waveform Generation
Harmonic Waveform Generation
Using the latest DSP technology, the iX Series II controller is capable of gen-
erating harmonic waveforms to test for harmonics susceptibility of a unit
under test. Included is a Graphical User Interface program that can be used
to define harmonic waveforms by specifying amplitude and phase for up
Harmonic waveform, Fund., 3rd, 5th, 7th, 9th,11th and 13th. to 50 harmonics. The waveform data points are generated and downloaded
by the GUI to the AC source through either the IEEE-488 or RS232C bus and
remain in nonvolatile memory. Up to 200 waveforms can be stored and
given a user defined name for easy recall.
The three phase configuration iX Series II offers independent waveform gen-
eration on each phase allowing three phase anomalies to be programmed.
It also allows simulation of unbalanced harmonic line conditions.
Two hundred user defined waveforms.
Arbitrary Waveform Generation [iX Series II only]
Using the provided GUI program or custom software, the user also has the
ability to define arbitrary AC waveforms. The arbitrary waveform method
of data entry provides an alternative method of specifying AC anomalies
by providing specific waveform data points. The GUI program includes a
catalog of custom waveforms. It also allows real-world waveforms cap-
Harmonically distorted waveform.
tured on a digital oscilloscope to be downloaded to one of the many AC
source's waveform memories.
Arbitrary waveform capability is a flexible way of simulating the effect of
real-world AC power line conditions in both engineering and production
environments.
iX and i Series II - AC and DC Transient Generation
The iX and i Series II controllers have a powerful AC and DC transient gen-
eration system that allows complex sequences of voltage, frequency and
waveshapes to be generated. This further enhances the i/iX's capability to
simulate AC line conditions or DC disturbances. When combined with the
multi phase arbitrary waveform capabilities, the AC and DC output possibili-
Transient List Data Entry from the front panel. ties are truly exceptional. In three phase i/iX system configurations, transient
generation is controlled independently yet time synchronized on all three
phases. Accurate phase angle control and synchronized transient list execu-
tion provide unparalleled accuracy in positioning AC output events.
Transient programming is easily accomplished from the front panel where
clearly laid out menu's guide the user through the transient definition
process. The front panel provides a convenient listing of the programmed
transient sequence and allows for transient execution, Start, Stop, Abort
and Resume operations.
User defined transient sequences can be saved to nonvolatile memory for
instant recall and execution at a later time. The included Graphical User In-
terface program supports transient definitions using a spreadsheet-like data
entry grid. A library of frequently used transient programs can be created
on disk using this GUI program.
Transient List Data Entry in GUI program.
www.california-instruments.com ISO 9001:2000
iX and i Series II - Measurement and Analysis
The i/iX Series II is much more than a programmable AC and DC power source.
It also incorporates an advanced digital signal processor based data acquisi-
tion system that continuously monitors all AC source and load parameters.
This data acquisition system forms the basis for all measurement and analysis
functions. These functions are accessible from the front panel and the remote
Measurement data for single phase (MX30/45 Display). control interface.
Conventional Measurements
Common AC and DC measurement parameters are automatically provided
by the data acquisition system on both iX and i Series models. These values
are displayed in numeric form on the front panel LCD display. The following
measurements are available: Frequency, Vrms, Irms, Ipk, Crest Factor, Real Power
Measurement data for all three phases (MX30/45 Display). (Watts), Apparent Power (VA) and Power Factor.
Harmonic Analysis [iX Series II only]
The iX Series II provides detailed amplitude and phase information on up to
50 harmonics of the fundamental voltage and current for either one or three
phases. Harmonic content can be displayed in both tabular and graphical
formats on the front panel LCD for immediate feedback to the operator. Al-
Absolute amplitude bar graph display of current harmonics with cur- ternatively, the included GUI program can be used to display, print and save
sor positioned at the fundamental (MX30/45 Display). harmonic measurement data. Total harmonic distortion of both voltage and
current is calculated from the harmonic data.
Voltage harmonic measurement table display in absolute values
(MX30/45 Display).
Waveform Acquisition [iX Series II only]
The measurement system is based on real-time digitization of the voltage and
current waveforms using a 4K deep sample buffer. This time domain informa-
tion provides detailed information on both voltage and current waveshapes.
Waveform acquisitions can be triggered at a specific phase angle or from a
Acquired Current waveform (MX30/45 Display). transient program to allow precise positioning of the captured waveform with
respect to the AC source output.
The front panel LCD displays captured waveforms with cursor readouts. The
included GUI program also allows acquired waveform data to be displayed,
printed and saved to disk.
Acquired Voltage waveform (MX30/45 Display).
Acquired three
phase voltage
waveforms
display on PC.
ISO 9001:2000 www.california-instruments.com
iX and i Series II - Specifications
Operating Modes
iX Series II: AC, DC or AC+DC
i Series II: AC or DC
AC Mode Output
Frequency: Range: 16.00-1000 Hz (Note: Voltage on 300 V range derates from 300 Vrms max at 500 Hz to 150 Vrms max at 1000 Hz; See V-F rating chart. below)
3001i/iX: 3000 VA, 5001i/iX: 5000 VA, 9003i/iX: 3000 VA 3ø, with mode iX: 9000 VA 1ø, 10001i/iX: 10000 VA, 15001i/iX: 15000
Total Power:
VA, 15003i/iX: 5000 VA/ø 3ø, with mode iX: 15000 VA/ø 1ø
Load Power Factor: 0 to unity at full output VA
AC Mode Voltage
Load Regulation
< 0.2%
Range: V Low V High (with ALC on):
Voltage Ranges: AC 0-150 V 0-300 V Load Regulation < 0.5% DC to 100 Hz, < 0.6% 100 Hz to 500 Hz in high voltage range,
AC+DC 0-150 V 0-300 V
(with ALC off ): < 2.2% 100 Hz to 500 Hz in low voltage range, < 3% 500 Hz to 1000 Hz
Line Regulation: < 0.1% for 10% line change
Output Noise (20 kHz to 1 MHz): < 250 mVrms typ., < 500 mVrms max.
Harmonic Distortion (Linear): < 1% from 16 - 66 Hz, < 2% at 400 Hz, < 3% at 800 Hz (Full resistive load)
DC Offset: < 20 mV External Amplitude Modulation: Depth: 0 - 10 %, Frequency: DC - 2 KHz
Isolation Voltage: 300 Vrms output to chassis
AC Mode Current
Model 3001i/iX 5001i/iX 9003i/iX 9003i/iX 3ø 10001i/iX 15001i/iX 15003i/iX 3ø
300 V range 11.1 18.5 33.3 11.1 37.0 55.5 18.5
Steady State AC Current:
150 V range 22.2 37.0 66.6 22.2 74.0 111.0 37.0
Note: Constant power mode provides increased current at reduced voltage (See chart below)
Model 3001i/iX 5001i/iX 9003i/iX 9003i/iX 3ø 10001i/iX 15001i/iX 15003i/iX 3ø
Peak Repetitive AC Current: High range 96.0 96.0 288.0 96.0 192.0 288.0 96.0
Low range 110.0 110.0 330.0 110.0 220.0 330.0 110.0
Voltage (rms): ± 0.2 of range, 16 to 1000 Hz, Frequency: ± 0.01 % of programmed value, Current Limit: ± 0.5 % of programmed
Programming Accuracy:
value, Phase: < 1.5° with balanced load at 50/60 Hz
Voltage (rms): 100 mV, Frequency: 0.01 Hz from 16 - 81.91 Hz, 0.1 Hz from 82.0 - 819.1 Hz, 1 Hz from 820-1000 Hz, Current Limit:
Programming Resolution:
0.1 A, Phase, 0.1°
Output Relay: Push-button controlled or bus controlled output relay
Output Impedance (iX Only): Programmable Z on 3001iX, 5001iX, 9003iX and 15003iX (3ø mode only) for 50 Hz fundamental
Resistive: Range: 17 - 1000 mOhm,resolution: 4 mOhm accuracy, 2 % FS
Inductive: Range: 230 - 1000 µH, resolution: 4 µH, accuracy: 2 % FS
Constant Power AC Mode - Available Max. AC Current V-F Rating Chart - 300V Range
100% 350
Current 90% 300
(RMS) 75%
250
Volt (RMS)
200
Full
50%
Power
150
25% 100
50
0
0% 5% 25% 50% 75% 90% 100% 16 200 400 500 600 800 1000
Frequency (Hz)
Voltage (RMS)
Note: Specifications are subject to change without notice. Specifications are warranted over an ambient temperature range of 25°± 5° C. Unless otherwise noted, specifications are per phase for a sinewave with a resistive
load and apply after a 30 minute warm-up period. For three phase configurations, all specifications are for L-N. Phase angle specifications are valid under balanced load conditions only.
www.california-instruments.com ISO 9001:2000
iX and i Series II - Specifications
Measurement
Parameter Frequency RMS Voltage RMS Current Peak Current Crest Factor Real Power Apparent Pwr Pwr Factor DC Voltage DC Current Power
Range 16-500 Hz 0-300 V 0-40 A 0-120 A 0.00-6.00 0-6 kW 0-6 kVA 0.00-1.00 0-300 V 0-40 A 0-6 kW
Accuracy* (±) < 100 Hz 0.5 V + 0.02% 0.5 A + 0.02% 0.5 A + 0.02% 0.05 10 W + 0.1% 10 VA + 0.1% 0.01 150 mV 70 mA 15 W
0.01% +
Measurements - Standard 0.01 Hz
(AC Measurements): 100 - 500 Hz 0.1 V + 0.02% 0.1 A + 0.02% 0.1 A + 0.02% 0.05 20 W + 0.1% 20 VA + 0.1% 0.02 150 mV 70 mA 15 W
Resolution* 0.01 Hz 10 mV 1 mA 1 mA 0.01 1W 1 VA 0.01 10 mV 1 mA 1W
* Measurement system bandwidth = DC to 48 kHz. Accuracy specifications are valid above 100 counts. Current and Power Accuracy specifications are times two for
10001i/iX /30003i/iX and times three for 15001i/iX and 9003i/iX and 15003i/iX with MODE-iX option when in single phase mode. For 10001i/iX, 15001i/iX and 30003i/
iX, resolution decreases by factor of 10, ranges for current and power increases by factor of three. PF accuracy applies for PF > 0.5 and VA > 50% of range.
Parameter Frequency Fundamental Harmonics Phase Voltage Current
Range 16.00-500.0 Hz / 32.00 Hz - 48.00 kHz 0.0 - 360.0° Fundamental Harmonics 2 - 50 Fundamental Harmonics 2 - 50
Measurements - Harmon- Accuracy* (±) 0.01% + 0.01 Hz 2° typ. 250 mV 0.1% + 250 mV+0.1% /1 kHz 50 mA/0.1% + 50 mA +0.1% /1 kHz
ics:
Resolution 0.01 Hz 0.5° 10 mV / 10 mV 10 mA / 10 mA
* Accuracy specifications are valid above 100 counts. Accuracy specifications are times three for three phase mode. Harmonics frequency range in three phase mode is
32 Hz - 16 kHz. Resolution decreases by factor of 10 for 9003iX and 15003iX with Mode-iX option in 1 phase mode and for 10001iX, 15001iX and 30003i/iX.
DC Mode Output
Power (Max at full scale of 3001i/iX: 2100 W, 5001i/iX: 3500 W, 9003i/iX: 2100 W/ø 3ø, 6300 W/ø 1ø, 10001i/iX: 7000 W, 15001i/iX: 10500 W, 15003i/iX: 3500
DC Voltage Range): W/ø 3ø, 10500 W/ø 1ø
Voltage Ranges: Range: Low: 200 Vdc, High: 400 Vdc
Line Regulation: < 0.1% FS or 10% line change
Output Noise:: < 250 mV rms typ., < 500 mV rms max., (20 kHz to 1 MHz)
Model 3001i/iX 5001i/iX 9003i/iX 9003i/iX 3ø 10001i/iX 15001i/iX 15003i/iX 3ø
Max DC Current (Maximum
400V range 7.8 13 23.4 7.8 26 39 13
current at 65% of V Range):
200V range 15.6 26 46.8 15.6 52 78 26
Current Limit Programmable from 0 A to maximum current for selected range
AC+DC Mode Output
Output Power (iX only): Full AC Power if DC component is less than 20% of full scale voltage, Full DC power if DC component is above 20%
Storage
Non Volatile Mem. storage: 16 instrument setups, 200 user defined waveforms
Waveforms
Waveform Types: i Series II: Sine, iX Series II: Sine, Square, Clipped sine, User defined
User defined waveform
Four groups of 50 user defined arbitrary waveforms of 1024 points for a total of 200 (One group can be active at a time)
storage:
System Interface
Inputs: Remote shutdown, External Sync, Clock/Lock (option)
Outputs: Function Strobe, Clock/Lock (option)
Protection
Over Load: Constant Current or Constant Voltage mode
Over Temperature: Automatic shutdown
Note: Specifications are subject to change without notice. Specifications are warranted over an ambient temperature range of 25°± 5° C. Unless otherwise noted, specifications are per phase for a sinewave with a resistive
load and apply after a 30 minute warm-up period. For three phase configurations, all specifications are for L-N. Phase angle specifications are valid under balanced load conditions only.
ISO 9001:2000 www.california-instruments.com
iX and i Series II - Specifications
Remote Control
IEEE-488 Interface: IEEE-488 (GPIB) talker listener. Subset: AH1, C0, DC1, DT1, L3, PP0, RL2, SH1, SR1, T6, IEEE-488.2 SCPI Syntax
9 pin D-shell connector, Handshake: CTS, RTS, Data bits: 7,8, Stop bits: 1, 2, Baud rate: 9600, 19200, 38400, 57600, 115200, IEEE-488.2
RS232C Interface:
SCPI Syntax (Supplied with RS232C cable).
AC Input
3001 and 9003: 208-240 ± 10% Vac, (L-N, 1ø), All other models: Standard: 208-240 ± 10% Vac, (L-L, 3ø), Option -400: 400-480 ± 10%
Voltage:
Vac, (L-L, 3ø) (Input range must be specified when ordering).
Model 3001i/iX 5001i/iX 9003i/iX 10001i/iX 15001i/iX 15003i/iX
Input Line Current (per
187-264V 25 A 23 A 75 A 46 A 69 A 69 A
phase):
360-528V N/A 12 A N/A 24 A 36 A 36 A
Inrush Current per chassis: < 100 Apk for 100 µs at 208-240 V, < 50 Apk for 100 µs at 400-480 V
Line Frequency: 50-60 Hz ± 10 %
Efficiency: 75% typical
Power Factor: 0.6 typical
AC Service
Regulatory: IEC61010, EN50081-2, EN50082-2, CE EMC and Safety Mark requirements
RFI Suppression: CISPR 11, Group1 , Class A
AC Input & Output terminal block with cover, IEEE-488 (GPIB) connector (rear panel), 9 pin D-Shell RS232C connector*, Remote volt-
Rear Panel Connectors:
age sense terminal block, System Interface Connector, *RS232 DB9 to DB9 cable supplied
Physical Dimensions
Dimensions: Height: 7" (178 mm), Width: 19" (483 mm), Depth: 24" (610 mm) (depth includes rear panel connectors)
Weight: per 5001i/iX Chassis: Net: 61 lbs / 28 Kg, Shipping: 115 lbs / 52 Kg
Vibration and Shock: Designed to meet NSTA project 1A transportation levels
Air Intake/Exhaust: Forced air cooling, side air intake, rear exhaust.
Operating Humidity: 0 to 95 % RAH, non condensing.
Temperature: Operating: 0 to 40° C, Storage: -20 to +85° C
i/X Series II Rear Panel
Note: Specifications are subject to change without notice. Specifications are warranted over an ambient temperature range of 25°± 5° C. Unless otherwise noted, specifications are per phase for a sinewave with a resistive
load and apply after a 30 minute warm-up period. For three phase configurations, all specifications are for L-N. Phase angle specifications are valid under balanced load conditions only.
www.california-instruments.com ISO 9001:2000
Ordering Information Feature Comparison
Standard controller versions with single voltage range:
Model Output Phase Max. current per phase Input Controller: i iX
Power Output Low V range High V range Voltage2
AC AC DC AC DC AC mode x x
3001i/iX 3 kVA 1 22 15.6 11 7.8 208-240V DC mode x x
5001i/iX 5 kVA 1 37 26 18.5 13 208-240V AC+DC mode x
5001i/iX-400 5 kVA 1 37 26 18.5 13 400-480V Transient
9003iX1 9 kVA 3 22 15.6 11 7.8 208-240V programming x x
10001i/iX 1
10 kVA 1 74 52 37 26 208-240V Arbitrary
waveforms x
10001i/iX-400 1
10 kVA 1 74 52 37 26 400-480V
10002i/iX1 10 kVA 2 37 26 18.5 13 208-240V
Measurements x x
10002i/iX-4001 10 kVA 2 37 26 18.5 13 400-480V Harmonic
measurements x
15001i/iX1 15 kVA 1 111 78 55.5 39 208-240V
Waveform
15001i/iX-4001 15 kVA 1 111 78 55.5 39 400-480V
acquisition x
15003i/iX1 15 kVA 3 37 26 18.5 13 208-240V
Programmable
15003i/iX-4001 15 kVA 3 37 26 18.5 13 400-480V Impedance x
Note (1): Supplied with System Interface cable(s). Controller in master unit only.
Note (2): All input voltage specifications are for Line to Line three phase except 3001iX and 9003iX which require single phase input only.
IEEE / RS232 x x
Note (3): For 10002iX split phase system specifications, refer to 5001iX for each phase.
Model -LNS Internal AC Line Sync.
Refer to table shown for model num- -MODE-iX Switches between 1 and 3 phase
bers and configurations. output modes, for 9003iX or
15003i/iX only.
Supplied with OMNI-1-18i Impedance matching network for
User Manual, Programming Manual, For lower power needs, inquire about the Com-
single phase 3001i/iX or 5001i/iX pact iX. The Compact iX features a spacesaving 2U
Software (all on CD ROM) and RS232C to support IEC-1000-3-3 flicker rackmountable chasis in both 750 VA and 1500 VA
serial cable. tests. configurations.
Options OMNI-3-18i Impedance matching network for Combination AC and DC Power Source and Power
three phase 9003iX or 15003i/iX Analyzer
-160 RTCA/DO-160D and EUROCAE Replace multiple instruments with a single
test firmware. Refer to -160 option systems to support IEC-1000-3-3
multifunction unit
data sheet for details. flicker tests.
OMNI-3-37i Impedance matching network Single & Three Phase Operation
-400 400-480 Volt Line to Line AC in- Maximum output of 750VA per phase
put. for three phase 30003i/iX systems
to support IEC-1000-3-3 flicker 750 VA to 2250 VA of Output Power
-411 IEC61000-4-11 test firmware. See Cost effective power source
also EOS1/3. tests.
-413 IEC61000-4-13 Harmonics and -RMS Rackmount Slides. Arbitrary Waveform Generation
-WHM Watt-Hour Measurement option. User defined voltage waveform and distortion
Interharmonics test firmware. programming (C-iX Series)
-704 Mil Std 704D/E test firmware. -XLS External AC Line Sync adaptor.
Refer to -704 option data sheet (-LNS and XLS are mutually exclu- Built-in Digital Power Analyzer
sive) Analyze frequency and time domain of both
for details. voltage and current (C-iX Series)
-EOS-1 IEC61000-4-11 Electronic Output -ABD Airbus Directive 0100.1.8 tests.
-B787 Boeing 787B3-0147 tests. Rackmountable
Switch (1 phase) Includes -411 Space-saving rackmount chassis
option. Refer to EOS data sheet Cabinets Multi box iX Series II systems can
for details. be factory installed and wired in Scope Capture Capability
19 inch cabinets. Cabinet con- Built in voltage and current waveform acquisition
-EOS-3 IEC61000-4-11 Electronic Output capability (C-iX Series)
Switch (3 phase) Includes -411 figurations can be ordered by
option. Refer to EOS data sheet preceeding the model number Powerful Programing Software
with a "C1-C4" prefix. Contact fac- Powerful, yet easy to use, instrument control
for details. software included (Avionics test software also
-LKM Clock/Lock Master tory for pricing and details.
available)
-LKS Clock/Lock Auxiliary
Constant Power Mode
Provides increased current at reduced voltage to
maximize efficiency
9689 Towne Centre Drive, San Diego, CA 92121-1964 l (858) 677-9040 l FAX: (858) 677-0940 l sales@calinst.com
CompuMess Elektronik
© Copyright 2007, California Instruments Corp. GmbH · Lise-Meitner-Str. 1 ·notice
Specifications subject to change without D-85716 Unterschleißheim
Printed in the USA iXSII 07/07
www.california-instruments.com