Hier Klicken für das PDF Dokument!!!
MENT1220
Universal 220 Watt Series
Medical Switch-Mode Power Supply
·Universal Input 100Vac 240Vac
·12V to 48V Single Output Models up to 220W
·Desktop Style/Non-vented case - No internal fan!
·Ribbed Case Surface to improve cooling
·Energy Star 2.0, Level V
·Certified to Medical (UL/IEC/EN60601-1) and
ITE (UL/IEC/EN60950-1) Standards
·IPX1 Rated Enclosure
·Class I and Class II input versions available.
·Modified and Custom Designs
Specifications All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
AC Input 100-240VAC, +/-10%, 47-63 Hz single phase MTBF >100,000 hours (calculated)
Input Current 115VAC: 4A max, 230VAC: 1.5A max Hold-up Time 21 ms min. @ 100VAC
Inrush Current 75A max. @ 264VAC, Cold Start Overvoltage Protection 130% of nominal output voltage
Input Fuses Internal 6.3A Primary Current Fuses are provided Overload Protection 110-150% of full rated load, auto-recovery
PFC, Harmonics IEC61000-3-2 Short Circuit Protection Hiccup Mode, auto-recovery
Earth Leakage Current <350 A typical @264VAC, 50Hz Switching Frequency 50-120kHz, variable
Efficiency 88% typical Medical Safety Standards EN/IEC/UL60601-1
No Load Power <0.5W ITE Safety Standards EN/IEC/UL60950-1
Transient Response 5ms max. for 50% load step Isolation: Input-Output: 4KVac, Input-GND: 1.5KVac
Max deviation is 5% Output-GND: 700Vdc
Output Voltage See chart Operating Temp. 0° - 70°C, derate above 40°C to
60% of full rated load at 70°C
Line Regulation: +/-0.5% max Storage Temperature -20 to +85°C
Load Regulation +/-3% max Relative Humidity 5% to 95%, non-condensing
Minimum Load Not required Altitude 0 to 10,000 ft
Ripple and Noise 1% pk-pk max., 20MHz BW, differential mode Case Material Black 94V-0 rated Polycarbonate
Turn-on Time 3 sec. maximum @ 115VAC Case Dimensions Desktop Style: 8.4" x 4.25" x 1.85"
Output Connections See chart and outline drawing Weight 1200g
______________________________________________________________________________________________________________________________ 1 of 3
SL Power Electronics Corp · 6050 King Drive Ventura,CompuMess Elektronik GmbH · Lise-Meitner-Str. 4 · D-85716 Unterschleißheim
CA 93003 · Phone 805.486.4565 · Fax 858.712.2040 · Email: info@slpower.com · www.slpower.com
Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
info@compumess.de · www.compumess.de · www.netzteile.de
MENT1220
Universal 220W Series
Medical Switch-Mode Power Supply
OUTPUT PARAMETERS
Model Number Output Volts Output Current (max) Max Watts Ripple (Vp-p max)
MENT1220A1251F01 12V 15.0A 180W 120mV
MENT1220A1551F01 15V 13.0A 195W 150mV
MENT1220A1851F01 18V 11.7A 210W 240mV
MENT1220A2451F01 24V 9.2A 220W 240mV
MENT1220A2851F01 28V 7.9A 220W 280mV
MENT1220A3251F01 32V 6.9A 220W 320mV
MENT1220A4851F01 48V 4.6A 220W 480mV
Model Number Key
MENT 1 220 A VV 51 F 01
Model Configuration: "01" = Standard. "02" and higher indicates a modified model.
Input Connector: "F" = IEC320 C14 grounded. See below for other options (consult factory for availability)
Output Connector: "51" - Molex 39-01-2060 or equivalent. Other options available, consult factory.
Output Voltage: "12" = 12Vdc, "15" = 15Vdc, "18" = 18Vdc, "24" = 24Vdc, "28" = 28Vdc, "32" = 32Vdc, "48" = 48Vdc
Model Configuration: "A" = Model Configuration
Output Power: "220" = 220 Watts
# of Outputs "1" = Single Output
Product Family: "M" = Medical, "E" = External, "NT" = Model Series Designator
EMI/EMC Compliance
Conducted Emissions FCC Part 15, Subpart B, EN55011 Class B
Radiated Emissions FCC Part 15, Subpart B, EN55011 Class B
Voltage Fluctuations EN61000-3-3, Line Flicker
Static Discharge Immunity EN61000-4-2, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity EN61000-4-3, 3V/m.
EFT/Burst Immunity EN61000-4-4, 2kV/5kHz..
Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode
Conducted RF Immunity EN61000-4-6, 3Vrms
Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m
Voltage Dip Immunity EN61000-4-11 crit. A, 100Vac 60Hz, 40%/5 cycles with 70% full load.
Line Frequency Harmonics EN61000-3-2, Class A
Input Receptacle Options Output Connector
"N" = IEC320 C8 "F" = IEC320 C14 "Q" = IEC320 C18
Ungrounded Grounded Ungrounded See outline drawings below for connector information and pinout.
______________________________________________________________________________________________________________________________ 2 of 3
SL Power Electronics Corp · 6050 King Drive Ventura,CompuMess Elektronik GmbH · Lise-Meitner-Str. 4 · D-85716 Unterschleißheim
CA 93003 · Phone 805.486.4565 · Fax 858.712.2040 · Email: info@slpower.com · www.slpower.com
Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
info@compumess.de · www.compumess.de · www.netzteile.de
MENT1220
Universal 220W Series
Medical Switch-Mode Power Supply
Outline Drawing: ("F" input receptacle shown)
Output Connector Pinouts
12V thru 18V Models: 24V thru 48V Models:
Pin Wire Color Desig. Wire Color Desig.
1 Red +Vout Red +Vout
2 Blue GND N/A N/C
3 Green GND Green GND
4 White +Vout White +Vout
5 Yellow +Vout N/A N/C
6 Black GND Black GND
OUTPUT CONNECTOR: Molex 39-01-2060 or equiv.
MATING CONNECTOR: Molex 39-01-2061 or equiv.
Data Sheet © 2010 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. Rev. 5-10-2010
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
______________________________________________________________________________________________________________________________ 3 of 3
SL Power Electronics Corp · 6050 King Drive Ventura,CompuMess Elektronik GmbH · Lise-Meitner-Str. 4 · D-85716 Unterschleißheim
CA 93003 · Phone 805.486.4565 · Fax 858.712.2040 · Email: info@slpower.com · www.slpower.com
Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
info@compumess.de · www.compumess.de · www.netzteile.de