Medically Approved Plug & Play Power Supply 1340W
vite
Medical
Medically Approved patents pending
Ultra-high efficiency 1U size
The XVE addition to the Xgen family of medically approved power supplies
provides up to 1340W (peak power 1450W) in the most compact package of 1U x
PLUG & PLAY POWER 268mm x 127mm. Boasting an industry leading 16W/in3 and 90% efficiency, the
XVE can provide up to 12 isolated DC outputs. The XVE employs innovative plug
next generation power source & play architecture allowing users to instantly configure a custom power solution
in less than 5 minutes!
FEATURES
Ultra high efficiencies and high power density are made possible through the
· UL60601-1 and EN60601-1 approved combination of low loss technologies and best field-proven techniques in planar
· Less than 300µA leakage current magnetics and surface mount electronics. The significant system space savings and
· 4000VAC isolation reduced heat dissipation radically simplify system design.
· 1450W peak power for 10sec
The XVE can be populated with up to 6 powerMods selected from the table of
· Extra low profile: 1U height (40mm)
powerMods shown below.
· Ultra high efficiency up to 90%
· Plug & Play Power All configurations carry full safety agency approvals including UL60601-1and EN60601-1
- allows fast custom configuration and are CE marked. For alternative power interfaces contact support@excelsys.com
- allow easy logistics
powerMods powerPacs
· Reduced system heat dissipation
MODEL Vmin Vnom Vmax Imax Watts MODEL Watts
· Few electrolytic capacitors (all long life)
· Visual LED indicators Xg1 1.5 2.5 3.6 50A 125W XVE 1340W
· Series / Parallel of multiple outputs Xg2 3.2 5.0 6.0 40A 200W
· 5V bias standby voltage provided Xg3 6.0 12.0 15.0 20A 240W
· Individual output control signals Xg4 12.0 24.0 30.0 10A 240W
Xg5 28.0 48.0 58.0 6A 288W
APPLICATIONS INCLUDE
Xg7 5.0 24.0 28.0 5A 120W
· Clinical diagnostic equipment
Xg8 V1 5.0 24.0 28.0 3A 72W
· Medical lasers
V2 5.0 24.0 28.0 3A 72W
· Dialysis equipment
· For Standard applications see XCE
EFFICIENCY (typical) MECHANICAL SPECIFICATIONS
93 1
Third angle projection
92
91 50.00 97.75
Efficiency %
90 46.00 122.00 92.00
89
23.50
SLOT F
88
87 SLOT E
19.00
J1
70C MAX OPERATING TEMPERATURE
SEE INSTRUCTION MANUAL
INPUT AC 100V-240V 50/60Hz
86 80.00 TOP VIEW SLOT D
19.00 127.00
85
J2
SLOT C
19.00
84 SLOT B
70 85 100 115 130 145 160 175 190 205 220 235 260
SLOT A
Line Voltage VAC 23.50
40.40 29.50
All dimensions in mm.
Mounting Holes
4 M4 threaded holes on Base. Max screw penetration is 6mm from Base.
Series Fleximount Side Mounting Slots
Use with self-clinching studs type FH-M4-X or FH-832-X (X= stud length) from PEM, or equivalent
Alternatively, use Xgen Side Clamps from Excelsys. Part No. Z165 (drawing 61401)
CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
info@compumess.de · www.compumess.de · www.netzteile.de
1340W Medically Approved Plug & Play Power Supply Series
SPECIFICATION applies to configured units consisting of powerMods modules plugged into the appropriate powerPac
INPUT
Parameter Conditions/Decription Min Nom Max Units
Input Voltage Range Universal Input 85 264 VAC
120 380 VDC
Input Frequency Range 47 63 Hz
Medical
Power Rating XVE See Xgen Designers' Manual for derating versus input line voltage 1340(1450) W
Input Current XVE 85VAC in 1000W out 14.0 A
Inrush Current 230VAC @ 25°C 25 A
Undervoltage Lockout Shutdown 65 74 VAC
Fusing XVE 250V F15A HRC
OUTPUT
Parameter Conditions/Description Min Nom Max Units
powerMod Power As per powerMod table
Output Adjustment Range Manual: Multi-turn potentiometer. As per powerMod table
Electronic: See Xgen Designers' Manual
Minimum Load 0 A
Line Regulation For ±10% change from nominal line ±0.1 %
Load & Cross Regulation For 25% to 75% load change ±0.2 %
Transient Response For 25% to 75% load change Voltage Deviation 10 %
Settling Time 250 µs
Ripple and Noise 20MHz Bandwidth 1.0 % pk-pk
Overvoltage Protection Two-level. 1st level: Vset Tracking. 2nd level: Vmax (Latching) 110 125 %
Overcurrent Protection Straight line with hiccup activation at <30% of Vnom 110 120 %
See Designer's Manual for full details
Remote Sense Max. line drop compensation. (except Xg7, Xg8) 0.5 VDC
Overshoot 2 %
Turn-on Delay From AC In / Enable signal 300 / 30 ms
Rise Time Monotonic 5 ms
Hold-up Time For nominal output voltages at full load. 230VAC/115VAC 20 / 15 ms
Output Isolation Output to Output / Output to Chassis 500 / 500 VDC
GENERAL
Parameter Conditions/Description Min Nom Max Units
Isolation Voltage Input to Output 4000 VAC
Input to Chassis 1500 VAC
Efficiency 230VAC, 1340W @ 24V 90 %
Safety Agency Approvals EN60601-1, UL2601-1, CSA601-1 UL File No. E230761
Earth Leakage Current 250VAC, 60Hz, 25°C 300 µA
Signals See Xgen Series datasheet
Bias Supply Always ON. Current 250mA 4.9 5.0 5.1 VDC
Reliability Failures per million hours at 25°C and full load powerMod 0.98 fpmh
See Designers' Manual. powerPac excludes fans powerPac 0.92 fpmh
EMC
Parameter Standard Level Units
Emissions
Conducted EN55011, EN55022, FCC Level B
Radiated EN55011, EN55022, FCC Level B
Harmonic Distortion EN61000-3-2 Compliant
Flicker and Fluctuation EN61000-3-3 Compliant
Immunity
Electrostatic Discharge EN61000-4-2 Level 4
Radiated RFI EN61000-4-3 Level 3
Fast Transients - burst EN61000-4-4 Level 4
Input Line Surges EN61000-4-5 Class 4
Conducted RFI EN61000-4-6 10 V/m
Voltage Dips EN61000-4-11 (EN55024) 10 ms
ENVIRONMENTAL
Parameter Conditions/Description Min Nom Max Units
Operating Temperature -20 +70 °C
2 Storage Temperature -40 +85 °C
Derating See Designers Manual for full deratings versus temperature
Relative Humidity Non-condensing 5 95 %RH
Shock 3000 Bumps, 10G (16ms) half sine
Vibration 1.5G 10 200 Hz
NOTES 1. This product is not intended for use as a stand alone unit and must be installed by qualified personnel.
2. The specifications contained herein are believed to be correct at time of publication and are subject to change without notice.
3. All specifications at nominal input, full load, 25°C unless otherwise stated.
4. 1450W peak for 10s; powerMod output power must not exceed normal ratings
XVE Prelim Apr 08
CompuMess Elektronik GmbH · Lise-Meitner-Str. 1 · D-85716 Unterschleißheim
Telefon (089) 32 15 01 - 0 · Telefax (089) 32 15 01 - 11
info@compumess.de · www.compumess.de · www.netzteile.de