MEDICAL LINEARS
FEATURES
MEDICAL LINEARS
· World-wide AC input ranges and safety standards · Medical leakage current does not exceed 10µA
· Industry standard packages · All electrolytic caps rated at 105ºC
· Approved to UL544, IEC601 and CSA234 L3 · Transformer insulation meets Class F (155ºC)
· Burn-in with cycling; 3-year warranty · Exceed FCC and CISPR11 Class B conducted emissions
· MTBF over 200,000 hours per Mil-Hdbk 217E · marked to LVD
SPECIFICATIONS overvoltage protection. OVP is not optional on L and LL case units;
however, it is built in on 5V models.
AC Input Short Circuit Protection
100, 120, 220 and 240Vac: +10%, -13%, 47-63Hz. Tolerance for 230Vac Automatic current limit/foldback.
operation is +15%, -10%. Derate output current 10% for 50Hz operation.
Remote Sensing
L and LL case units have AC input of 95-132/190-264Vac, 47-63Hz.
Provided on all models 3 amps and above; open sense lead protection
Range can be extended to 90-132/180-264Vac by derating output
built-in.
current 20%.
Stability
DC Output ±0.05% for 24 hours after warmup.
See voltage rating chart. Adjustment range ±5% minimum.
Temperature Rating
Line Regulation 0 to 50°C full rated, derated linearly to 40% at 70°C.
±0.05% for a 10% change.
Temperature Coefficient
Load Regulation ±0.01%/°C maximum.
± 0.05% for a 50% load change.
Efficiency
Output Ripple 5V units: 45%; 12 and 15V units: 55%; 20 and 24V units: 60%.
3mV +0.05% of output voltage, peak to peak maximum.
Vibration
Transient Response Per Mil-Std-810D, Method 514.3, Category 1, Procedure 1.
50 microseconds for 50% load change.
Shock
Overvoltage Protection Per Mil-Std-810D, Method 516.3, Procedure III.
Built-in on all 5V models, set at 6.2V ±0.4V. Other models use optional
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
118
Model (Single) Voltage Current Leakage Current Case
ML5-OV-A 5V 1.0A 2.0µA L
MB5-3-OV-A 5V 3.0A 6.0µA B
MC5-6-OV-A 5V 6.0A 6.0µA C
MD5-12/OV-A 5V 12.0A 7.5µA D
ML12-0.5-A 12V 0.5A 2.0µA L
MB12-1.7-A 12V 1.7A 6.0µA B
MC12-3.4-A 12V 3.4A 6.0µA C
MD12-6.8-A 12V 6.8A 7.5µA D
ML15-0.4-A 15V 0.4A 2.0µA L
MB15-1.5-A 15V 1.5A 6.0µA B
MC15-3-A 15V 3.0A 6.0µA C
MD15-6-A 15V 6.0A 7.5µA D
ML24-0.28-A 24V 0.28A 2.0µA L
MB24-1.2-A 24V 1.2A 6.0µA B
MC24-2.4-A 24V 2.4A 6.0µA C
MD24-4.8-A 24V 4.8A 7.5µA D
MB28-1-A 28V 1.0A 6.0µA B
MC28-2-A 28V 2.0A 6.0µA C
MD28-4-A 28V 4.0A 7.5µA D
Model (Dual) Output 1 Output 2 Leakage Current Case
MLL12-0.25-A +12V@0.25A -12V@0.25A 2.0µA L
MEDICAL LINEARS
MLL15-0.2-A +15V@0.20A -15V@0.20A 2.0µA L
MAA15-0.8-A +12/15V@1.0/0.8A 12/15V@1.0/0.8A 6.0µA AA
MBB15-1.5-A +12/15V@1.7/1.5A -12/15V@1.7/1.5A 6.0µA BB
MCC15-3-A +12/15V@3.43/3.0A -12/15V@3.43/3.0A 7.5µA CC
MAA512-A 5V@2.0A 9-15V@0.5A 6.0µA AA
MBB512-A 5V@3.0A 9-15V@1.25A 6.0µA BB
MCC512-A 5V@6.0A 9-15V@2.5A 7.5µA CC
MAA524-A 5V@2.0A 18-24V@0.3A 6.0µA AA
MBB524-A 5V@3.0A 18-24V@0.8A 6.0µA BB
MCC524-A 5V@6.0A 18-24V@2.0A 7.5µA CC
Model (Triple) Output 1 Output 2 Output 3 Leakage Current Case
MTLL-5W-A +5V@0.5A +12 to 15V@0.1A -12 to 15V@0.1A 2.0µA L
MTAA-16W-A 5V@2.0A +12 to 15V@0.4A -12 to 15V@0.4A 2.0µA AA
MBAA40W-A 5V@3.0A +12/15V@1.0/0.8A -12/15V@1.0/0.8A 6.0µA BAA
MCAA60W-A 5V@6.0A +12/15V@1.0/0.8A --12/15V@1.0/0.8A 6.0µA D
MCBB75W-A 5V@6.0A +12/15V@1.71/1.5A -12/15V@1.7/1.5A 7.5µA CBB
NOTE: For medical linear mechanical specifications, see pp. 56-59.
CompuMess Elektronik GmbH · Lise-Meitner-Str.1 · 85716 Unterschleissheim
Tel 089-321501-0 · Fax 089-321501-11 · www.compumess.de · www.netzteile.de
119