  Digibridge                                         1715 LCR Meter
                                          Passive Component Tester with RS-232
USES:                                     Introduction
· Production Testing of LCR Components    The QuadTech 1715 LCR Digibridge is an economically priced, easily pro-
                                          grammable, micro-processor based passive component tester for fast, precise
· Component Design and Evaluation
                                          impedance measurements. It measures eleven different parameters with a
· Component Inspection and Screening      basic accuracy of 0.2%. The 1715 provides four different test frequencies and
· Quality Assurance Testing               two test voltages for testing flexibility. For remote operation and automated
                                          system applications, the unit comes standard with an RS-232 interface, plus
· Material Testing                        IEEE-488 and Handler interfaces are available as options.
                                          Description
FEATURES:                                 11 Measurement Parameters: Measure and display two of 11 impedance
· 11 Measurement Parameters               parameters simultaneously with a basic accuracy of 0.2%. Test results can also
                                          be shown as Deviation or %Deviation from nominal value.
· 0.2% Basic Measurement Accuracy
                                          Test Signal: Four user selectable test frequencies: 100Hz, 120Hz, 1kHz, and
· Selectable Test Frequency: 100Hz,       10kHz (9.6kHz), plus two test voltages: 0.25V and 1.0V.
  120Hz, 1kHz, 10kHz                      Display: 40 x 4 character LCD display for user friendly menu programming
· Selectable Test Voltage: 0.25V, 1.0V    and measured results. Simultaneously displays primary/secondary values and
· Measurement Speed to 25/second          binning or comparator results.
                                          Setup Storage: Storage of 1 setup with parameter, binning and compare infor-
· LCD Display                             mation saved. Auto recall of this setup upon power-up of instrument.
· RS-232 Interface Standard               Auto Ranging: For maximum measurement integrity the 1715 automatically
· IEEE and Handler Interfaces, Optional   selects the appropriate measurement range and instrument source impedance
· Automatic Open/Short Zeroing            (25, 100, 1k, 10k or 100k), dependent upon the impedance of the
                                          device under test.
· Automatic Hi/Lo Comparator Limits       Pass/Fail Testing: The 1715 provides a programmable Hi/Lo comparator
· Pass/Fail Sorting (8 Bins)              function as well as eight sorting bins for categorization of components.
· Storage of 1 Setup                      Open/Short Zeroing: Reduce the effects of error sources (stray
                                          admittance/residual impedance) between the device under test and the cali-
· Automatic Source Impedance Selection
                                          brated connection to the measuring instrument.
· Keypad Lockout                          Interfaces: For remote control, or adaption to a production environment, the
· Easy Operation                          1715 instrument includes an RS-232 interface as standard equipment. Both an
· Programmable Delay Time                 IEEE-488 and a Handler interface are available as optional interfaces.

                                                                     CompuMess Elektronik GmbH
                                                                       Lise-Meitner-Str.1, 85716 Unterschleissheim
                                                                          Tel 089-321501-0 Fax 089-321501-11
                                                                     http://www.compumess.de oder http://www.netzteile.de




                                                    HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS
1715 Digibridge LCR Features
Parameter                 Measurement Range                          Interfaces (Standard): RS-232
    Ls, Lp                0.01uH   - 9.999kH                         Interfaces (Optional):   IEEE-488 and Handler
    Cs, Cp                0.01pF   - 99.999mF
    Rs, Rp                0.1m     - 99.99M                          Test Terminals:          Front: 4 BNC & 1 GND
    D                     0.0001   - 9999                            Dimensions:              (w x h x d): 12.5 x 4.0 x 8.0in
    Q                     0.0001   - 9999                                                     (312.5 x 100.0 x 200.0mm)
    lZl                   0.1m     - 99.99M
                                                                     Weight:                  10.85 lbs. (5kg) net, 15.2 lbs. (7kg) ship
    Xs                    0.1m     - 99.99M
    Phase                 -180.00° - +180.00°                        Environmental:           Specifications: +15°C to +35°C, 75% RH
Accuracy:                  Basic LCR         ±0.2%                                            Operating: 10°C to +40°C
                                                                                              Storage: 0°C to +50°C
                           Basic DQ          ±0.002
                                                                                              Humidity: 10 -90% RH
Test Frequency:            100Hz, 120Hz, 1kHz and 10kHz (9.6kHz)                              Pollution Degree 2
                               Accuracy ±0.2%                                                 Installation Category II
Test Signal:              0.25V or 1.0V RMS, ±(10% + 3mV)            Power:                   · 90 - 125V AC
Measurement Speed:        Fast:        25 measurements/second                                 · 190 - 250V AC
                          Medium:      8 measurements/second                                  · 48 - 62 Hz
                          Slow:        2 measurements/second                                  · 50VA max
Measurement Mode:         Continuous or Trigger
Trigger:                  Internal
                          External (IEEE or Handler)
                          Manual
Trigger Delay:            0 - 9999ms
Ranging:                  Automatic or Hold Range
Equivalent Circuit:       Series or Parallel
Zeroing:                   Open and short circuit compensation
Source Impedance:         25, 100, 1k, 10k or 100k
                          (based on DUT impedance)
Comparator:               Primary Nominal
                          Primary Hi/Lo Limits (Value or %)
                          Secondary Hi/Lo Limuts (Value or %)
Indication:               Audible Alarm programmable: HI LO or
                          OFF for Pass or Fail Result
Binning:                  Primary Nominal
                          Primary Hi/Lo Limits (8 Bins in %)
                          Secondary Hi/Lo Limits (Value)
Display:                  40 by 4 Character LCD display
Results Format:           Value
                          Deviation
                          % Deviation
                          Bin Number
Setup Storage:            Store 1 setup, Auto recall upon power-up
Lock:                     Keypad Lockout

 Ordering Information
1715 LCR Digibridge                                                  Optional Accessories:
                                                                               B/A Calibration Data
Includes:
                                                                     1700-01   Axial/Radial Component Test Fixture
QuadTech P/N          Item
                                                                     1700-02   Axial/Radial Remote Test Fixture
150710                1715 Instruction Manual
                                                                     1700-03   Lead Set: 4 BNC Connectors to 2 Kelvin Clips
1700-03               Lead Set: 4 BNC Connectors to 2 Kelvin Clips
                                                                     1700-04   Lead Set: 4 BNC Connectors to 4 Banana Plugs
4200-0300             AC Power Cord
                                                                     1700-05   Lead Set: 4 BNC Connectors to Chip Component Tweezers
520023/520025         Power Line Fuses (0.630A 250V/0.315A 250V)
                                                                     7000-07   Low Voltage Chip Component Test Fixture
                      Calibration Certificate Traceable to NIST
                                                                     700171    IEEE-488 & Handler Interface




                CompuMess Elektronik GmbH
                  Lise-Meitner-Str.1, 85716 Unterschleissheim
                     Tel 089-321501-0 Fax 089-321501-11
               http://www.compumess.de oder http://www.netzteile.de
   Digibridge                          1730T LCR Meter
USES:                                  with Turns Ratio for Transformer Testing
· Transformer/Coil Testing             Introduction
· Production Testing                   The QuadTech 1730T LCR Meter is an easily programmable, microprocessor-
                                       based high performance passive component tester with excellent low impedance
· Component Screening                  measurement and fast measurement speeds even at 100/120Hz. It measures 16 dif-
· Material Testing                     ferent parameters with a basic accuracy of 0.1%. Programmable test frequency,
                                       speed, mode and voltage permits great versatility in component and materials test-
· Quality Assurance Testing            ing. Standard IEEE-488 and Handler interfaces provide data transfer and compo-
· Component Design and Evaluation      nent automation applications for faster production line testing.

FEATURES:                              Description
                                       14 Measurement Parameters: Measure and display two of 16 impedance param-
· 14 Measurement Parameters            eters simultaneously with a basic accuracy of 0.1%. Test results can also be shown
· 0.1% Basic Measurement Accuracy      as Deviation or %Deviation from nominal value.
                                       10 Test Frequencies: User selectable: 50Hz, 60Hz, 100Hz, 120Hz, 1kHz, 10kHz,
· 10 Test Frequencies: 50Hz-100kHz     20kHz, 40kHz, 50kHz and 100kHz.
· Test Voltage: 10mV-1V                Transformer Testing: Use the Turns Ratio function to measure the DCR of the pri-
· Measurement Speeds up to 62/second   mary coil, N and 1/N for both coils, and M the mutual inductance between the coils.
                                       LCD Display: Large, high resolution graphic display for clear, concise results and
· Monitor DUT Voltage and Current      user friendly menu interface for easy programming.
· Large High Res. Graphic Display      Setup Store/Recall: Store/Recall up to 50 test setups from instrument memory.
· IEEE and Handler Interfaces, Std.    Lock out front panel to ensure procedures are run the same way every time.
                                       Programmable Source Impedance: Program tester impedance to 100,
· Automatic Open/Short Zeroing         100/25, 25 or 10 constant current when comparing results with measurements
· Pass/Fail Binning (8)                made on other testers. Measurement results can vary significantly based solely on
                                       the source impedance of the tester being used.
· Storage/Recall of 50 Setups
                                       Voltage/Current Monitoring: Use the test signal monitoring function to display
· Measurement Averaging (1-256)        the actual voltage level across and current through the DUT.
· Measurement Delay (0-10 seconds)     DC Bias: External DC Bias voltage from 0-5V, permits capacitors to be tested
                                       under real DC bias conditions. Measurement of External DC Bias Current to 20A
· Programmable Source Impedance        is possible using the 1730T and the 1320 Bias Current Source.
· DC Bias Voltage (0-5V)               Open/Short Zeroing: Reduce the effects of error sources (stray admittance/resid-
· Keypad Lockout                       ual impedance) between the DUT and the calibrated connection to the measuring
                                       instrument.
· Easy Operation                       Fast Testing: Select from one of three measurement speeds: slow, medium or fast
· Cable Correction 0-4M                for varying degrees of accuracy and speed up to 62 measurements/second.
· Turns Ratio & DCR Measurements       Interfaces Standard: The 1730T unit includes IEEE-488 and Handler interfaces
                                       for adaption into production environments or remote control applications.
                                                                                               For more detailed
                                                                                              specifications, visit
                                                                                     www.quadtech.com
                                                                              For more information about special
                                                                              purchase, rent & lease options, call
                                                                                                  1-800-253-1230
                                                                                             Fax 1-978-461-4295
                                                                                            Intl. 1-978-461-2100




                                         HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS · CABLE TESTERS
                         1730T LCR Meter Features
Product Specifications
                         Parameter                Measurement        Range                          Measurement Delay:           Programmable from 0-10s in 1ms steps
                             Ls, Lp               0.001uH -          99.999kH                       Averaging:                  Programmable from 1-256
                             Cs, Cp               0.001pF -          9.9999F
                             Rs, Rp               0.01m -            99.99M                         Binning:                    Eight pass bins for LCR
                             L2A, L2B             0.001uH -          99.999kH                                                   One fail bin, LCR and DQ
                             D                    0.0001   -         9999                           Interfaces:                  IEEE-488 and Handler Standard
                             Q                    0.0001   -         9999
                                                                                                    Equivalent Circuit:         Series or Parallel
                             ESR                  0.01m -            99.99M
                             lZl                  0.01m -            99.99M                         Zeroing:                     Open and short circuit compensation.
                             Xs                   0.01m -            99.99M                         Buzzer:                     Programmable: HI LO or OFF for Pass or
                             Phase                -180.00° -         +180.00°                                                   Fail Result
                             M                    0.001µH -          99.999kH
                                                                                                    Comparator:                 Set Upper/ Lower Limit as absolute value
                             N                    0.001T   -         9999.9
                                                                                                                                or % value for primary and secondary
                             1/N                  0.001T   -         9999.9
                                                                                                                                parameters
                             DCR                  0.01m -            99.99M
                                                                                                    Test Signal:                Monitor voltage across and current
                         Accuracy:                Basic LCR           ±0.1%.                                                    through device under test.
                                                  Basic DQ            ±0.001                        Test Terminals:             Front: 4 BNC & 1 GND
                         Test Frequencies:        100Hz, 120Hz, 1kHz, 10kHz, 20kHz,                                             Rear: EXT BIAS: 1 BNC, 2-terminal strip
                                                  50kHz and 100kHz                                  Setup Storage:              Store/Recall 50 test setups
                                                  Accuracy ±(0.01% ± 0.01Hz)                        Lock:                       Keypad Lockout
                         Measurement Speed:       Fast:         62 measurements/second              Dimensions:                 (w x h x d): 12.5 x 4.0 x 13.5in
                                                  Medium:       19 measurements/isecond                                         (312.5 x 100.0 x 337.5mm)
                                                  Slow:         3 measurements/second
                                                                                                    Weight:                      12.4 lbs. (5.7kg) net, 17 lbs. (7.8 kg) ship
                         Ranging:                 Automatic or User Selectable
                         Trigger:                 Internal                                          Enviromental:               Specifications: +15°C to +35°C, 75% RH
                                                  External (IEEE or Handler)                                                    Operating: 0°C to +50°C
                                                  Manual                                                                        Storage: -40°C to +75°C
                                                                                                                                Humidity: <85%
                         Source Impedance:        100/25, 100, 25 or 10c.c. (constant
                                                  current)                                          Power:                      · 90 - 125V AC
                                                                                                                                · 190 - 250V AC
                         AC Test Signal:          Voltage: 10mV - 1V, 10mV steps                                                · 50 or 60 Hz
                         DC Bias Voltage:         External: 0-5V                                                                · 65W max
                                                  Internal: 0-5V with RBIAS external resistor
                         DC Bias Current:         Internal: 0-200mA with RBIAS ext resistor
                                                  External: 0-20A with 1320 BC Source;
                                                  External: 0-100A with 1320 BC Source
                                                  and four 1320S Slave Units
                         Display:                 240 by 64 LCD display
                         Results Format:          Value
                                                  Deviation
                                                  % Deviation
                                                  Bin Number
                                                  GO/NO GO



                          Ordering Information
                         1730T LCR Meter                                                            Optional Accessories:
                                                                                                                  B/A Calibration Data
                         Includes:
                                                                                                    1700-01       Axial/Radial Component Test Fixture
                         QuadTech P/N        Item
                                                                                                    1700-02       Axial/Radial Remote Test Fixture
                         4200-0300           AC Power Cord
                                                                                                    1700-03       Kelvin Lead Set: 4 BNC Connectors to 2 Kelvin Clips
                         150821              Instruction Manual
                                                                                                    7000-01       BNC to BNC Cable Lead Set (1m)
                         520026/520138       Power Line Fuses (1.0A 250V/0.5A 250V)
                                                                                                    7000-02       BNC to BNC Cable Lead Set (2m)
                         1700-03             Kelvin Test Leads: 4 BNC to 2 Kelvin Clips
                                                                                                    7000-05       Lead Set: 4 BNC Connectors to Chip Component Tweezers
                         630166              Transformer Test Fixture: 4 BNC to 4 Alligator Clips
                                                                                                    7000-07       Low Voltage Chip Component Test Fixture
                                             Calibration Certificate Traceable to NIST
                                                                                                    630166        Transformer Test Fixture: 4 BNC to 4 Alligator Clips




                                      For more detailed specifications, visit www.quadtech.com
                                For more information about special purchase, rent & lease options, call Unterschleissheim
                                  CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716
                                       Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
                                    1-800-253-1230 · Fax 1-978-461-4295 · Intl. 1-978-461-2100
                                                          5 Clock Tower Place, 210 East, Maynard MA 01754                                                       P/N 030148/A1
                                          1910 Inductance Analyzer
USES:                                        A High Performance Tester
· Production Testing of Inductive            Introduction
  Components & Wire Wound Devices            The 1910 is a production oriented, Inductance Analyzer, with a programmable frequency
· AC Impedance & DC Resistance               range of 20Hz to 1MHz, designed to perform inductance and other impedance measure-
  Measurements on a Wide Variety of          ments on coils and wire wound devices. The unit is also capable of measuring DC resist-
                                             ance between 0.1m and 100k and provides voltage or current monitoring to the test
  Components                                 device. The 1910 also provides a sequence testing mode, allowing up to 6 unique tests
· Inductance Measurements Under DC           with one single start command. With its user friendly menu programming, automatic lead
  Bias Current Conditions                    zeroing, and 5 digit measurement resolution the instrument provides fast and accurate
                                             test results.
· Component Design & Evaluation
                                             Description
                                             20 Measurement Parameters Measure and display any two of 15 impedance param-
FEATURES:                                    eters simultaneously, with a basic accuracy of 0.1%. Additionally the 1910 can measure
· 20 Measurement Parameters                  the DC resistance, or display the current through or voltage across a test device ensuring
                                             the operator of the real test conditions.
· Frequency Range 20 Hz to 1MHz
                                             Wide Frequency Over 27,000 user programmable test frequencies to fully characterize
· 0.1% Basic Measurement Accuracy            devices over the range of 20 Hz to 1 MHz.
· Measurement Speeds Up to 40/sec            Automatic Test Sequencing For increased productivity and throughput the 1910 can
· DC Resistance Measurements                 perform up to six different tests in sequence with a single push of the start button. Each
                                             test can have different measurement parameters, test conditions and limits.
· Monitoring of DUT Voltage and
                                             DC Current Bias The instruments programmable internal DC bias current source from
  Current                                    1 mA to 1 Amp allows components to be tested under real operating conditions. This can
· 5 Digit Measurement Resolution             be expanded to 100 Amps when the 1910 is used in conjunction with QuadTech's Model
                                             1320 Bias Current Source and slave units
· Constant Voltage (Voltage Leveling)
                                             Setup Storage/Recall The test operator has the ability to store and recall, from inter-
· IEEE-488, RS-232 & Handler                 nal memory, up to 30 single test setups and 10 sequential setups (six tests in sequence).
  Interfaces, all Standard                   The front panel can be locked out, with password protection, to ensure that procedures
                                             are run the same way every time.
· Open/Short Zeroing & Cable
  Compensation                               Load Correction Substantially improves instrument accuracy by allowing the operator
                                             to specify the value of a known standard, measure it, and apply a correction to ongoing
· Load Correction                            measurements.
· 14 Pass/Fail Bins                          Programmable Source Impedance The operator is able to set instrument source
                                             impedance at 5, 25, 50 or 100 ohms, an important feature when comparing measure-
· Keypad Lockout
                                             ments to those made on other another testers. Measurement results can vary substantial-
                                             ly based solely on the source impedance of the tester being used.




                                                                                  CompuMess Elektronik GmbH
                                                                                   Lise-Meitner-Str.1, 85716 Unterschleissheim
                                                                                      Tel 089-321501-0 Fax 089-321501-11
                                                                                 http://www.compumess.de oder http://www.netzteile.de




                                        HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS · TRANSFORMER TESTERS
                         1910 Inductance Analyzer
Product Specifications
                          Parameter                          Measurement Range                                                Basic Measurement Accuracy*
                                                                                                                    Low                 Medium                              High
                          Ls,Lp                              0.001nH to 99.999H                                  +/-0.5%                   +/-0.25%                     +/-0.1%
                          Cs,Cp                              0.01pF to 9.9999F                                   +/-0.5%                   +/-0.25%                     +/-0.1%
                          DF                                 0.00001 to 99.999                                   +/-0.005                  +/-0.0025                    +/-0.001
                          Q                                  0.00000 to 9999.9                                   +/-0.005                  +/-0.0025                    +/-0.001
                          Y,Gp,Bp                            10 nS to 9999.9 S                                   +/-0.5                    +/-0.25                      +/-0.1
                          |Z|,Rs,Rp,Xs,ESR                   0.00001 m to 99.999M                                +/-0.5%                   +/-0.25%                     +/-0.1%
                          Phase Angle                        -180.00 to +179.99 degrees                          +/-1.8°                   +/-0.9°                      +/-0.18°
                          DCR                                0.1m to 100K                                        +/-0.5%                   +/-0.25%                     +/-0.1%
                          DUT AC Voltage                     20mV to 1.0V                                                          +/-(2% + 5mV)@1kHz
                          DUT AC Current                     1µA to 150mA                                                          +/-(2% + 5µA)@1kHz
                          DUT DC Voltage                     20mV to 1.0V                                                             +/-(2% + 5mV)
                          DUT DC Current                     1µA to 150mA                                                             +/-(2% + 5µA)
                                                                                                     *At optimum test signal levels, frequencies, DUT values and without calibration uncertainty.


                         Test Frequency:                Range: 20Hz to 1MHz, Continuous                          Measurement Delay:               Programmable from 0 to 1000 ms in 1ms steps
                                                        Resolution: 1Hz from 20Hz to 1KHz,
                                                                                                                 Averaging:                       Programmable from 1 to 1000
                                                        4 digits >1KHz
                                                        Accuracy: +/- (0.02% +0.02 Hz)                           Median Value:                    Averaged over last three measurements

                         Measurement Speed:             Speed                  Accuracy Setting                  Setup Storage:                   30 Single Tests
                                                        40 meas/sec            Low, No Display                                                    10 Sequential (6 tests in each)
                                                        25 meas/sec            Low
                                                                                                                 Other:                           Constant Voltage Mode (voltage leveling)
                                                        10 meas/sec            Medium
                                                                                                                                                  Cable Compensation (1M, 2M, no cable)
                                                             1 meas/sec        High
                                                                                                                                                  Open/Short Zeroing
                         Ranging:                       Automatic, Range Hold or user selectable                                                  Distortion Check

                         Trigger:                       Internal (automatic)                                     Calibration·:                    Recommended interval 1 year
                                                        External (via RS-232,IEEE-488.2 or Handler                                                NIST traceable calibration
                                                        interfaces)                                                                               Built-in automatic calibration procedure
                                                        Manual
                                                                                                                 Usage & Cal Data:                Displays last calibration date, standard values used
                         Source Impedance:              5, 25, 50, 100                                                                            in calibration

                         AC Test Signal:                Voltage: 20mV to 1.0V (open circuit) in 5 mV             Self Test:                       Verifies critical instrument operation at power-up or
                                                        steps                                                                                     when selected from menu

                         DC Test Signal:                Voltage: 20mV to 1.0V (open circuit) in 5 mV             Test Terminals:                  Front panel, four terminal (BNC)
                                                        steps, 5  source impedance                                                                Optional Test Fixtures Available

                         DC Bias Current:               Internal: 0.0 to 1.0A in 1mA steps                       Mechanical:                      Bench mount with tilt bail
                                                        External: 0.0 to 20.0A in 5mA steps using                                                 Rack mount kit optional
                                                        QuadTech 1320
                                                                                                                 Dimensions:                      (w x h x d): 17 x 5.25 x 16in (432 x 133 x 406 mm)
                         Display:                       LCD Display with backlight
                                                                                                                 Weight:                          15lbs (8kg) net, 21lbs (9.9kg) shipping
                                                        Pass/Fail and status indicators
                                                                                                                 Enviromental:                    Meets MIL-28800E, Type 3, Class 5, Style E & F
                         Results Formats:               Engineering or scientific format
                                                                                                                                                  Operating: 0° to +50°C
                                                        %Deviation from nominal of primary parameter
                                                                                                                                                  Humidity: < 75% for 11° to 30°C operating
                                                        Deviation from nominal of primary parameter
                                                                                                                                                  Storage: -40° to +71° C
                                                        Pass/Fail
                                                        No Display Mode for maximum throughput                   Power:                           100-240 VAC         50/60 Hz
                                                                                                                                                  100 W Max
                                                        Sequencing:
                                                        Displays up to 6 sequential test results, primary
                                                        and/or secondary                                                                            CompuMess Elektronik GmbH
                                                                                                                                                      Lise-Meitner-Str.1, 85716 Unterschleissheim
                         Standard Interfaces:           IEEE-488.2, RS-232, Handler                                                                      Tel 089-321501-0 Fax 089-321501-11
                         Ordering Information                                                                                                      http://www.compumess.de oder http://www.netzteile.de

                         1910 Inductance Analyzer                         1700-02     Axial/Radial Remote Test Fixture
                         Includes:                                        1700-03     4 BNC Connectors to 2 Kelvin Clips Lead Set
                         4200-0300 AC Power Cord                          1700-04     4 BNC Connectors to 4 Banana Plugs
                         150491    Instruction Manual                     1700-05     4 BNC Connectors to Chip Component Tweezers
                                   Calibration Certificate                2000-16     Rack Mount Flanges
                                     Traceable to NIST                    7000-01     BNC to BNC Cable Set (1M)
                                                                          7000-02     BNC to BNC Cable Set (2M)
                         Optional Accessories
                                                                          7000-07     Low Voltage Chip Component Test Fixture
                         1700-01     Axial/Radial Component
                                       Test Fixture



                                                                                                                                                                                                          PN030091/A1
                      Series              1920 Precision LCR Meter
USES:                                       High Performance Testing to 1 MHz
· Fast Production Testing of LCR            Introduction
  Components and Materials                  The 1920 is a high performance LCR Meter designed to perform fast, automated
· AC Impedance & DC Resistance              impedance measurements on a variety of electronic components and materials. The
  Measurements                              instrument has a basic accuracy specification of 0.1% for accurate test results over
                                            a wide frequency range, from 20 Hz to 1 MHz. Besides 15 impedance parameters the
· Component Characterization Over           1920 is also capable of measuring DC resistance as well as monitoring the voltage
  a Wide Frequency Range                    across or current through the device under test. The unit incorporates a distinctive
                                            sequence test mode, allowing up to 6 uniquely different tests to be performed quickly
· Component Screening, Evaluation           on a single start command. Additionally, the 1920 includes IEEE-488, RS-232, and
  and Design                                handler interfaces, all standard.

                                            Description
FEATURES:                                   20 Measurement Parameters Measure and display any two of 15 impedance
· 20 Measurement Parameters                 parameters simultaneously, with a basic accuracy of 0.1%. Additionally the 1920 can
                                            measure the DC resistance, or display the current through or voltage across a test device
· Frequency Range 20 Hz to 1MHz             ensuring the operator of the real test conditions.
· 0.1% Basic Measurement Accuracy           Wide Frequency Over 27,000 user programmable test frequencies to fully characterize
· Measurement Speeds Up to 40/sec           devices over the range of 20 Hz to 1 MHz.

· DC Resistance Measurements                Automatic Test Sequencing For increased productivity and throughput the 1920 can
                                            perform up to six different tests in sequence with a single push of the start button. Each
· Monitoring of DUT Voltage and             test can have different measurement parameters, test conditions and limits.
  Current                                   DC Bias Voltage The instruments internal DC bias voltage source, programmable from
· 5 Digit Measurement Resolution            0 to 2 V in 1 mV steps, allows capacitors to be tested under real DC bias conditions.

· Programmable DC Bias Voltage, 0-2V        Setup Storage/Recall The test operator has the ability to store and recall, from
                                            internal memory, up to 30 single test setups and 10 sequential setups (six tests in
· Constant Voltage (Voltage Leveling)       sequence). The front panel can be locked out, with password protection, to ensure that
· IEEE-488, RS-232 & Handler                procedures are run the same way every time.
  Interfaces, all Standard                  Load Correction Substantially improves instrument accuracy by allowing the operator
                                            to specify the value of a known standard, measure it, and apply a correction to ongoing
· Open/Short Zeroing & Cable                measurements.
  Compensation
                                            Programmable Source Impedance The operator is able to set instrument source
· Load Correction                           impedance at 5, 25, 50 or 100 ohms, an important feature when comparing measure-
                                            ments to those made on other testers. Measurement results can vary substantially based
· 14 Pass/Fail Bins
                                            solely on the source impedance of the tester being used.
· Keypad Lockout
                                                                             CompuMess Elektronik GmbH
                                                                               Lise-Meitner-Str.1, 85716 Unterschleissheim
                                                                                  Tel 089-321501-0 Fax 089-321501-11
                                                                            http://www.compumess.de oder http://www.netzteile.de

                                                                                      F




                                        HIPOT & GROUND BOND TESTERS · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS
                         1920 Precision LCR Meter
Product Specifications
                          Parameter                          Measurement Range                                                Basic Measurement Accuracy*
                                                                                                                    Low                    Medium                                    High
                          Ls,Lp                              0.001nH to 99.999H                                  +/-0.5%                      +/-0.25%                           +/-0.1%
                          Cs,Cp                              .01pF to 9.9999F                                    +/-0.5%                      +/-0.25%                           +/-0.1%
                          DF                                 0.00001 to 99.999                                   +/-0.005                     +/-0.0025                          +/-0.001
                          Q                                  0.00000 to 9999.9                                   +/-0.005                     +/-0.0025                          +/-0.001
                          Y,Gp,Bp                            10 nS to 9999.9 S                                   +/-0.5%                      +/-0.25%                           +/-0.1%
                          |Z|,Rs,Rp,Xs,ESR                   0.00001 m to 99.999M                                +/-0.5%                      +/-0.25%                           +/-0.1%
                          Phase Angle                        -180.00 to +179.99 degrees                          +/-1.8°                      +/-0.9°                            +/-0.18°
                          DCR                                0.1m to 100K                                        +/-0.5%                      +/-0.25%                           +/-0.1%
                          DUT AC Voltage                     20mV to 1.0V                                                              +/-(2% + 5mV)@1kHz
                          DUT AC Current                     1µA to 150mA                                                              +/-(2% + 5µA)@1kHz
                          DUT DC Voltage                     20mV to 1.0V                                                                  +/-(2% + 5mV)
                          DUT DC Current                     1µA to 150mA                                                                  +/-(2% + 5µA)
                                                                                                     *At optimum test signal levels, frequencies, DUT values and without calibration uncertainty.


                         Test Frequency:                Range: 20Hz to 1MHz, Continuous                          Measurement Delay:               Programmable from 0 to 1000 ms in 1ms steps
                                                        Resolution: 1Hz from 20Hz to 1KHz,
                                                                                                                 Averaging:                       Programmable from 1 to 1000
                                                        4 digits >1KHz
                                                        Accuracy: +/- (0.02% +0.02 Hz)                           Median Value:                    Averaged over last three measurements

                         Measurement Speed:             Speed                  Accuracy Setting                  Setup Storage:                   30 Single Tests
                                                        40 meas/sec            Low, No Display                                                    10 Sequential (6 tests in each)
                                                        25 meas/sec            Low
                                                                                                                 Other:                           Constant Voltage Mode (voltage leveling)
                                                        10 meas/sec            Medium
                                                                                                                                                  Cable Compensation (1M, 2M, no cable)
                                                             1 meas/sec        High
                                                                                                                                                  Open/Short Zeroing
                         Ranging:                       Automatic, Range Hold or user selectable                                                  Distortion Check

                         Trigger:                       Internal (automatic)                                     Calibration:                     Recommended interval 1 year
                                                        External (via RS-232,IEEE-488.2 or Handler                                                NIST traceable calibration
                                                        interfaces)                                                                               Built-in automatic calibration procedure
                                                        Manual
                                                                                                                 Usage & Cal Data:                Displays last calibration date, standard values used
                         Source Impedance:              5, 25, 50, 100                                                                            in calibration

                         AC Test Signal:                Voltage: 20mV to 1.0V (open circuit) in 5 mV             Self Test:                       Verifies critical instrument operation at power-up or
                                                        steps                                                                                     when selected from menu

                         DC Test Signal:                Voltage: 20mV to 1.0V (open circuit) in 5 mV             Test Terminals:                  Front panel, four terminal (BNC)
                                                        steps, 5  source impedance                                                                Optional Test Fixtures Available

                         DC Bias Voltage:               Internal: 0 to 2V in 1mV steps                           Mechanical:                      Bench mount with tilt bail
                                                                                                                                                  Rack mount kit optional
                         Display:                       LCD Display with backlight
                                                        Pass/Fail and status indicators                          Dimensions:                      (w x h x d): 17 x 5.25 x 16in (432 x 133 x 406 mm)

                         Results Formats:               Engineering or scientific format                         Weight:                          15lbs (8kg) net, 21lbs (9.9kg) shipping
                                                        %Deviation from nominal of primary parameter
                                                                                                                 Enviromental:                    Meets MIL-28800E, Type 3, Class 5, Style E & F
                                                        Deviation from nominal of primary parameter
                                                                                                                                                  Operating: 0° to +50°C
                                                        Pass/Fail
                                                                                                                                                  Humidity: < 75% for 11° to 30°C operating
                                                        No Display Mode for maximum throughput
                                                                                                                                                  Storage: -40° to +71° C
                                                        Sequencing:
                                                                                                                 Power:                           100-240 VAC         50/60 Hz
                                                        Displays up to 6 sequential test results, primary
                                                                                                                                                  100 W Max
                                                        and/or secondary

                         Standard Interfaces:           IEEE-488.2, RS-232, Handler
                                                                                                                                                  CompuMess Elektronik GmbH
                                                                                                                                                     Lise-Meitner-Str.1, 85716 Unterschleissheim
                         Ordering Information                                                                                                           Tel 089-321501-0 Fax 089-321501-11
                                                                                                                                                  http://www.compumess.de oder http://www.netzteile.de
                         1920 Precision LCR Meter                         1700-02     Axial/Radial Remote Test Fixture
                         Includes:                                        1700-03     4 BNC Connectors to 2 Kelvin Clips Lead Set
                         4200-0300 AC Power Cord                          1700-04     4 BNC Connectors to 4 Banana Plugs
                         150566    Instruction Manual                     1700-05     4 BNC Connectors to Chip Component Tweezers
                                   Calibration Certificate                2000-16     Rack Mount Flanges
                                     Traceable to NIST                    7000-01     BNC to BNC Cable Set (1M)
                                                                          7000-02     BNC to BNC Cable Set (2M)
                         Optional Accessories
                                                                          7000-07     Low Voltage Chip Component Test Fixture
                         1700-01     Axial/Radial Component
                                       Test Fixture



                                                                                                                                                                                                          PN030092/A1
               Series                      7000 Precision LCR Meters
USES:                                          CE Marked Impedance Analyzer
· Production Testing of LCR
  Components                                   Introduction
                                               The 7000 LCR Meter is designed to perform precision impedance measurements over a
· Frequency Response Characterization          wide frequency range, 10 Hz to 500 kHz for the model 7400 and 10 Hz to 2 MHz for the
  of Components                                7600. The instrument is capable of measuring 14 different parameters with 0.05% accu-
· Component Screening                          racy to meet today's requirements for component and material testing. The ease of use
                                               and user friendly menu programming makes the 7000 Series ideal for applications in
· Material Testing                             product development, incoming inspections, or production line testing.
· Quality Assurance Testing
                                               Description
· Measuring Dielectric Constant Using          14 Different Impedance Parameters Measure and display any two parame-
  Standard Test Cell                           ters simultaneously to achieve coverage and flexibility not previously available.

FEATURES:                                      Automatic Test Sequencing Run up to six different tests in sequence with a sin-
· Frequency Range 10 Hz to 2 MHz               gle push of the start button. Each test can have different conditions and limits.
  (7600), 500 kHz (7400)                       Swept Measurements Fast and accurate swept parameter measurements, graphi-
· 0.05% Basic Measurement Accuracy             cal or tabular, for verification of component and material response to changes in AC test
                                               frequency, AC test voltage or AC test current, without the need for complex programming
· 7 Digit Measurement Resolution               or an external controller.
· Programmable Test Voltage and
                                               Program and Data Storage Test setups can be stored and recalled from either
  Current
                                               internal memory or from standard DOS formatted 3 1/2" floppy disks. The front panel
· Auto Ranging                                 can be locked out, with password protection, to ensure procedures are run the same way
· Test Setup and Measurement Data              every time. Measured data can be stored on a floppy disk and then transferred to PC for
                                               data reduction and analysis.
  Storage
· Four-Terminal Kelvin Connections             Load Correction Substantially improves instrument accuracy by performing meas-
                                               urements on a known standard and applying correction to subsequent measurements.
· IEEE-488, RS-232, Handler, and               Ideal for repetitive testing of identical devices at like test conditions.
  Parallel Printer Interfaces
                                               Automatic Calibration Procedure The 7000 can be calibrated without return-
· Graphical and Tabular Display of
                                               ing the unit to the factory using the NIST traceable QuadTech Calibration Kit, reducing
  Swept Frequency, Voltage and Current         downtime and calibration costs.
  Measurements
· Sequence Testing of Up To 6                  Easy to Use Large LCD graphics display and user friendly menu driven interface allows
                                               the 7000 to be put on line fast, providing useful measurements by operators with little or
  Individual Tests
                                               no training.
· Load Correction                                                            CompuMess Elektronik GmbH
                                                                                Lise-Meitner-Str.1, 85716 Unterschleissheim
· Binning (15)                                                                     Tel 089-321501-0 Fax 089-321501-11
                                                                             http://www.compumess.de oder http://www.netzteile.de
· Built-in Automatic Calibration Routine




                                            HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS · DECADES · STANDARDS
                                                                                                                                              Result Formats:                Engineering or scientific notation
                         7400/7600
                                                                                                                                                                             % Deviation from nominal of primary parameter
                         Measured Parameters:                          Any two of 14 parameters measured and
                                                                                                                                                                             Deviation from nominal of primary parameter
Product Specifications
                                                                       displayed simultaneously, user selectable
                                                                                                                                                                             Pass/Fail
                                                                                                Basic Measurement Accuracy*                                                  Binning summary
                            Parameter                  Measurement Range                                   Speed                                                             No Display for maximum throughput
                                                                                                 Fast     Medium     Slow
                            Cs, Cp                00000.01 fF to 9.999999 F                    ±0.5%      ±0.25%    ±0.05%                    Sweep Result:                  Primary parameter vs. frequency, voltage or
                            Ls, Lp                0000.001 nH to 99.99999 H                    ±0.5%      ±0.25%    ±0.05%                                                   current
                            D                     .0000001 to 99.99999                         ±0.005     ±0.0025   ±0.0005                                                  Graphical or Tabular Format
                            Q                     .0000001 to 999999.9                         ±0.005     ±0.0025   ±0.0005                                                  Up to 200 measurement points per sweep
                            Z , Rs, Rp,           000.0001 m to 99.99999 M                     ±0.5%      ±0.25%    ±0.05%                    Sequencing Result:             Displays up to 6 sequential test results, primary
                            ESR, Xs
                                                                                                                                                                             and/or secondary
                            Y , Gp, Bp            00000.01 µS to 9.999999 MS                   ±0.5%           ±0.25%   ±0.05%
                            Phase Angle           -180.0000 to +179.9999 degrees               ±1.8°           ±0.9°    ±0.18°                AutoAcc:                       Automatic calculation and display of overall
                         *At optimum test signal levels, frequencies, DUT values and without calibration uncertainty.                                                        instrument accuracy for selected settings, test
                         Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q)                                                        conditions and device under test
                         Factors, Impedance Z , Admittance Y , Phase Angle (0 Equivalent Series Resistance (ESR),
                                                                             -),
                         Conductance (Gp), Reactance (Xs), Susceptance (Bp)                                                                   Standard Interfaces:           IEEE-488.2, RS-232, Handler, Printer Port,
                         Note: s = series, p = parallel, ESR equivalent to Rs
                                                                                                                                                                             3.5" Disk Drive
                         Test Frequency:                               7400         Range: 10 Hz to 500 kHz, continuous                       Charged Capacitor Protection: 8/C for Vmax  250 V; 2/C for Vmax  1000V
                                                                                    Resolution: 0.1 Hz from 10 Hz to                                                         C = Capacitance in farads of the device under test
                                                                                    10 kHz, 5 digits>10kHz                                                                   Additional Fuse Protection on 7400C
                                                                                    Accuracy: +/-(0.002% + 0.02Hz)
                                                                       7600         Range: 10 Hz to 2 MHz, continuous                         Measurement Delay:             Programmable from 0 - 1000 ms in 1 ms steps
                                                                                    Resolution: 0.1 Hz from 10 Hz to                          Averaging:                     Programmable from 1 - 1000
                                                                                    10 kHz, 5 digits>10kHz                                                                   Median value mode
                                                                                    Accuracy: +/-(0.25% + 0.02Hz)
                                                                                                                                              Data Storage:                  40,000 measurements/disk
                         Measurement Speed:                            7400         Basic: 40 meas/sec                                                                       ASCII format   DOS compatible
                                                                                    Enhanced: 8 meas/sec
                                                                                    Extended: 1 meas/sec                                      Program Storage:               125 per disk                          25 setups internal
                                                                       7600         Fast: 25 meas/sec                                                                        Password protected                    ASCII format
                                                                                    Medium: 8 meas/sec
                                                                                                                                              Calibration:                   Recommended Calibration Interval 1 year
                                                                                    Slow: 1 meas/sec
                                                                                                                                                                             Complete NIST Traceable Calibration using
                         Ranging:                                      Automatic, Range Hold or user selectable                                                              QuadTech 7000-09 Cal Kit
                                                                                                                                                                             Built-in automatic calibration procedure
                         Trigger:                                      Internal (automatic), External (RS-232,
                                                                       IEEE-488.2 or Handler interfaces) and Manual                           Usage & Cal Data:              Displays last calibration date, standard values
                                                                                                                                                                             used in calibration and # of hours operation
                         AC Test Signal:                               Voltage: 20 mV to 5.0 V (open circuit)
                                                                                up to 500kHz in 5 mV steps                                    Self-Test Routine:             Verifies critical instrument operation at power-
                                                                       7600     20 mV to 1.0 V (open circuit)                                                                up or when selected from menu
                                                                                500kHz-1MHz in 5 mV steps
                                                                                                                                              Contact Check:                 Time to detect, 2ms
                                                                                20 mV to 0.5 V (open circuit)
                                                                                >1MHz in 5 mV steps                                           Test Terminals:                Front panel, four terminal (BNC) guarded

                                                                       Current: 250 µA to 100 mA (short circuit) in                           Mechanical:                    Bench mount with tilt bail
                                                                       50 µA steps                                                                                           Dimensions: (w x h x d): 16 x 6 x 14in
                                                                       (Max. Compliance 3V < 500kHz.                                                                                      (410 x 150 x 360mm)
                                                                                                                                                                             Weight: 17 lbs (8kg) net, 23 lbs (10.5kg)
                         Source Impedance:                             25, 400, 6.4k, or 100k, range dependent
                                                                                                                                                                             shipping
                         DC Bias Voltage:                              Internal: 2.0 V
                                                                                                                                              Environmental:                 Meets MIL-T-28800E, Type 3, Class 5,
                                                                       External: 0 to +/-200V
                                                                                                                                                                             Style E & F
                                                                                 0 to +/-500V on 7400A/7600A
                                                                                                                                                                             Operating: 0 to + 50° C
                         Display:                                      LCD Graphics with back light and adjustable                                                           Humidity: <75% for 11 to 30° operating
                                                                       contrast                                                                                              Storage: - 40 to + 71° C

                                                                                                                                              Power:                         · 90 - 250Vac            · 47/63Hz
                                                                                                                                                                             · 100W max


                           Ordering Information
                            7400/7600 Precision LCR Meters                                        7400C       LCR Meter, Charged Capacitor         7000-01      BNC Cable Set, 1 meter
                            7600      LCR Meter                                                               Protection                           7000-02      BNC Cable Set, 2 meters
                            7600-CE LCR Meter, CE Marked                                          7400C-CE LCR Meter, Charged Capacitor            7000-03      Kelvin Clip Leads
                            7600A     LCR Meter, +/-500V External Bias                                        Protection, CE Marked                7000-04      Alligator Clip Leads
                            7600A-CE LCR Meter, +/-500V External Bias,                            Includes:                                        7000-05      Chip Component Tweezers
                                      CE Marked                                                   Instruction Manual                               7000-06      Axial/Radial Lead Component
                            7400      LCR Meter                                                   Power Cord                                                    Test Fixture
                            7400-CE LCR Meter, CE Marked                                          Calibration Certificate Traceable to NIST        7000-07      Chip Component Test Fixture
                            7400A     LCR Meter, +/-500V External Bias                            Optional Accessories:                            7000-08      High Voltage Test Fixture
                            7400A-CE LCR Meter, +/-500V External Bias,                            7000-00 Rack Mount Kit                           7000-09      Calibration Kit
                                      CE Marked

                                                                          CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
                                                                                Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
                                                                                                                                                                                                                                   PN030087/A1
         1855                           Leakage Current/IR Meter
                                        Capacitor Leakage Testing
USES:
· Production Testing of Leakage         Introduction
  Current on Electrolytic Capacitors,   The 1855 Capacitor Leakage Current/IR Meter is a fast and reliable digital
  Zener diodes, Neon lamps, and other   leakage current meter. Primarily used for electrolytic capacitor leakage cur-
  Capacitive Components                 rent testing, and aluminum-foil withstand voltage testing, the 1855 instrument
· Production Testing of Insulation      also measures insulation resistance. The 1855 instrument comes standard with
  Resistance on Switches, Connectors,   an RS-232 interface plus it has an optional IEEE-488 & Handler interface for
  and Other Devices                     high speed automated measurements. Whether its component evaluation on
· Testing Insulation Breakdown in       the production line or bench-top testing in R&D applications, the 1855
  accordance with EIAJ RC-2364A         Capacitor Leakage Current/IR Meter is right for the task.
· Component Evaluation, Testing, and    Description
  Design
                                        Leakage Current: The 1855 instrument measures and displays the undesir-
FEATURES:                               able flow of current through a capacitor or over any surface of insulation. The
                                        leakage current function provides a measurement range from 1nA to 20mA and
· 0.3% Basic Measurement Accuracy
                                        programmable test voltage from 1- 650V DC.
· Constant Current: 0.5mA ± 0.05mA      Powerful Charge Current: The 1855 Model offers flexibility with program-
  meets EIAJ RC-2364A WV test           mable charge current from 0.5mA to 500mA in increments of 0.5mA.
· Programmable DC Voltage: 1-650V       Programmable Source Voltage: The DC voltage can be programmed from
· Programmable Charge Current: 0.5-     1.0V to 100V in 0.1V increments (with current to 500mA) and from 101V to
  500mA                                 650V in 1V increments (with current to 150mA).
· Programmable Charge & Dwell           Precision Measurement: With a basic accuracy of 0.3%, the compact 1855
  Times: 0-999 seconds                  unit makes consistent, stable, and reliable test results.
· Leakage Current Measurement from:     Insulation Resistance: Serves as an economical megohmmeter by measuring
  1nA - 20.00mA                         and displaying a product's insulation resistance. This resistance can be meas-
· Insulation Resistance Measurement     ured over the range of 10 to 99G with a programmable test voltage from
  from 10 - 99G                         1.0V - 650V DC.
· Auto Ranging                          Aluminum-Foil Withstand Voltage: The 1855 meets the requirements of the
                                        EIAJ RC-2364A standard with a low programmable constant current for
· Speeds Up to 18 Measurements/second   charging aluminum electrolytic capacitors. The rise time (Tr) and final test
· Measurement Averaging (1-8)           voltage (Vf) are shown on the instrument display.
· Standard RS-232 Interface             65 Watt Discharge Circuit: For operator safety and for rapid, complete dis-
· Optional IEEE & Handler Interface     charge of large capacitors, a power discharge circuit is built into the 1855 unit.
· Front Panel Lockout
· Comparator and Pass/Fail Function
· Enhanced LCD Display




                                                                             CompuMess Elektronik GmbH
                                                                               Lise-Meitner-Str.1, 85716 Unterschleissheim
                                                                                  Tel 089-321501-0 Fax 089-321501-11
                                                                            http://www.compumess.de oder http://www.netzteile.de
1855 Leakage Current/IR Meter
Leakage Current Test                                                            Measurement Mode:             Continuous or Trigger (INT, EXT, MAN)
Measurement Range:         0.001uA - 20.0mA                                     Ranging:                      Automatic or Hold

Accuracy:                  ± (0.3% + 0.005uA)                                   Measurement Rate:             Fast    18 Measurements/second
                                                                                                              Medium: 14 Measurements/second
Test Voltage:              1.0V - 650V DC, 0.1V/Step                                                          Slow:   7 Measurements/second
Voltage Accuracy:          ± (0.5% + 0.2V)                                      Averaging:                    1-8
Test Current:              0.5mA - 500mA, 0.5mA/Step DC <100V                   Compare:                      Upper & Lower Limits for LC & IR Tests
                           0.5mA - 150mA, 0.5mA/Step DC >100V                   Indication:                   Audible Alarm, Programmable High, Low
Charge Current                                                                                                or Off for Pass or Fail
Accuracy:                  ±(3% + 0.05mA)                                       Front Panel Lockout:          Keypad Lock
                                                                                Display:                      240 x 64 LCD Graphic Display
Insulation Resistance Test
                                                                                Standard Interfaces:          RS-232
Measurement Range:         10 - 99.99G 
                                                                                Optional Interfaces:          IEEE-488 & Handler
IR Accuracy:               + 0.6% (voltage & load dependent)
                                                                                Connectors:                   1 BNC Terminal: Input
Test Voltage:              1.0V - 650V DC, 0.1V/Step                                                          2 Banana Terminals: HV(+), HV(-)
Voltage Accuracy:          ± (0.5% + 0.2V)                                                                    1 Banana Socket: Chassis Ground

Test Current:              0.5mA - 500mA, 0.5mA/step DC <100V                   Dimensions:                   (w x h x d): 12.5 x 4.0 x 13.5 inches
                                                                                                                        (317.2 x 101.5 x 342.6 mm)
                           0.5mA - 150mA, 0.5mA/Step DC >100V
                                                                                Weight:                       18lbs (8.2kg) net, 22lbs (10kg) shipping
Charge Current
Accuracy:                  ±(3% + 0.05mA)                                       Environmental:                Operation: 10°C to +40°C
                                                                                                              Storage: -10°C to +50°C
WithStand Voltage Test                                                                                        Humidity: <90%
                                                                                                              Pollution Degree 2
Rise Time (Tr):            0.05 - 120s
                                                                                                              Installation Category I
Withstand Voltage (Vf): 1.0V - 650V DC, 0.1V/Step                               Power:                        90-125V AC
Test Current:              0.5mA - 150mA, 0.5mA/Step                                                          190-250V AC
                                                                                                              50/60 Hz
Charge Current                                                                                                400 W Maximum
Accuracy:                  ±(3% + 0.05mA)
Measure Time:              30 - 600s
MAX Charge Time:           5 - 600s
General Features
Test Types:                Automatic Sequence Test
                           Manual Step Test
Null:                      Correction for Lead Leakage
Monitor Voltage (Vm): Monitor & Display Voltage across DUT
Charge Time:               0 - 999seconds
Dwell Time:                0.2 - 990seconds
Discharge:                 65 Watt Discharge Circuit
Trigger:                   Delay: 0 - 9.995seconds
                           Edge: Falling or Rising


Ordering Information
1855            Capacitor Leakage Current/IR Meter                             Optional Accessories:
                                                                               QT P/N      Description
Includes:
                                                                               N/A         Before & After Calibration Data
QT P/N          Description
                                                                               700171      IEEE-488/Handler Interface
150767          Instruction Manual
                                                                               630157      RS-232 Cable
1855-01         Test Leads: Banana to Alligator Clip & BNC to Alligator Clip
4200-0300-00    AC Power Cable
520149          Power Line Fuse 115V Operation: T4A 250V
520148          Power Line Fuse 230V Operation: T2A 250V
N/A             Calibration Certificate traceable to NIST




   CompuMess Elektronik GmbH
        Lise-Meitner-Str.1, 85716 Unterschleissheim
           Tel 089-321501-0 Fax 089-321501-11
  http://www.compumess.de oder http://www.netzteile.de
            Series                               1865 Megohmmeter/IR Tester
USES:                                                               High Resistance Meter
· Production Testing of Insulation on       Introduction
  Switches, Connectors, Cables,             The 1865 Megohmmeter/IR tester, with digital display, is used for high accuracy measurements
  Printed Circuit Boards,                   of insulation resistance on a wide variety of components, materials and electronic devices.
                                            Resistance values greater than 100 teraohms can be measured with a basic accuracy of 0.5% at
  Transformers, Motors and other
                                            DC voltages up to 1000V. A programmable limit makes the unit ideal for pass/fail testing in a
  Devices                                   production environment and is also adaptable for use in automated system applications with its
· Insulation Properties of Natural and      optional IEEE-488 interface.
  Synthetic Materials such as Oils,         Description
  Plastics, Rubber or Porcelains            Precision Measurements: The 1865 provides resistance measurement capability over a range of
                                            1k to 100T (test voltage dependent), with a basic accuracy of 0.5%. To meet the test
· Screening and Sorting of High             requirements for a wide range of devices, the test voltage is fully programmable up to 1000V
  Resistance Components                     DC. Sensing the proper resistance measurement range is done automatically, thus eliminating
                                            setup errors: and in order to eliminate lead or fixture errors, the operator can easily initiate an
· Incoming Inspection and Quality
                                            automatic zeroing routine of test leads.
  Assurance Testing
                                            Easy To Use: The 1865's menu programming system, simple controls and indicators combine
· Product Design and Evaluation of
                                            for efficiency of test and productivity. It's multi-function keypad provides the operator with an
  Insulation Properties                     easy way to program and make measurements.

                                            Pass/Fail: Measured results are automatically compared to an operator programmed limit for
FEATURES:                                   pass/fail testing. Psss/Fail indicator lights or a pass/fail output from the instruments I/O inter-
· Measurement Range to 100T                 face give a clear indication of the measurement results.
· Programmable Test Voltage from 1          Program Data Storage: Test setup conditions and measured results can be stored in instrument
  to 1000 V DC                              memory or on 3 1/2" disk when using the extended program storage option.

· 0.5% Measurement Accuracy                 Automated Testing: For automated system applications, the instrument includes an I/O inter-
                                            face connection with remote start and pass/fail outputs. An optional IEEE-488 interface is also
· Automatic Ranging                         available (or added later) which enables the 1865 to be used under computer control.
· Programmable Test Times
                                            Current Display Mode: Besides a readout of insulation resistance in ohms, the operator can
· High Resolution LCD Display               select a display mode for reading the measured current directly.
· Storage of Test Setup and Results         Safety Features: The 1865 provides safety features such as current limiting, a warning indicator
· Automatic Zeroing of Test Leads           when high voltage is active, and safety interlock, all for protection of the operator.

· Limit Entry for Pass/Fail Testing         Component Test Fixture: An accessory fixture is available for use with the 1865 which acco-
                                            modates a variety of component types, including radial, axial, and chip components. Its shield-
· Direct Reading of Measured Current        ed case reduces electrical noise and interfernce and includes a cover interlock switch and remote
· Keypad Lockout with Password              start for maximum operator safety.
  Protection
                                                                                              For more detailed specifications,
· RS-232 and I/O Interfaces                                                                                                visit
· Optional IEEE-488 Interface                                                                            www.quadtech.com
                                                                                                  For more information about special
                                                                                                  purchase, rent & lease options, call
                                                                                                                    1-800-253-1230
                                                                                                               Fax 1-978-461-4295
                                                                                                              Intl. 1-978-461-2100




                  CompuMess Elektronik GmbH, Lise-Meitner-Str. MEGOHMMETERS · MILLIOHMMETERS
                                         HIPOT · LCR METERS ·
                                                               1, 85716 Unterschleißheim
            Tel.: (089) 321 501-0; Fax: (089) 321 501-11; eMail: info@compumess.de; www.compumess.de; www.netzteile.de
                         1865 Megohmmeter                                                                  Dimensions:               Benchtop unit with tilt back bail
                                                                                                                                     (w x h x d) :(17.5 x 5.25 x 16in)
                         Resistance Range:          7 ranges (automatic or manual)
Product Specifications
                                                                                                                                     (444.5 x 133.4 x 406.4mm)
                                                    1 x 103 to >1 x 1014  (voltage dependent)
                                                    1 x 106 to >1 x 1014 at 1000V DC                       Weight:                   18 lbs (8.2kg) - Net
                                                    1 x 105 to 1 x 1013 at 100V DC                                                   26 lbs (11.8kg) - Shipping
                                                    1 x104 to 1 x 1012 at 10V DC
                                                    1x 103 to 1x 1011 at 1V DC                             Power:                    90 to 250V,
                                                                                                                                     47 or 63 Hz
                         Resistance                                                                                                  60W max
                         Accuracy:                  ± [0.45% + {(Rx/Vx)(0.0005 FS + 2pA) +
                                                    30  /Rx}100%]                                          Environmental:            In accordance with MIL-T-28800D,
                                                    Rx: Measured resistance in ohms                                                  Type 3, Class 5, Syle E & F
                                                    Vx: Programmed voltage in volts                                                  Operating: 0°C to +50°C,
                                                    FS: Full scale current range in amperes                                          Storage: -40°C to +70°C

                         Voltage Range:             1 to 1000 volts programmable in two ranges

                         Voltage Accuracy:          1 - 100V: 1% of setting + 1V, 25mV resolution
                                                    100 - l000V: 1% of setting + 2V, 250mV
                                                    resolution
                         Output Voltage
                         Impedance:                 1k, ± 5%

                         Current Limited:           < 2mA

                         Current Measure:           1 x10-13 - 1 x 10-3A

                         Current* Accuracy: 1nA- 1mA ± [0.5% + (0.0005FS + 2 pA)]
                                            100pA- 1nA ±[1% + (0.0005FS + 2pA)]
                                            1pA - 100pA ± [10% + (0.0005FS + 2 pA)]
                                            *Specified at front panel connectors                                      1865-51 Shielded Lead Set (optional accessory)

                         Input Impedance:           5k, ± 5%

                         Measure Limits:            Pass/Fail (1 limit)

                         Display:                   · LCD graphics display
                                                    · Voltage warning indicator
                                                    · Pass/Fail indicator

                         Test Cycle:                Manual:            Charge, Measure, Discharge
                                                    Automatic:
                                                    Charge: (0 - 300 sec), Measure: (0 - 999 sec)
                                                    Dwell: (0 - 300 sec), Discharge: (0 - 300 sec)

                         Interfaces:                Standard: RS-232,I/O Port (w/safety interlock)
                                                    Optional: IEEE-488 & Data/ Program Storage                   1865-52 Component Test Fixture (optional accessory)
                                                    31/2:" disk
                                                                                                           1865-52 Component Test Fixture
                         Input Terminals:           Four sheathed banana jacks, front or rear
                                                                                                           Components                Leaded Module: Body Size < 1inch
                                                    (optional) mount
                                                                                                           Accomodated:                                  Lead Size < 1.4mm dia
                                                    · + unknown (red)            · guard (blue)
                                                                                                                                     SMD Module:         Width: 0.5 to 10mm
                                                    · - unknown (black)          · ground (green)
                                                                                                                                                         Height: 0.5 to 10mm
                         Other Features:            · Menu Programming           · Stop on Pass                                                          (Diameter 0.5 to 3.0mm)
                                                    · Automatic Zeroing          · Safety Interlock                                                      Length: 0.1 to 8mm
                                                    · Stored Test Conditions     · Keypad Lockout                                    Alligator Clips: < 5mm diameter
                                                    · Measurement units (engineering or scientific)        Dimensions:               (w x h x d): (8 x 5.5 x 9in), (203 x 140 x 230mm)
                                                    · Measurement averaging (1 - 400)                      Weight:                   5 lbs (2.3kg) - Net, 9 lbs (4kg) - Shipping

                         Ordering Information
                          1865-00     Megohmmeter/IR Tester                                           Optional Accessories:
                          Includes:                                                                   1865-01 IEEE Interface Option (factory installed)
                                                                                                      1865-02 Data/Extended Program Storage Option (factory installed)
                          150073      Instruction Manual
                                                                                                      1865-03 Rear Panel Input Option (factory installed)
                          XX          Calibration Certificate Traceable to NIST
                                                                                                      1865-50 Rack Mount Kit
                          700070      AC Power Cable
                                                                                                      1865-51 Shielded Lead Set
                          800014      100k Capacitor Adaptor
                                                                                                      1865-52 Component Test Fixture
                          800015      1M Capacitor Adaptor
                                                                                                      1865-70 IEEE Interface Option (field retrofit)
                          520049      Spare T2.5A, 250V Power Line Fuse
                                                                                                      Calibration Data

                                                    For more detailed specifications, visit www.quadtech.com
                                            For more information about special purchase, rent & lease options, call
                                           1-800-253-1230 · Fax 1-978-461-4295 · Intl. 1-978-461-2100
                                                         5 Clock Tower Place, 210 East, Maynard MA 01754                                                                 P/N 030047/A4

                                   CompuMess Elektronik GmbH, Lise-Meitner-Str. 1, 85716 Unterschleißheim
                         Tel.: (089) 321 501-0; Fax: (089) 321 501-11; eMail: info@compumess.de; www.compumess.de; www.netzteile.de
  1868 Series 1868A-D Megohmmeters
USES:                                           High Resistance Testing To 1P
· Production Testing of Insulation on
  Switches, Cables, Transformers, Motors and    Introduction
  Other Devices                                 The 1868 Megohmmeters (Models A&D) are precise, full-featured digital instruments
· Testing of Capacitive Components              with both resistance and current measurement modes. They are capable of measuring
                                                Insulation Resistance from 10k to 1P at programmable DC test voltages between 10
· Insulation Properties of Natural and
                                                and 1000V (1868A) or 50 and 5000V (1868D). For charging high capacitive devices,
  Synthetic Materials such as Oils, Plastics,
                                                the 1868A supplies a powerful charge current up to 80mA and the 1868D up to 18mA.
  Rubber or Porcelains
                                                Remote I/O, RS232 and IEEE-488.2 interfaces for remote control operation and com-
· High Resistance Testing of Electrical         munication with other instrumentation are standard on both models.
  Connections
                                                Description
FEATURES:                                       Wide Measurement Ranges: Measure resistance from 10k to 1P with basic accu-
                                                racy of +2% up to 1T. Measure Leakage Current down to 1pA with basic accuracy
· Measurement Range: 10k to 1P                  of +2%.
· Basic Accuracy: ± 2% to 1T                    Programmable Test Voltage: From 10V to 1kV in 1V increments on the 1868A
· Programmable Test Voltage: 10V to 1kV         Model, and from 50V to 5kV in 1V increments and on the 1868D Model.
  (1868A), 50V to 5kV (1868D)                   Powerful Charge Capability: The 1868A Model offers current ranges of
· Selectable Charge Current: to 80mA            2mA/25mA/80mA and the 1868D offers current ranges of 2mA/10mA/18mA.
  (1868A), to 18mA (1868D)                      Currents above 2mA are password protected for safety.
· Leakage Current Measurement Down to           Programmable Charge, Delay & Discharge Times: Allows testing of highly capac-
  1pA (± 2% Basic Accuracy)                     itive devices without false failures.
· Programmable Charge, Delay and                Standard Interfaces: The 1868 models come standard with RS232 and GPIB inter-
  Discharge Times                               faces for PC connection in automated applications. Incorporate the 1868 unit into an
                                                automated manufacturing environment with remote start and bin results output
· 5 Programmable Limit Sets
                                                through the rear panel PLC interface.
· Internal Storage of 4 Tests (20 Steps each)
                                                Graphical Display: The backlit display enables quick, easy setup and display of the
  Plus Additional Storage with Optional
                                                test results as text, bar graph or X-Y graph. The X-Y Graph is an excellent tool to view
  PCMCIA Memory Card
                                                current or resistance measurements over time.
· Standard RS232, IEEE and Remote I/O           Test Storage: Store 2 Global Setups including instrument setup and sequence setups,
  Interfaces                                    4 user-programmable test sequences (up to 20 steps each), and 4 pre-programmed test
· Graphic LCD with Text, Bar Graph and          sequences internally. The PCMCIA interface increases test setup storage with the
  X-Y Graph Displays                            capability of uploading/downloading test setups to a memory card.
· Current or Resistance Display Mode
· 4 Ranges Plus Automatic Ranging
· Automatic Zeroing
· Contact Check Function
                                                                                         CompuMess Elektronik GmbH
                                                                                           Lise-Meitner-Str.1, 85716 Unterschleissheim
· Floating Power Supply (1868A) for                                                           Tel 089-321501-0 Fax 089-321501-11
                                                                                         http://www.compumess.de oder http://www.netzteile.de
  Measurements on Grounded Devices




                                                             HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS
1868A-D Megohmmeter
Resistance Range:           10k - 1P (dependent on test voltage)      Display:               LCD graphic display
                            4 Ranges or Auto Range:                                          3.5 digit resolution
                            Range 1: 1pA to 10nA                                             Bar Graph: 8 ranges/2 decades/graph,
                            Range 2: 2nA to 100nA                                            Fixed or auto-scale
                            Range 3: 20nA to 10uA                                            Line Graph: X-Y: I or R versus Time
                            Range 4: 2uA to 1mA                                              Indicators: HV, Measure, Bin, Remote
Resistance Accuracy: For Rx <20G (Vx): ±(2% of reading +5cnt)         Standard Interfaces:   RS232, IEEE 488, Remote I/O
                     For Rx >20G (Vx): ±(Vx/1pA) in ohms.             Internal Storage:      2 Global Setups (Save Instrument Setup
                     Max reading: Vx/0.1pA                                                   and Sequence Setups)
                     Rx = Measured resistance in ohms                                        4 User-Programmable Sequence Setups,
                     Vx = Programmed voltage in volts                                        (Up to 20 Steps each)
Voltage Range:              10 to 1000V DC in 1V steps, 1868A                                4 Pre-Programmed Sequence Setups
                            50 to 5000V DC in 1V steps, 1868D         External Storage:      PCMCIA Interface for 2M SRAM Memory
Voltage Accuracy:           2% of setting                                                    Cards (20 Setups/20 Steps each)
                            ±1V in uncalibrated mode                  Input Terminals:       HV: BNC Connector
Source Resistance:          100, 1kV (1868A); 3.3k, 5kV (1868D)                              I IN: Triaxial Connector
                                                                                             GUARD: Sheathed Banana Plug
Charge Current:             2/25/80mA (1868A); 2/10/18mA (1868D)
                                                                      Safety:                Interlock Plug
Discharge Resistance: 100k 
                                                                                             Password-protected Charge Current
Current Measure:            1pA to 1mA                                                       (For I > 2mA)
Current Accuracy:           ±[2% of reading +5cnts +1pA]                                     HV Break Red LED/Push Button
Input Impedance:            For Ix > 10nA: 10k ; For Ix < 10nA: 1M    Dimensions:            (w x h x d):12.5 x 4.5 x 15.5 inches
                            Ix = Programmed current in amperes                               (312.5 x 112.5 x 387.5 mm)
Measure Limits:             5 Programmable Limit Sets                 Weight:                28 lbs (12.7kg) net
Measurement Result:         Resistance or Current                                            41 lbs (18.6kg) ship

Measurement Mode:           Continuous or Triggered                   Environmental:         Operating: 10°C to +40°C, 80% RH

Averaging:                  1-100 measurements                        Power:                 110V - 240V
                                                                                             50 or 60Hz
Test Speed:                 Trigger Mode: 40msec
                                                                                             400W Max (1.8A at 220V AC,
                            Continuous Mode: 90-4000msec                                     3.2A at 125V AC)
                            (Depending on # to Average)
Programmed Time:            Charge: 0 - 9999msec (Charge R: 200)
                            Measurement Delay: 0 - 9999msec
                            Discharge: 0 - 100sec or Auto mode
                            Auto: 0.1(Cx) as Vx goes to 1% Vt
                            Cx = Measured capacitance (µF)
                            Vx = Programmed voltage (V)
Zero/Offset:                Automatic Zeroing of All Ranges or
                            Individual Zeroing of 4 Ranges




Ordering Information
 1868A Megohmmeter: 2/25/80mA, 1kV                                   Optional Accessories:
 1868D Megohmmeter: 2/10/18mA, 5kV                                              Calibration Data
                                                                     2100-04 2MEG SRAM PCMCIA Memory Cards
 Includes:
 150753        Instruction Manual
 700070        Power Cable
 1868-01       1868A Alligator Clip Lead Set (1868A only)
 1868-02       19" Rack Mount Brackets (Set of 2)
 1868-03       1868D Alligator Clip Lead Set (1868D only)
 1868-04       Interlock Plug
 N/A           Calibration Certificate Traceable to NIST




             CompuMess Elektronik GmbH
                Lise-Meitner-Str.1, 85716 Unterschleissheim
                   Tel 089-321501-0 Fax 089-321501-11
             http://www.compumess.de oder http://www.netzteile.de
       LR Series                             LR2000 Milliohmmeter
                                             A Precision Low Resistance Meter
USES:
                                             Introduction
· Production Testing of Contact Resistance
                                             The QuadTech LR2000 Milliohmmeter with its LCD display and menu type
  of Switches, Relays, Connectors, Cables,
                                             front panel programming assures that low resistance measurements on switch-
  and Other Low Resistance Devices
                                             es, relays, connectors, cables, and other devices can be made quick and easy.
· Testing of Low Value Resistors, Fuses,     With a basic accuracy of 0.05% the instrument offers a wide measurement
  Squibs, and Heating Elements               range from 1µohm to 2Mohms. For remote operation and production appli-
                                             cations the unit comes standard with an RS-232 interface, plus IEEE-488 and
· Winding Resistance of Motors,
                                             handler interfaces are available as options. For measurement integrity, contact
  Transformers, Solenoids, and Ballasts
                                             to the test device is made via a 4-terminal Kelvin connection that incorporates
· Conductivity Evaluation in Product         an automatic zeroing function to compensate for lead errors.
  Design                                     Description
· Incoming Inspection and Quality            Wide Measurement Range: Eight measurement ranges from 20m to 2M
  Assurance Testing                          with constant current between 1A and 1µA. For "dry" contact measurements
                                             (those contacts whose resistance can be altered by excessive voltage potential)
FEATURES:                                    the LR2000 can be limited to 20mV on selected measurement ranges.
                                             Test Signal: Besides the standard DC test signal, the LR2000 provides a signal
· 1µ - 2M Measurement Range
                                             reversal mode for eliminating thermal EMF's, and pulsed current mode for
· 1µA - 1A Constant Current                  minimizing errors caused by device heating.
· 0.05% Basic Measurement Accuracy           Precision Measurements: With a basic measurement accuracy of 0.05% the
                                             instrument makes for consistent, reliable test results.
· Measurement Speed to 15/second
                                             Graphical Display: 240 x 64 dot matrix graphical LCD display for user friend-
· Graphical LCD Display                      ly menu programming and display of measured results.
· Four Terminal Kelvin Connection            Measurement Rate: Three measure modes of 15, 6 and 1.5 measurements per
· Automatic Zeroing                          second with varying degrees of accuracy.
· RS-232 Interface Standard                  Pass/Fail Testing: Provides a programmable Hi/Lo comparator function in
                                             absolute value or %, as well as 8 sorting bins for categorization of components.
· IEEE and Handler Interfaces, Optional
                                             Zeroing: An automatic zeroing functions reduces the effects of lead resistance
· Automatic Hi/Lo Comparator Limits          through the front panel 4-terminal Kelvin connection.
· Pass/Fail Sorting (8 Bins)                 Interfaces: For remote control, or adaptation to a production type environ-
· Voltage Limiting for Dry Contact Testing   ment, the LR2000 includes an RS-232 interface standard. An IEEE-488 and
                                             Handler interface are available as an option.
· Keypad Lockout
· Programmable Delay Times
                                                                               CompuMess Elektronik GmbH
                                                                                 Lise-Meitner-Str.1, 85716 Unterschleissheim
                                                                                    Tel 089-321501-0 Fax 089-321501-11
                                                                               http://www.compumess.de oder http://www.netzteile.de




                                                       HIPOT · LCR METERS · MEGOHMMETERS · MILLIOHMMETERS
LR2000 Milliohmmeter Features
Resistance Range:

Range F.S.                    Resolution                         Accuracy                                  Test I (Typical)

20m                           1µ                                 ±(0.1% of rdg + .006m)                    1A
200m                          10µ                                ±(0.05% of rdg + .06m)                    100mA
2                             100µ                               ±(0.05% of rdg + .6m)                     10mA
20                            1m                                 ±(0.05% of rdg + 6m)                      1mA
200                           10m                                ±(0.05% of rdg + 40m)                     1mA
2k                            100m                               ±(0.05% of rdg + .2)                      1mA
20k                           1                                  ±(0.1% of rdg + 2)                        100µA
200k                          10                                 ±(0.2% of rdg + 20)                       10µA
2M                            100                                ±(0.4% of rdg + 200)                      1µA

Test Signal:             Modes: DC+, DC-, Pulse+, Pulse-,                   Dimensions:                (w x h x d): 12.5 x 4.0 x 13.5in
                         Pulse+/- and STBY                                                             (312.5 x 100.0 x 337.5mm)
                         Dry Circuit: Open Circuit Voltage <20mV            Weight:                     10.85 lbs. (5kg) net, 15.2 lbs. (7kg) ship
                         for 200m, 2 and 20 ranges
Measurement Rate:        Fast:        15 measurements/second                Environmental:             Specifications: +15°C to +35°C, 75% RH
                         Medium:      6 measurements/second                                            Operating: 10°C to +40°C
                         Slow:        1.5 measurements/second                                          Storage: 0°C to +50°C
                                                                                                       Humidity: 10 - 90% RH
Measurement Mode:        Continuous or Trigger                                                         Pollution Degree 2
Trigger:                 Internal                                                                      Installation Category II
                         Manual                                             Power:                     · 90 - 125V AC
                         External (IEEE or Handler)                                                    · 190 - 250V AC
Delay Times:             Trigger Delay: 5 - 1000ms                                                     · 48 - 62 Hz
                         Measurement Delay: 0 - 100s                                                   · 80W max
Ranging:                 Automatic or Hold Range
Zeroing:                 Open and short circuit compensation
Averaging:               1 - 10
Comparator:              Hi/Lo Limits (Value or %)
Binning:                 Hi/Lo Limuts (8 bins in %)
Indication:              Audible Alarm programmable: HI LO or
                         OFF for Pass or Fail Result
Display:                 240 by 64 dot matrix LCD display
Setup Storage:           Auto recall on power-up
Lock:                    Keypad Lockout
Interfaces (Standard): RS-232
Interfaces (Optional):   IEEE-488 and Handler
Test Terminals:          Front: 4 Sheathed Banana & 1 GND



 Ordering Information
LR2000 Milliohmmeter                                                        Optional Accessories:
                                                                                        B/A Calibration Data
Includes:
                                                                            LR2000-50   Kelvin Clip Lead Set
QuadTech P/N        Item
                                                                            700171      IEEE-488/Handler Interface
150713              LR2000 Instruction Manual
LR2000-50           Lead Set: 4 Banana Connectors to 2 Kelvin Clips
4200-0300           AC Power Cord
520026/520138       Power Line Fuses (1.0A 250V/0.5A 250V)
                    Calibration Certificate Traceable to NIST




              CompuMess Elektronik GmbH
              For more detailed specifications, visit www.quadtech.com
        For more information about special purchase, rent & lease options, call