XDL Series
105 to 215 W Programmable Linear DC Power Supply
XDL 35-5 XDL 35-5T
Single Output Triple Output
Xantrex Technology Inc.
Precision laboratory linear DC power
Headquarters
The Xantrex XDL series represents the `next generation' of high performance laboratory power supplies. 8999 Nelson Way
The XDL provides multiple ranges for increased current capability at lower voltages and uses pure Burnaby, British Columbia
linear technology. Unlike other digitally controlled units, the XDL series provides both numeric and Canada V5A 4B5
rotary control while the illuminated keys and display legends provide instant confirmation of settings 604 422 8595 Phone
and status. 800 667 8422 Sales & Support
604 421 3056 Fax
For added convenience the Xantrex XDL series provides storage of up to 10 power supply set-ups in prg.info@xantrex.com
non-volatile memory (30 set-ups for a triple). There are also fully adjustable over-voltage and over-
current trips.The XDL series also provides full remote sense capability via dedicated sense terminals. 5916 195th Street NE
Arlington, Washington
The XDL triple output model features link and copy mode for convenience. When linked, keyboard and USA 98223
jog wheel control operates both outputs simultaneously. The copy function copies all settings for output 360 435 8826 Phone
1 to output 2. 360 925 5144 Fax
Product Features Edificio Diagonal 2A,
Multiple voltage/current ranges C/ Constitución 3, 4º2ª
Direct numeric entry and incremental rotary control of voltage and current 08960 Sant Just Desvern
Remote or local sense Barcelona, Spain
Illuminated keys and display legends General Tel: +34 93.470.5330
Up to ten store/recall set-ups (30 set-ups for triple output) General Fax: +34 93.473.6093
Power output display
Link and copy mode
Protection Features
Over voltage protection
Over current protection
Over temperature protection
Sense protection
Options
GPIB, RS-232 or USB interface (P models)
CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
info@compumess.de · www.compumess.de · www.netzteile.de www.xantrex.com
© 2004 Xantrex Technology Inc. All rights reserved. Xantrex is a registered trademark of Xantrex International. DS20040608R04_xdl. Printed in Canada
XDL Series
105 to 215 W Programmable Linear DC Power Supply
Electrical Specifications
Models 35-5 35-5T 35-5P 35-5TP 56-4 56-4P
Output Ranges: Range 1 0-35 V, 0-3A 0-35V, 0-3A 0-35V, 0-3A 0-35V, 0-3A 0-56V, 0-2A 0-56V, 0-2A
Range 2 0-15V, 0-5 A 0-15V, 0-5A 0-15V, 0-5A 0-15V, 0-5A 0-25V,0-4A 0-25V, 0-4A
Range 3 0-35V,0-500.0 mA 0-35V,0-500.0 mA 0-35V,0-500.0 mA 0-35V,0-500.0 mA 0-56V,0-500.0 mA 0-56V,0-500.0 mA
Outputs 1 3 1 3 1 1
Output Power 105 W 215 W 105 W 215 W 112 W 112 W
Interface (GPIB/RS-232/USB) No No Yes Yes No Yes
Voltage Setting By floating point numeric entry or rotary jog wheel; resolution 1mV
Current Setting By floating point numeric entry or rotary jog wheel; resolution 1mA or 0.1mA depending on range
Setting Accuracy Voltage - 0.03% ± 5 mV. Current 0.2% ± 5mA, 0.5mA
Output mode Operation in constant voltage or constant current modes with automatic cross-over and mode indication by LEDs.
DC Output Switch Sets output voltage and current levels to zero when Off.
Output Terminals 4mm terminals on 19mm (0.75") spacing
Load Regulation Voltage: <0.01% + 2mV Current: <0.01% + 250µA; <0.01% + 50µA on 500mA range (measured at output terminals using remote sense)
Line Regulation Voltage: <0.01% + 2mV for 10% line change Current: <0.01% + 250µA; <0.01% + 50µA on 500mA range
Ripple and Noise Typically <0.35%Vrms 2mVp-p CV mode, and <0.2mArms, <20µArms (500 mA range) CI mode
Transient Response 50µs to within 15mV of set level for a change in load current from full load to half load or vice versa
Temperature Coeffcient <±(50ppm+0.5mV)/ °C (voltage)
Remote Sense Eliminates up to 0.5V drop per lead. Remote sense operation selected from front panel and indicated by LED
Sense Terminals Recessed sprung sockets for direct insertion of wires. Duplicated on rear terminal block (P versions only)
General Specifications1
Operational AC Input Voltage 115V or 230V ± 10% (adjustable internally, option HV for factory set 230 VAC input), 50/60 Hz. Installation Category II
Operating Temperature Range 5 °C to 40 °C, 20% to 80% RH
Storage Temperature Range -40 °C to 70 °C
Dimensions (HxWxD) 6.3 x 5.5 x 11.4"/160 x 140 x 290 mm (XDL 35-5, XDL 35-5P, XDL 56-4, XDL 56-4P),
6.3 x 11.0 x 11.4"/160 x 280 x 290 mm (XDL 35-5T, XDL 35-5TP)
Weight 11.9 lb/5.4 kg (XDL 35-5, XDL 56-4), 12.1 lb/5.5 kg (XDL 35-5P, XDL 56-4P),
23.1 lb/10.5 kg (XDL 35-5T), 23.3 lb/10.6 kg (XDL 35-5TP)
Benchtop Operation Folding legs are incorporated that can be used to angle the front panel upwards when required
Rack Mount Operation 19 inch 4U mount for up to three single output units or one triple plus one single
Blanking plates available for un-used sections
Warranty 3 years
Approvals CE-marked units meet: EN61010-1 and EN61326
Output Protection and Metering Bus Interfaces (P Suffix versions)
Output Protection Output will withstand forward voltages of up to 2OV above rated out- USB Standard USB hardware connection Supplied with device
put voltage. Reverse protection by diode clamp for current upto 3A driver for Win 98 or above. Operates as a virtual COM port.
Fault Condition Trip The output will be shut down if any of the four trip conditions RS-232 Variable baud rate 19,200 max. Single instrument or Addressable
listed below occur. In addtion to the output being set Off, an RS232
isolated rear panel signal is also activated.
Over Voltage (OVP) Settable 1V to 40V (XDL 35-5) or 62V (XDL 56-4) in 0.1V steps GPIB Conforming with IEEE-488.1 and IEEE-488.2 (N.B. All three
interfaces incorporate full control, readback and status reporting)
Over Current (OCP) Settable 0.1A to 5.5A (XDL 35-5) or 4,4A (XDL 56-4) in 0.01A steps
BUS Type Selection From front panel (GPIB/RS232/USB)
Over Temperature Monitors internal temperature rise to protect against excess
ambient temperature or blocked ventilation slots. Address Selection From front panel (1 to 31)
Sense Error Monitors the voltage between the remote sense terminals and Baud Selection RS-232 only. From front panel (600 to 19200 baud)
output terminals to protect against mis-wiring Setting Resolution Voltage - 1mV, Current 0.1mA (0.1 mA on 500 mA range)
Trip Output Signal Isolated open-collector output signal on rear panel Accuracy See specifications under Outputs and Metering
Meter Resolution and Accuracy Remote Control Response Time
Voltage (CI mode): Resolution 10mV Interface Typically <80 ms
Accuracy ± (0.1% of reading + 10mV) Output Voltage Response time varies with range and load conditions. Typical time
Current (CV mode): Resolution 0.001A; 0.1A on 500mA range to settle to within 1% of the total excursion on a 35V/3A range with
Accuracy ± (0.2% + 0.005A); ± (0.2% + 0.5mA) on 500mA range full load is <25ms. With no load it is <7 ms for an upward charge
and <600ms for downward.
V x A: Resolution 0.01W; 0.001W on 500mA range
Accuracy ± (0.3% + 0.05W); ± (0.3% + 0.005W) on 500mA range
Auxiliary Output-XDL 35-5T and TP
Store/Recall Settings Output Voltage Switchable 2.7V, 3.3V or 5.0V. Accuracy better than ± 5%
Output Current >1.0A maximum. LED indication of over-current.
Number of Stores 10 (30 total on XDL 35-5T) plus power-down store
DC Ouput Switch Sets output voltage level to zero when Off.
Memory Type Non-volatile using EEPROM
Output Terminals 4mm terminals on 19mm (0.75") spacing. Duplicate terminals at
Parameters Stored Range, Set volts, Set current, OVP, OCP
rear (P versions only)
Recall System Settings are previewed onto the displays before being actioned
Output Protection Output will withstand up to 16 V forward voltage. Diode clamped
1
General Specifications apply for 5 to 40°C temperature range. Accuracy specifications apply for 18 to 28°C for reverse voltages and 3 Amps reverse current.
temperature range after 1 hour warm-up with no load and calibration at 23°C. Typical specifications are Load Regulation <1% for 90% load change
determined by design and not guaranteed.
Line Regulation <0.1% for 10% line change
Note: Specifications are subject to change without notice.